A smartphone-based optical platform for colorimetric analysis of microfluidic device

被引:97
作者
Kim, Sang C. [1 ]
Jalal, Uddin M. [1 ]
Im, Sung B. [1 ]
Ko, Sungho [2 ]
Shim, Joon S. [1 ]
机构
[1] Kwangwoon Univ, Dept Elect Convergence Engn, Bio IT Convergence Lab, Seoul, South Korea
[2] CHA Univ, Dept Biotechnol, Seongnam, Gyeonggi Do, South Korea
基金
新加坡国家研究基金会;
关键词
Point-of-Care (POC) testing; Microfluidic; Smartphone; Light diffuser; Gray-scale-valuation (GSV); Hematocrit; OF-CARE DIAGNOSTICS; HEALTH; TECHNOLOGIES; HEMATOCRIT; BLOOD; CHIP; LAB;
D O I
10.1016/j.snb.2016.07.159
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this work, a smartphone based optical platform for colorimetric analysis of blood hematocrit using a disposable microfluidic device is designed, implemented and fully characterized. Using an integrated camera in the smartphone, pictures of human blood in the microchannel were taken and analyzed by a mobile application. To avoid the image burning and ambient light effect, a unique light-diffusing model inside a white acrylic-imaging box was included in this platform. With the image-processing program on the smartphone, the developed device was successfully applied to determine various hematocrit levels of human blood from 10% to 65%. Furthermore, the characterization of the depth of the microfluidic channel demonstrated that a shallower depth of the microchannel enhanced the sensitivity of the hematocrit determination. The limit of detection (LOD) obtained from the developed platform was 0.1% of hematocrit with a sensitivity of 0.53 GSV (a.u.)/hematocrit%. Thus, utilizing the advantage of the microfluidic effect, a rapid and sensitive hematocrit determination was achieved successfully. With this, the hematocrit of human blood could be conveniently and accurately determined using the disposable microfluidic device, and then processed by the smartphone camera and mobile image analysis application. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:52 / 59
页数:8
相关论文
共 44 条
[1]  
ARMBRUSTER DA, 2008, CLIN BIOCHEM REV, V29, DOI DOI 10.3390/S20236922
[2]   Portable smartphone quantitation of prostate specific antigen (PSA) in a fluoropolymer microfluidic device [J].
Barbosa, Ana I. ;
Gehlot, Poonam ;
Sidapra, Kalpita ;
Edwards, Alexander D. ;
Reis, Nuno M. .
BIOSENSORS & BIOELECTRONICS, 2015, 70 :5-14
[3]   Cellphone-Based Hand-Held Microplate Reader for Point-of-Care Testing of Enzyme-Linked Immunosorbent Assays [J].
Berg, Brandon ;
Cortazar, Bingen ;
Tseng, Derek ;
Ozkan, Haydar ;
Feng, Steve ;
Wei, Qingshan ;
Chan, Raymond Yan-Lok ;
Burbano, Jordi ;
Farooqui, Qamar ;
Lewinski, Michael ;
Di Carlo, Dino ;
Garner, Omai B. ;
Ozcan, Aydogan .
ACS NANO, 2015, 9 (08) :7857-7866
[4]   Mobile Phone Based Clinical Microscopy for Global Health Applications [J].
Breslauer, David N. ;
Maamari, Robi N. ;
Switz, Neil A. ;
Lam, Wilbur A. ;
Fletcher, Daniel A. .
PLOS ONE, 2009, 4 (07)
[5]  
Chem R., 2011, BIOMICROFLUIDICS, V8
[6]  
Efthimia Bilissi, 2008, LANGFORDS ADV PHOTOG
[7]   MULTIPLE-SCATTERING EFFECTS IN TRANSMISSION OXIMETRY [J].
FINE, I ;
WEINREB, A .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1993, 31 (05) :516-522
[8]   Label-free biodetection using a smartphone [J].
Gallegos, Dustin ;
Long, Kenneth D. ;
Yu, Hojeong ;
Clark, Peter P. ;
Lin, Yixiao ;
George, Sherine ;
Nath, Pabitra ;
Cunningham, Brian T. .
LAB ON A CHIP, 2013, 13 (11) :2124-2132
[9]  
Ginsberg Barry H, 2009, J Diabetes Sci Technol, V3, P903
[10]  
Guider Romain, 2015, Sensing and Bio-Sensing Research, V6, P99, DOI 10.1016/j.sbsr.2015.08.002