Visible and near infrared fluorescence spectral flow cytometry

被引:30
作者
Nolan, John P. [1 ]
Condello, Danilo [1 ]
Duggan, Erika [1 ]
Naivar, Mark [2 ]
Novo, David [3 ]
机构
[1] La Jolla Bioengn Inst, San Diego, CA USA
[2] Darkling X LLC, Los Alamos, NM USA
[3] De Novo Software, Los Angeles, CA USA
关键词
spectroscopy; ccd; dispersion; quantum dots; SPECTROSCOPY; SENSITIVITY; RESOLUTION; PARTICLES; CELLS;
D O I
10.1002/cyto.a.22241
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
There is a long standing interest in measuring complete emission spectra from individual cells in flow cytometry. We have developed flow cytometry instruments and analysis approaches to enable this to be done routinely and robustly. Our spectral flow cytometers use a holographic grating to disperse light from single cells onto a CCD for high speed, wavelength-resolved detection. Customized software allows the single cell spectral data to be displayed and analyzed to produce new spectra-derived parameters. We show that familiar reference and calibration beads can be employed to quantitatively assess instrument performance. We use microspheres stained with six different quantum dots to compare a virtual bandpass filter approach with classic least squares (CLS) spectral unmixing, and then use antibody capture beads and CLS unmixing to demonstrate immunophenotyping of peripheral blood mononuclear cells using spectral flow cytometry. Finally, we characterize and evaluate several near infrared (NIR) emitting fluorophores for use in spectral flow cytometry. Spectral flow cytometry offers a number of attractive features for single cell analysis, including a simplified optical path, high spectral resolution, and streamlined approaches to quantitative multiparameter measurements. The availability of robust instrumentation, software, and analysis approaches will facilitate the development of spectral flow cytometry applications. (c) 2012 International Society for Advancement of Cytometry
引用
收藏
页码:253 / 264
页数:15
相关论文
共 30 条
  • [1] Asbury CL, 1996, CYTOMETRY, V24, P234, DOI 10.1002/(SICI)1097-0320(19960701)24:3<234::AID-CYTO6>3.0.CO
  • [2] 2-H
  • [3] BUICAN TN, 1990, P SOC PHOTO-OPT INS, V1205, P126, DOI 10.1117/12.17787
  • [4] Chase ES, 1998, CYTOMETRY, V33, P267, DOI 10.1002/(SICI)1097-0320(19981001)33:2<267::AID-CYTO24>3.0.CO
  • [5] 2-R
  • [6] Dubelaar GBJ, 1999, CYTOMETRY, V37, P247, DOI 10.1002/(SICI)1097-0320(19991201)37:4<247::AID-CYTO1>3.3.CO
  • [7] 2-0
  • [8] Fuller RR, 1996, CYTOMETRY, V25, P144, DOI 10.1002/(SICI)1097-0320(19961001)25:2<144::AID-CYTO3>3.3.CO
  • [9] 2-K
  • [10] Gauci MR, 1996, CYTOMETRY, V25, P388, DOI 10.1002/(SICI)1097-0320(19961201)25:4<388::AID-CYTO11>3.0.CO