Magnetic Resonance Imaging: An Accurate, Radiation-Free, Alternative to Computed Tomography for the Primary Imaging and Three-Dimensional Reconstruction of the Bony Orbit

被引:25
作者
Schmutz, Beat [1 ]
Rahmel, Benjamin [1 ]
McNamara, Zeb [1 ]
Coulthard, Alan [1 ]
Schuetz, Michael [1 ]
Lynham, Anthony [1 ]
机构
[1] Queensland Univ Technol, Inst Hlth & Biomed Innovat, Brisbane, Qld 4059, Australia
关键词
VOLUME MEASUREMENT; FRACTURES; IMPLANTS; COIL; MRI; CAD/CAM; WALL;
D O I
10.1016/j.joms.2013.08.030
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: To determine the extent to which the accuracy of magnetic resonance imaging (MRI) based virtual 3-dimensional (3D) models of the intact orbit can approach that of the gold standard, computed tomography (CT) based models. The goal was to determine whether MRI is a viable alternative to CT scans in patients with isolated orbital fractures and penetrating eye injuries, pediatric patients, and patients requiring multiple scans in whom radiation exposure is ideally limited. Materials and Methods: Patients who presented with unilateral orbital fractures to the Royal Brisbane and Women's Hospital from March 2011 to March 2012 were recruited to participate in this cross-sectional study. The primary predictor variable was the imaging technique (MRI vs CT). The outcome measurements were orbital volume (primary outcome) and geometric intraorbital surface deviations (secondary outcome) between the MRI- and CT-based 3D models. Results: Eleven subjects (9 male) were enrolled. The patients' mean age was 30 years. On average, the MRI models underestimated the orbital volume of the CT models by 0.50 +/- 0.19 cm(3). The average intraorbital surface deviation between the MRI and CT models was 0.34 +/- 0.32 mm, with 78 +/- 2.7% of the surface within a tolerance of +/-0.5 mm. Conclusions: The volumetric differences of the MRI models are comparable to reported results from CT models. The intraorbital MRI surface deviations are smaller than the accepted tolerance for orbital surgical reconstructions. Therefore, the authors believe that MRI is an accurate radiation-free alternative to CT for the primary imaging and 3D reconstruction of the bony orbit. (C) 2014 American Association of Oral and Maxillofacial Surgeons
引用
收藏
页码:611 / 618
页数:8
相关论文
共 23 条
  • [1] The technology of MRI - the next 10 years?
    Blamire, A. M.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2008, 81 (968) : 601 - 617
  • [2] Computer-assisted secondary reconstruction of unilateral posttraumatic orbital deformity
    Gellrich, NC
    Schramm, A
    Hammer, B
    Rojas, S
    Cufi, D
    Lagrèze, W
    Schmelzeisen, R
    [J]. PLASTIC AND RECONSTRUCTIVE SURGERY, 2002, 110 (06) : 1417 - 1429
  • [3] Grimm J, 2006, MED SOLUTIONS, P94
  • [4] Hempel M, 1993, Recent Results Cancer Res, V130, P231
  • [5] RADIATION-DOSE TO THE LENS AND CATARACT FORMATION
    HENK, JM
    WHITELOCKE, RAF
    WARRINGTON, AP
    BESSELL, EM
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (05): : 815 - 820
  • [6] Orbital reconstruction with individually copy-milled ceramic implants
    Hoffmann, J
    Cornelius, CP
    Groten, M
    Pröbster, L
    Pfannenberg, C
    Schwenzer, N
    [J]. PLASTIC AND RECONSTRUCTIVE SURGERY, 1998, 101 (03) : 604 - 612
  • [7] A novel high-resolution magnetic resonance imaging microscopy coil as an alternative to the multislice computed tomography in postoperative imaging of orbital fractures and computer-based volume measurement
    Kolk, A
    Pautke, C
    Wiener, E
    Ploder, O
    Neff, A
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2005, 63 (04) : 492 - 498
  • [8] High Resolution Magnetic Resonance Imaging With an Orbital Coil as an Alternative to Computed Tomography Scan as the Primary Imaging Modality of Pediatric Orbital Fractures
    Kolk, Andreas
    Stimmer, Herbert
    Klopfer, Matthias
    Wolff, Klaus-Dietrich
    Hohlweg-Majert, Bettina
    Ploder, Oliver
    Pautke, Christoph
    [J]. JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2009, 67 (02) : 348 - 356
  • [9] Imaging of Orbital Trauma
    Kubal, Wayne S.
    [J]. RADIOGRAPHICS, 2008, 28 (06) : 1729 - 1739
  • [10] LAI KF, 2001, JHK COLL RADIOL, V4, P181