Determination of Nanogram Quantities of Emodin in Pharmaceutical Preparations and Biofluids by Luminol-Myoglobin Chemiluminescence System

被引:3
作者
Zhang, Yun [1 ]
He, Xili [1 ,2 ]
Song, Zhenghua [1 ]
机构
[1] NW Univ Xian, Key Lab Synthet & Nat Funct Mol Chem, Minist Educ, Coll Chem & Mat Sci, Xian 710069, Peoples R China
[2] Xi An Jiao Tong Univ, Sch Sci, Xian 710049, Peoples R China
来源
SPECTROSCOPY-AN INTERNATIONAL JOURNAL | 2012年 / 27卷 / 02期
关键词
Emodin; luminol; myoglobin; chemiluminescence; serum; flow injection; ALOE-EMODIN; FLUORESCENCE; ANTHRAQUINONES; BEHAVIOR; CELLS;
D O I
10.1155/2012/741053
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Based on the quenching effect of emodin on the luminol-myoglobin (Mb) reaction, a sensitive method for the determination of nanogram level emodin by flow injection chemiluminescence (FI-CL) is presented for the first time. It was found that the CL intensity from luminol-Mb system could be inhibited in the presence of emodin, and the decrement of CL intensity was linearly proportional to the logarithm of emodin concentration in the range of 0.5-300 ng . mL(-1) (R = 0.9956) with the detection limit of 0.2ng . mL(-1) (3 sigma). At a flow rate of 2.0mL . min(-1), a complete determination of emodin, including sampling and washing, could be accomplished in 0.5 min with the relative standard deviations (RSDs) of less than 3.5% (n = 5). The proposed method was successfully applied to the determination of emodin in pharmaceutical preparations and human serum samples. The possible CL mechanism of luminol-Mb-emodin reaction was explained.
引用
收藏
页码:73 / 81
页数:9
相关论文
共 22 条
[21]   Electrocatalysis of emodin at multi-wall nanotubes [J].
Yin, Zhao-Hui ;
Xu, Qiao ;
Tu, Yi ;
Zou, Qiu-Ju ;
Yu, Jiu-Hong ;
Zhao, Yuan-Di .
BIOELECTROCHEMISTRY, 2008, 72 (02) :155-160
[22]   Identification and determination of active anthraquinones in Chinese teas by micellar electrokinetic capillary chromatography [J].
Zheng, WJ ;
Wang, SF ;
Chen, XG ;
Hu, ZD .
BIOMEDICAL CHROMATOGRAPHY, 2004, 18 (03) :167-172