Reduced-Dose Low-Voltage Chest CT Angiography with Sinogram-affirmed Iterative Reconstruction versus Standard-Dose Filtered Back Projection

被引:85
作者
Pontana, Francois [1 ]
Pagniez, Julien [1 ]
Duhamel, Alain [2 ]
Flohr, Thomas [3 ]
Faivre, Jean-Baptiste [1 ]
Murphy, Colm [1 ]
Remy, Jacques [1 ]
Remy-Jardin, Martine [1 ]
机构
[1] Univ Lille Nord France, Hop Calmette EA 2694, Dept Thorac Imaging, F-59037 Lille, France
[2] Univ Lille Nord France, Hop Calmette EA 2694, Dept Med Stat, F-59037 Lille, France
[3] Siemens Healthcare, Computed Tomog Div, Forchheim, Germany
关键词
COMPUTED-TOMOGRAPHY ANGIOGRAPHY; IMAGE QUALITY; VASCULAR ENHANCEMENT; RADIATION-EXPOSURE; PULMONARY-ARTERIES; REDUCTION; ADULTS;
D O I
10.1148/radiol.12120414
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate image quality of low-voltage chest computed tomographic (CT) angiography with raw data-based iterative reconstruction (sonogram-affirmed iterative reconstruction) in comparison with image quality of standard-dose standard-voltage filtered back projection (FBP) CT. Materials and Methods: This prospective study was approved by the institutional review board, and the informed consent requirement was waived. Eighty consecutive patients who were referred for follow-up chest CT angiography underwent reduced-dose CT (hereafter, T2 examination) under technical conditions similar to those of the initial examination (hereafter, T1 examination), except the voltage selection was reduced by 20 kV with adaptation of the tube current to ensure a 50% reduction in CT dose index, and regular FBP was replaced by iterative reconstruction with sonogram-affirmed iterative reconstruction. The two techniques were compared by using paired tests (Student t test, Wilcoxon test, or McNemar test, according to the nature of variables). Results: When compared with standard-dose T1 studies, reduced-dose T2 images showed: (a) significantly less objective noise at the level of the trachea on mediastinal and lung parenchymal images (P < .001) and no significant difference in objective noise at the level of the aorta on mediastinal images (P = .507); (b) significantly higher signal-to-noise and contrast-to-noise (P < .001) ratios; (c) similar visual perception of noise on mediastinal (P = .132) and lung (P = .366) images, mainly rated as moderate; and (d) similar overall subjective image quality (P = .405). Conclusion: Raw data-based iterative reconstruction yielded equivalent subjective and improved objective image quality of low-voltage half-dose CT angiograms compared with standard-dose FBP CT images for an average dose-length product of less than 80 mGy . cm in this population. (C) RSNA, 2013
引用
收藏
页码:609 / 618
页数:10
相关论文
共 20 条
[1]   Reduced Lung-Cancer Mortality with Low-Dose Computed Tomographic Screening [J].
Aberle, Denise R. ;
Adams, Amanda M. ;
Berg, Christine D. ;
Black, William C. ;
Clapp, Jonathan D. ;
Fagerstrom, Richard M. ;
Gareen, Ilana F. ;
Gatsonis, Constantine ;
Marcus, Pamela M. ;
Sicks, JoRean D. .
NEW ENGLAND JOURNAL OF MEDICINE, 2011, 365 (05) :395-409
[2]   Dose Reduction in Helical CT: Dynamically Adjustable z-Axis X-Ray Beam Collimation [J].
Christner, Jodie A. ;
Zavaletta, Vanessa A. ;
Eusemann, Christian D. ;
Walz-Flannigan, Alisa I. ;
McCollough, Cynthia H. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (01) :W49-W55
[3]  
CPMP, 2001, BRIT J CLIN PHARMACO, V52, P223
[4]   The risks of radiation exposure related to diagnostic imaging and how to minimise them [J].
Davies, H. E. ;
Wathen, C. G. ;
Gleeson, F. V. .
BMJ-BRITISH MEDICAL JOURNAL, 2011, 342 :589-593
[5]   Evaluation of Peripheral Pulmonary Arteries at 80 kV and at 140 kV: Dual-Energy Computed Tomography Assessment in 51 Patients [J].
Gorgos, Andrei ;
Remy-Jardin, Martine ;
Duhamel, Alain ;
Faivre, Jean-Baptiste ;
Tacelli, Nunzia ;
Delannoy, Valerie ;
Remy, Jacques .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2009, 33 (06) :981-986
[6]   Image quality and radiation exposure at pulmonary CT angiography with 100-or 120-kVp protocol: prospective randomized study [J].
Heyer, Christoph M. ;
Mohr, Patrick S. ;
Lemburg, Stefan P. ;
Peters, Soeren A. ;
Nicolas, Volkmar .
RADIOLOGY, 2007, 245 (02) :577-583
[7]   Multidetector Computed Tomography Chest Examinations With Low-Kilovoltage Protocols in Adults: Effect on Image Quality and Radiation Dose [J].
Kim, Mi Jung ;
Park, Chul Hi ;
Choi, Soo Jin ;
Hwang, Kyung Hoon ;
Kim, Hyung Sik .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2009, 33 (03) :416-421
[8]   Radiation dose reduction in chest CT: A review [J].
Kubo, Takeshi ;
Lin, Pei-Jan Paul ;
Stiller, Wolfram ;
Takahashi, Masaya ;
Kauczor, Hans-Ulrich ;
Ohno, Yoshiharu ;
Hatabu, Hiroto .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2008, 190 (02) :335-343
[9]   Low-Tube-Voltage, High-Tube-Current Multidetector Abdominal CT: Improved Image Quality and Decreased Radiation Dose with Adaptive Statistical Iterative Reconstruction Algorithm-Initial Clinical Experience [J].
Marin, Daniele ;
Nelson, Rendon C. ;
Schindera, Sebastian T. ;
Richard, Samuel ;
Youngblood, Richard S. ;
Yoshizumi, Terry T. ;
Samei, Ehsan .
RADIOLOGY, 2010, 254 (01) :145-153
[10]   Vascular Enhancement and Image Quality of MDCT Pulmonary Angiography in 400 Cases: Comparison of Standard and Low Kilovoltage Settings [J].
Matsuoka, Shin ;
Hunsaker, Andetta R. ;
Gill, Ritu R. ;
Oliva, Isabel B. ;
Trotman-Dickenson, Beatrice ;
Jacobson, Francine L. ;
Hatabu, Hiroto .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 192 (06) :1651-1656