Comparison of cardiac output derived from FloTrac/Vigileo and impedance cardiography during major abdominal surgery

被引:11
作者
Kim, Ji-Yeon [1 ]
Kim, Bo-Ram [1 ]
Lee, Kang-Hun [1 ]
Kim, Kyung-Woo [1 ]
Kim, Jun-Hyun [1 ]
Lee, Sang-Il [1 ]
Kim, Kyung-Tae [1 ]
Choe, Won-Joo [1 ]
Park, Jang-Su [1 ]
Kim, Jung-Won [1 ]
机构
[1] Inje Univ, Ilsan Paik Hosp, Dept Anaesthesiol & Pain Med, Goyang, South Korea
关键词
Cardiac output; impedance cardiography; niccomo; continuous haemodynamic monitoring; FloTrac; Vigileo; THORACIC ELECTRICAL BIOIMPEDANCE; PRESSURE WAVE-FORM; NONINVASIVE ASSESSMENT; CLINICAL MEASUREMENT; ASSESSING AGREEMENT; HEART-FAILURE; THERMODILUTION; EQUIVALENCE; INDEX;
D O I
10.1177/0300060513487649
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Objectives Impedance cardiography (ICG) is a noninvasive technique that provides reasonably accurate measurements of cardiac output, but the usefulness of ICG in patients undergoing open abdominal surgery has not been validated. Methods Cardiac output was measured while patients underwent open gastrectomy using an ICG monitor (niccomo; ICG-CO); the results were compared with those measured using a FloTrac/Vigileo monitor (Flo-CO), which measures cardiac output by analysing the arterial waveform. Data collection commenced at the beginning of anaesthetic induction and continued until the patient was awake. Data were compared using the Bland-Altman analysis, and the clinical significance of the difference between the two methods was evaluated by calculating the percentage error (%). Results Eleven patients were monitored during surgery. The bias of the Flo-CO and ICG-CO values was -0.45l/min. The upper and lower limits of agreement were 0.96l/min and -1.85l/min, respectively. The percentage error was 28.5%. Electrocautery induced interference that transiently impaired the performance of the ICG monitor. Conclusions ICG provided useful information in evaluating the cardiac output of patients during abdominal surgery.
引用
收藏
页码:1342 / 1349
页数:8
相关论文
共 22 条
[1]   Equivalence of the bioimpedance and thermodilution methods in measuring cardiac output in hospitalized patients with advanced, decompensated chronic heart failure [J].
Albert, NM ;
Hail, MD ;
Li, JB ;
Young, JB .
AMERICAN JOURNAL OF CRITICAL CARE, 2004, 13 (06) :469-479
[2]   A NEW STROKE VOLUME EQUATION FOR THORACIC ELECTRICAL BIOIMPEDANCE - THEORY AND RATIONALE [J].
BERNSTEIN, DP .
CRITICAL CARE MEDICINE, 1986, 14 (10) :904-909
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Methodologies for assessing agreement between two methods of clinical measurement: are we as good as we think we are? [J].
Cecconi, Maurizio ;
Grounds, Michael ;
Rhodes, Andrew .
CURRENT OPINION IN CRITICAL CARE, 2007, 13 (03) :294-296
[5]   CARDIAC-OUTPUT MEASUREMENT IN CRITICAL CARE PATIENTS - THORACIC ELECTRICAL BIOIMPEDANCE VERSUS THERMODILUTION [J].
CLANCY, TV ;
NORMAN, K ;
REYNOLDS, R ;
COVINGTON, D ;
MAXWELL, JG .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1991, 31 (08) :1116-1121
[6]   A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques [J].
Critchley, LAH ;
Critchley, JAJH .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) :85-91
[7]   Impedance cardiography: Recent advancements [J].
Cybulski, Gerard ;
Strasz, Anna ;
Niewiadomski, Wiktor ;
Gasiorowska, Anna .
CARDIOLOGY JOURNAL, 2012, 19 (05) :550-556
[8]   Arterial pressure-based cardiac output monitoring: a multicenter validation of the third-generation software in septic patients [J].
De Backer, Daniel ;
Marx, Gernot ;
Tan, Andrew ;
Junker, Christopher ;
Van Nuffelen, Marc ;
Huter, Lars ;
Ching, Willy ;
Michard, Frederic ;
Vincent, Jean-Louis .
INTENSIVE CARE MEDICINE, 2011, 37 (02) :233-240
[9]   Cardiac output monitoring [J].
de Waal, Eric. E. C. ;
Wappler, Frank ;
Buhre, Wolfgang F. .
CURRENT OPINION IN ANESTHESIOLOGY, 2009, 22 (01) :71-77
[10]   Comparison of impedance cardiography with invasive hemodynamic measurements in patients with heart failure secondary to ischemic or nonischemic cardiomyopathy [J].
Drazner, MH ;
Thompson, B ;
Rosenberg, PB ;
Kaiser, PA ;
Boehrer, JD ;
Baldwin, BJ ;
Dries, DL ;
Yancy, CW .
AMERICAN JOURNAL OF CARDIOLOGY, 2002, 89 (08) :993-+