Dose Received By Occupationally Exposed Workers At A Nuclear Medicine Department

被引:1
作者
Avila, O. [1 ]
Sanchez-Uribe, N. A. [1 ,2 ]
Rodriguez-Laguna, A. [3 ]
Medina, L. A. [4 ,5 ]
Estrada, E. [3 ]
Buenfil, A. E. [4 ]
Brandan, M. E. [4 ]
机构
[1] Inst Nacl Invest Nucl, AP 18-1027, Mexico City 11801, DF, Mexico
[2] Univ Nacl Autonoma Mexico, Fac Ciencias, Mexico City 04510, DF, Mexico
[3] Inst Nacl Cancerologia INCan, Mexico City 14080, DF, Mexico
[4] Univ Nacl Autonoma Mexico, Inst Fis, Mexico City 01000, DF, Mexico
[5] Univ Nacl Autonoma Mexico, INCan, Unidad Invest Biomed & Canc, Mexico City 4080, DF, Mexico
来源
MEDICAL PHYSICS | 2012年 / 1494卷
关键词
Radiation Protection; Thermoluminescent Dosimetry; Nuclear medicine;
D O I
10.1063/1.4764602
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Personal Dose Equivalent (PDE) values were determined for occupational exposed workers (OEW) at the Nuclear Medicine Department (NMD) of "Instituto Nacional de Cancerologia" (INCan), Mexico, using TLD-100 thermoluminescent dosemeters. OEW at NMD, INCan make use of radiopharmaceuticals for diagnosis and treatment of diseases. Radionuclides associated to a pharmaceutical compound used at this Department are I-131, F-18, Ga-68, Tc-99m, In-111 and C-11 with main gamma emission energies between 140 and 511 keV. Dosemeter calibration was performed at the metrology department of "Instituto Nacional de Investigaciones Nucleares" (ININ), Mexico. Every occupational worker used dark containers with three dosimeters which were replaced monthly for a total of 5 periods. Additionally, control dosemeters were also placed at a site free of radioactive sources in order to determine the background radiation. Results were adjusted to find PDE/day and estimating annual PDE values in the range between 2 mSv (background) and 9 mSv. The mean annual value is 3.51 mSv and the standard deviation SD is 0.78 mSv. Four of the 16 OEW received annual doses higher than the average +1 SD (4.29 mSv). Results depend on OEW daily activities and were consistent for each OEW for the 5 studied periods as well as with PDE values reported by the firm that performs the monthly service. All obtained values are well within the established annual OEW dose limit stated in the "Reglamento General de Seguridad Radiolegica", Mexico (50 mSv), as well as within the lower limit recommended by the "International Commission on Radiation Protection" (ICRP), report no.60 (20 mSv). These results verify the adequate compliance of the NMD at INCan, Mexico with the norms given by the national regulatory commission.
引用
收藏
页码:73 / 75
页数:3
相关论文
共 2 条
  • [1] *ICRP, 1990, INT COMM RAD PROT PU, V60
  • [2] *ICRU, 1992, 47 ICRU INT COMM RAD