Reliability and Validity of Non-invasive Blood Pressure Measurement System Using Three-Axis Tactile Force Sensor

被引:12
|
作者
Yoo, Sun-Young [1 ]
Ahn, Ji-Eun [2 ]
Cserey, Gyoergy [3 ]
Lee, Hae-Young [4 ,5 ]
Seo, Jong-Mo [1 ,5 ]
机构
[1] Seoul Natl Univ, Inst Engn Res, Interuniv Semicond Res Ctr, Dept Elect & Comp Engn, Seoul 08826, South Korea
[2] Seoul Natl Univ, Interdisciplinary Program Bioengn, Seoul 03080, South Korea
[3] Pazmany Peter Catholic Univ, Fac Informat Technol & Bion, H-1083 Budapest, Hungary
[4] Seoul Natl Univ Hosp, Dept Internal Med, Seoul 03080, South Korea
[5] Seoul Natl Univ Hosp, Biomed Res Inst, Seoul 03080, South Korea
来源
SENSORS | 2019年 / 19卷 / 07期
关键词
non-invasive blood pressure measurement; radial arterial pressure; continuous blood pressure monitoring; three-axis tactile force sensor; ARTERY APPLANATION TONOMETRY; STATISTICAL-METHODS; VALIDATION; AGREEMENT;
D O I
10.3390/s19071744
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Blood pressure (BP) is a physiological parameter reflecting hemodynamic factors and is crucial in evaluating cardiovascular disease and its prognosis. In the present study, the reliability of a non-invasive and continuous BP measurement using a three-axis tactile force sensor was verified. All the data were collected every 2 min for the short-term experiment, and every 10 min for the long-term experiment. In addition, the effects on the BP measurement of external physical factors such as the tension to the radial artery on applying the device and wrist circumference were evaluated. A high correlation between the measured BP with the proposed system and with the cuff-based non-invasive blood pressure, and reproducibility, were demonstrated. All data satisfied the Association for the Advancement of Medical Instrumentation criteria. The external physical factors did not affect the measurement results. In addition to previous research indicating the high reliability of the arterial pulse waveforms, the present results have demonstrated the reliability of numerical BP values, and this implies that the three-axis force sensor can be used as a patient monitoring device.