Fast GC-MS Pesticide Multiresidue Analysis of Apples

被引:26
作者
Huskova, Renta [1 ]
Matisova, Eva [1 ]
Kirchner, Michal [1 ]
机构
[1] Slovak Univ Technol Bratislava, Fac Chem & Food Technol, Inst Analyt Chem, Bratislava 81237, Slovakia
关键词
Gas chromatography; Fast GC-MS; Multiresidue analysis; Pesticide residues; Calibration;
D O I
10.1365/s10337-008-0562-1
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A fast gas chromatographic-mass spectrometric (GC-MS) method is proposed for pesticide multiresidue analysis of apples. The QuEChERS method was used for sample preparation. GC-MS analysis was performed with a PTV, an autoinjector, and a quadrupole benchtop MS detector. Electron-impact ionization (70 eV) was used with two modes of selected ion monitoring. Compounds were separated under temperature-programmed conditions on a narrow-bore diphenyldimethylsiloxane column. In one chromatographic run 61 pesticides of different chemical classes, and triphenyl phosphate as internal standard, were determined in 11 min. Calibration was performed with matrix-matched standard solutions and response to the pesticides was a linear function of concentration in the range 1-500 ng mL(-1) (equivalent to 1-500 mu g kg(-1) in real samples). High values of the determination coefficients (R-2; 0.9900-1.0000) were obtained for most of the pesticides. Limits of detection and quantification were determined. When the method was used for analysis of pesticide residues in real samples, five pesticides were detected at concentrations in the range 1.00-21.47 mu g kg(-1). Repeatability of measurements, expressed as relative standard deviations of absolute peak areas, normalized relative to TPP, and of the concentrations determined, met the EU criterion of RSD 20%. Use of the internal standard moderately improved quantitative results.
引用
收藏
页码:S49 / S55
页数:7
相关论文
共 30 条
[1]   PULSED FLAME PHOTOMETER DETECTOR FOR GAS-CHROMATOGRAPHY [J].
AMIRAV, A ;
JING, HW .
ANALYTICAL CHEMISTRY, 1995, 67 (18) :3305-3318
[2]   Evaluation of analyte protectants to improve gas chromatographic analysis of pesticides [J].
Anastassiades, M ;
Mastovská, K ;
Lehotay, SJ .
JOURNAL OF CHROMATOGRAPHY A, 2003, 1015 (1-2) :163-184
[3]  
Anastassiades M, 2003, J AOAC INT, V86, P412
[4]   Use of automated direct sample introduction with analyte protectants in the GC-MS analysis of pesticide residues [J].
Cajka, T ;
Mastovská, K ;
Lehotay, SJ ;
Hajslová, J .
JOURNAL OF SEPARATION SCIENCE, 2005, 28 (9-10) :1048-1060
[5]   Gas chromatography-high-resolution time-of-flight mass spectrometry in pesticide residue analysis:: advantages and limitations [J].
Cajka, T ;
Hajslová, J .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1058 (1-2) :251-261
[6]  
Dallüge J, 2002, J SEP SCI, V25, P608, DOI 10.1002/1615-9314(20020601)25:9&lt
[7]  
608::AID-JSSC608&gt
[8]  
3.0.CO
[9]  
2-
[10]  
Dömötörová M, 2005, ACTA CHIM SLOV, V52, P422