Radiofrequency catheter ablation using cooled electrodes: Impact of irrigation flow rate and catheter contact pressure on lesion dimensions

被引:75
作者
Weiss, C
Antz, M
Eick, O
Eshagzaiy, K
Meinertz, T
Willems, S
机构
[1] Univ Hamburg, Hosp Eppendorf, Dept Cardiol, D-20246 Hamburg, Germany
[2] Gen Hosp St Georg, Hamburg, Germany
[3] Bakken Res Ctr, Maastricht, Netherlands
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2002年 / 25卷 / 04期
关键词
catheter ablation; ablation electrodes; electrophysiology;
D O I
10.1046/j.1460-9592.2002.00463.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Irrigation of radiofrequency current (RF) ablation reduces the risk of thrombus formation. The aim of this study was to investigate the impact of different irrigation catheter flow rates and contact pressures from the catheter on the development of lesion dimension and thrombus formation. A thigh muscle preparation was achieved in six sheep to create a cradle that was filled and perfused with heparinized blood (250 mL/min, 37Cdegrees). RF ablation (30 s, 30 W) was initially performed with three different irrigation flow rates (5 mL/min, 10 mL/min, and 20 mL/min) and a perpendicular position (0.1 N contact pressure) of the irrigated ablation catheter ("Sprinklr," Medtronic, Inc., Minneapolis, MN, USA). The next lesions were induced with constant contact pressure of 0.05 Newton (N); 0.1 N; 0.3 and 0.5 N and a parallel or perpendicular orientation of the catheter, respectively. A constant irrigation flow of 10 mL/min was maintained during these RF applications. Cross sections of the lesions were investigated with regard to maximal depth and maximal diameter at and below the surface. During high flow irrigation (20 mL/min) the surface diameter was significantly smaller (0.63 +/- 0.1 cm) compared to irrigation flow rates of 5 mL/min (0.88 +/- 0.2 cm) and 10 mL/min (1 +/- 0.1 cm). Thrombus formation was not observed during any RF application. Only in perpendicular catheter orientations with a contact pressure of 0.5 N were significantly deeper lesions (0.85 +/- 0.12 cm) induced compared to 0.05 N (0.55 +/- 0.02 cm), 0.1 N (0.7 +/- 0.01 cm) and 0.3 N (0.67 +/- 0.01 cm) contact pressure. There was no significant difference in lesion depth with different flow rates. Irrigated RF oblation even with low flow rates and high catheter contact pressure prevented thrombus formation at the electrode. Smaller lesion diameters have been created with high irrigation flow rates. The deeper lesion created with high catheter contact pressure might be caused by a greater power transmission to the tissue.
引用
收藏
页码:463 / 469
页数:7
相关论文
共 18 条
[1]   Popping phenomena in temperature-controlled radiofrequency ablation: When and why do they occur? [J].
Eick, OJ ;
Gerritse, B ;
Schumacher, B .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2000, 23 (02) :253-258
[2]   OBSERVATIONS ON ELECTRODE-TISSUE INTERFACE TEMPERATURE AND EFFECT ON ELECTRICAL-IMPEDANCE DURING RADIOFREQUENCY ABLATION OF VENTRICULAR MYOCARDIUM [J].
HAINES, DE ;
VEROW, AF .
CIRCULATION, 1990, 82 (03) :1034-1038
[3]   TISSUE HEATING DURING RADIOFREQUENCY CATHETER ABLATION - A THERMODYNAMIC MODEL AND OBSERVATIONS IN ISOLATED PERFUSED AND SUPERFUSED CANINE RIGHT VENTRICULAR FREE WALL [J].
HAINES, DE ;
WATSON, DD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (06) :962-976
[4]   COAGULATION OF VENTRICULAR MYOCARDIUM USING RADIOFREQUENCY ALTERNATING-CURRENT - BIO-PHYSICAL ASPECTS AND EXPERIMENTAL FINDINGS [J].
HAVERKAMP, W ;
HINDRICKS, G ;
GULKER, H ;
RISSEL, U ;
PFENNINGS, W ;
BORGGREFE, M ;
BREITHARDT, G .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (01) :187-195
[5]   RADIOFREQUENCY CURRENT DIRECTED ACROSS THE MITRAL ANULUS WITH A BIPOLAR EPICARDIAL-ENDOCARDIAL CATHETER ELECTRODE CONFIGURATION IN DOGS [J].
JACKMAN, WM ;
KUCK, KH ;
NACCARELLI, GV ;
CARMEN, L ;
PITHA, J .
CIRCULATION, 1988, 78 (05) :1288-1298
[6]   Successful irrigated-tip catheter ablation of atrial flutter resistant to conventional radiofrequency ablation [J].
Jaïs, P ;
Haïssaguerre, M ;
Shah, DC ;
Takahashi, A ;
Hocini, M ;
Lavergne, T ;
Lafitte, S ;
Le Mouroux, A ;
Fischer, B ;
Clémenty, J .
CIRCULATION, 1998, 98 (09) :835-838
[7]   Electrophysiological correlates of transmural linear ablation [J].
Liem, LB ;
Pomeranz, M ;
Riseling, K ;
Anderson, S ;
Berry, GJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2000, 23 (01) :40-46
[8]   Inverse relationship between electrode size and lesion size during radiofrequency ablation with active electrode cooling [J].
Nakagawa, H ;
Wittkampf, FHM ;
Yamanashi, WS ;
Pitha, JV ;
Imai, S ;
Campbell, B ;
Arruda, M ;
Lazzara, R ;
Jackman, WM .
CIRCULATION, 1998, 98 (05) :458-465
[9]   COMPARISON OF IN-VIVO TISSUE TEMPERATURE PROFILE AND LESION GEOMETRY FOR RADIOFREQUENCY ABLATION WITH A SALINE-IRRIGATED ELECTRODE VERSUS TEMPERATURE CONTROL IN A CANINE THIGH MUSCLE PREPARATION [J].
NAKAGAWA, H ;
YAMANASHI, WS ;
PITHA, JV ;
ARRUDA, M ;
WANG, XZ ;
OHTOMO, K ;
BECKMAN, KJ ;
MCCLELLAND, JH ;
LAZZARA, R ;
JACKMAN, WM .
CIRCULATION, 1995, 91 (08) :2264-2273
[10]   Lesion dimensions during temperature-controlled radiofrequency catheter ablation of left ventricular porcine myocardium - Impact of ablation site, electrode size, and convective cooling [J].
Petersen, HH ;
Chen, X ;
Pietersen, A ;
Svendsen, JH ;
Haunso, S .
CIRCULATION, 1999, 99 (02) :319-325