Analytical, Numerical and Experimental Study of a Horizontal Electrothermal MEMS Microgripper for the Deformability Characterisation of Human Red Blood Cells

被引:40
作者
Cauchi, Marija [1 ]
Grech, Ivan [2 ]
Mallia, Bertram [3 ]
Mollicone, Pierluigi [1 ]
Sammut, Nicholas [2 ]
机构
[1] Univ Malta, Fac Engn, Dept Mech Engn, MSD-2080 Msida, Malta
[2] Univ Malta, Fac Informat & Commun Technol, Dept Microelect & Nanoelect, MSD-2080 Msida, Malta
[3] Univ Malta, Fac Engn, Dept Met & Mat Engn, MSD-2080 Msida, Malta
关键词
MEMS microgripper; red blood cells; electrothermal actuation; analytical model; finite element model; fabrication; PolyMUMPs (TM); temperature; displacement; stresses; VISCOELASTIC CHARACTERIZATION; DRIVEN MICROGRIPPER;
D O I
10.3390/mi9030108
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Microgrippers are typical microelectromechanical systems (MEMS) that are widely used for micromanipulation and microassembly in both biological and micromanufacturing fields. This paper presents the design, modelling, fabrication and experimental testing of an electrothermal microgripper based on a 'hot and cold arm' actuator design that is suitable for the deformability characterisation of human red blood cells (RBCs). The analysis of the mechanical properties of human RBCs is of great interest in the field of medicine as pathological alterations in the deformability characteristics of RBCs have been linked to a number of diseases. The study of the microgripper's steady-state performance is initially carried out by the development of a lumped analytical model, followed by a numerical model established in CoventorWare (R) (Coventor, Inc., Cary, NC, USA) using multiphysics finite element analysis. Both analytical and numerical models are based on an electothermomechanical analysis, and take into account the internal heat generation due to the applied potential, as well as conduction heat losses through both the anchor pads and the air gap to the substrate. The models are used to investigate key factors of the actuator's performance including temperature distribution, deflection and stresses based on an elastic analysis of structures. Results show that analytical and numerical values for temperature and deflection are in good agreement. The analytical and computational models are then validated experimentally using a polysilicon microgripper fabricated by the standard surface micromachining process, PolyMUMPs (TM)(Durham, NC, USA). The microgripper's actuation is characterised at atmospheric pressure by optical microscopy studies. Experimental results for the deflection of the microgripper arm tips are found to be in good agreement with the analytical and numerical results, with process-induced variations and the non-linear temperature dependence of the material properties accounting for the slight discrepancies observed. The microgripper is shown to actuate to a maximum opening displacement of 9 mu m at an applied voltage of 3 V, thus being in line with the design requirement of an approximate opening of 8 mu m for securing and characterising a RBC.
引用
收藏
页数:18
相关论文
共 30 条
[1]   Fabrication of Novel MEMS Microgrippers by Deep Reactive Ion Etching With Metal Hard Mask [J].
Bagolini, Alvise ;
Ronchin, Sabina ;
Bellutti, Pierluigi ;
Chiste, Matteo ;
Verotti, Matteo ;
Belfiore, Nicola Pio .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2017, 26 (04) :926-934
[2]  
Cauchi M, 2016, P INT CAE C 2016 PAR, P1
[3]   Electrothermally activated SU-8 microgripper for single cell manipulation in solution [J].
Chronis, N ;
Lee, LP .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2005, 14 (04) :857-863
[4]  
Comtois J. H., 1996, Technical Digest. Solid-State Sensor and Actuator Workshop, P174
[5]   Thermal Management Using MEMS Bimorph Cantilever Beams [J].
Coutu, R. A., Jr. ;
LaFleur, R. S. ;
Walton, J. P. K. ;
Starman, L. A. .
EXPERIMENTAL MECHANICS, 2016, 56 (07) :1293-1303
[6]  
Cowen A., 2011, POLYMUMPS DESIGN HDB
[7]   New MEMS Tweezers for the Viscoelastic Characterization of Soft Materials at the Microscale [J].
Di Giamberardino, Paolo ;
Bagolini, Alvise ;
Bellutti, Pierluigi ;
Rudas, Imre J. ;
Verotti, Matteo ;
Botta, Fabio ;
Belfiore, Nicola P. .
MICROMACHINES, 2018, 9 (01)
[8]   A Comprehensive Survey on Microgrippers Design: Operational Strategy [J].
Dochshanov, Alden ;
Verotti, Matteo ;
Belfiore, Nicola P. .
JOURNAL OF MECHANICAL DESIGN, 2017, 139 (07)
[9]  
Guckel H., 1992, Technical Digest. IEEE Solid-State Sensor and Actuator Workshop (Cat. No.92TH0403-X), P73, DOI 10.1109/SOLSEN.1992.228273
[10]  
Hannon Bruce, 2009, P3