Digital tomosynthesis with an on-board kilovoltage imaging device

被引:104
作者
Godfrey, DJ [1 ]
Yin, FF [1 ]
Oldham, M [1 ]
Yoo, S [1 ]
Willett, C [1 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiat Oncol, Durham, NC 27710 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2006年 / 65卷 / 01期
关键词
image-guided radiotherapy; tomosynthesis; cone-beam CT; patient positioning; treatment verification;
D O I
10.1016/j.ijrobp.2006.01.025
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To generate on-board digital tomosynthesis (DTS) and reference DTS images for three-dimensional image-guided radiation therapy (IGRT) as an alternative to conventional portal imaging or on-board cone-beam computed tomography (CBCT). Methods and Materials: Three clinical cases (prostate, head-and-neck, and liver) were selected to illustrate the capabilities of on-board DTS for IGRT. Corresponding reference DTS images were reconstructed from digitally reconstructed radiographs computed from planning CT image sets. The effect of scan angle on DTS slice thickness was examined by computing the mutual information between coincident CBCT and DTS images, as the DTS scan angle was varied from 0 degrees to 165 degrees. A breath-hold DTS acquisition strategy was implemented to remove respiratory motion artifacts. Results: Digital tomosynthesis slices appeared similar to coincident CBCT planes and yielded substantially more anatomic information than either kilovoltage or megavoltage radiographs. Breath-hold DTS acquisition improved soft-tissue visibility by suppressing respiratory motion. Conclusions: Improved bony and soft-tissue visibility in DTS images is likely to improve target localization compared with radiographic verification techniques and might allow for daily localization of a soft-tissue target. Breath-hold DTS is a potential alternative to on-board CBCT for sites prone to respiratory motion. (c) 2006 Elsevier Inc.
引用
收藏
页码:8 / 15
页数:8
相关论文
共 16 条
  • [1] Initial application of digital tomosynthesis with on-board imaging in radiation oncology
    Baydush, AH
    Godfrey, DJ
    Oldham, M
    Dobbins, JT
    [J]. Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 : 1300 - 1305
  • [2] Cho Z.H, 1993, FDN MED IMAGING, P71
  • [3] Digital x-ray tomosynthesis: current state of the art and clinical potential
    Dobbins, JT
    Godfrey, DJ
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (19) : R65 - R106
  • [4] Digital tomosynthesis of hand joints for arthritis assessment
    Duryea, J
    Dobbins, JT
    Lynch, JA
    [J]. MEDICAL PHYSICS, 2003, 30 (03) : 325 - 333
  • [5] PRACTICAL CONE-BEAM ALGORITHM
    FELDKAMP, LA
    DAVIS, LC
    KRESS, JW
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06): : 612 - 619
  • [6] Godfrey D., 2005, THESIS DUKE U
  • [7] Optimization of the matrix inversion tomosynthesis (MITS) impulse response and modulation transfer function characteristics for chest imaging
    Godfrey, DJ
    McAdams, HP
    Dobbins, JT
    [J]. MEDICAL PHYSICS, 2006, 33 (03) : 655 - 667
  • [8] Emergent technologies for 3-dimensional image-guided radiation delivery
    Jaffray, DA
    [J]. SEMINARS IN RADIATION ONCOLOGY, 2005, 15 (03) : 208 - 216
  • [9] Flat-panel cone-beam computed tomography for image-guided radiation therapy
    Jaffray, DA
    Siewerdsen, JH
    Wong, JW
    Martinez, AA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05): : 1337 - 1349
  • [10] A new irradiation unit constructed of self-moving gantry-CT and linac
    Kuriyama, K
    Onishi, H
    Sano, N
    Komiyama, T
    Aikawa, Y
    Tateda, Y
    Araki, T
    Uematsu, M
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 55 (02): : 428 - 435