In vitro wear assessments of fixed and mobile UHMWPE total knee replacement

被引:10
作者
Affatato, Saverio [1 ]
Bracco, Pierangiola [2 ,3 ]
Sudanese, Alessandra [1 ,4 ]
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, I-40136 Bologna, Italy
[2] Univ Turin, Dept Chem, I-10125 Turin, Italy
[3] Univ Turin, NIS Ctr Excellence, I-10125 Turin, Italy
[4] Ist Ortoped Rizzoli, I-40136 Bologna, Italy
关键词
Wear; Mobile TKR; Fixed TKR; FTIR analyses; Crystallinity; MOLECULAR-WEIGHT POLYETHYLENE; FOLLOW-UP; BEARING; ARTHROPLASTY; PROSTHESES; OXIDATION; STERILIZATION; IRRADIATION;
D O I
10.1016/j.matdes.2012.11.046
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This work discusses the wear behaviour of two different ultra-high-molecular-weight-polyethylene tibial component designs. Mobile and fixed bearings were tested on a knee wear simulator for 5 million cycles using bovine calf serum as lubricant. We correlated the wear results with the chemical characterisation of the investigated materials: Fourier Transformed Infra Red Spectroscopy analyses, Differential Scanning Calorimetry and cross-link density measurements were used to assess the chemical features of this polyethylene. Mobile and fixed polyethylene inserts showed a different wear behaviour: the mobile designs components showed lower weight losses than the fixed components (109 +/- 6 mg and 163 +/- 80 mg, respectively). Significant statistical differences were observed in wear rate (P = 0.035, Kolmogorov-Smirnov Test for two samples). From a molecular point of view, typical radiation-induced oxidation profiles were observed in all the tested polyethylene samples, but the overall degradation was more significant in the fixed bearing inserts and this is likely to play a role on the wear performances. (C) 2012 Elsevier Ltd. All rights reserved.
引用
收藏
页码:44 / 51
页数:8
相关论文
共 51 条
[1]   Unicompartmental knee prostheses:: in vitro wear assessment of the menisci tibial insert after two different fixation methods [J].
Affatato, S. ;
Spinelli, M. ;
Zavalloni, M. ;
Carmignato, S. ;
Lopomo, N. ;
Marcacci, M. ;
Viceconti, M. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (19) :5357-5369
[2]   Can the method of fixation influence the wear behaviour of ZrN coated unicompartmental mobile knee prostheses? [J].
Affatato, S. ;
Spinelli, M. ;
Lopomo, N. ;
Grupp, T. M. ;
Marcacci, M. ;
Toni, A. .
CLINICAL BIOMECHANICS, 2011, 26 (02) :152-158
[3]   Fluid absorption study in ultra-high molecular weight polyethylene( UHMWPE) sterilized and unsterilized acetabular cups [J].
Affatato, S ;
Vandelli, C ;
Bordini, B ;
Toni, A .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2001, 215 (H1) :107-111
[4]   Investigation on wear of knee prostheses under fixed kinematic conditions [J].
Affatato, Saverio ;
Leardini, Walther ;
Rocchi, Mirko ;
Toni, Aldo ;
Viceconti, Marco .
ARTIFICIAL ORGANS, 2008, 32 (01) :13-18
[5]   In vitro wear performance of standard, crosslinked, and vitamin-E-blended UHMWPE [J].
Affatato, Saverio ;
Bracco, Pierangiola ;
Costa, Luigi ;
Villa, Tomaso ;
Quaglini, Virginio ;
Toni, Aldo .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART A, 2012, 100A (03) :554-560
[6]   Comparison of mobile-bearing and fixed-bearing total knee arthroplasty - A prospective randomized study [J].
Aglietti, P ;
Baldini, A ;
Buzzi, R ;
Lup, D ;
De Luca, L .
JOURNAL OF ARTHROPLASTY, 2005, 20 (02) :145-153
[7]   Wear and toughness of crosslinked polyethylene for total knee replacements: A study using a simulator and small-punch testing [J].
Akagi, Masao ;
Asano, Taiyo ;
Clarke, Ian C. ;
Niiyama, Norio ;
Kyomoto, Masayuki ;
Nakamura, Takashi ;
Hamanishi, Chiaki .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2006, 24 (10) :2021-2027
[8]  
Armitage P, 1994, KOLMOGOROV SMIRNOV T, P397
[9]   Fixed Versus Rotating Platform Total Knee Arthroplasty: A Prospective, Randomized, Single-Blind Study [J].
Ball, Scott T. ;
Sanchez, Hugo B. ;
Mahoney, Ormonde M. ;
Schmalzried, Thomas P. .
JOURNAL OF ARTHROPLASTY, 2011, 26 (04) :531-536
[10]   Investigation of wear of knee prostheses in a new displacement/force-controlled simulator [J].
Barnett, PI ;
McEwen, HMJ ;
Auger, DD ;
Stone, MH ;
Ingham, E ;
Fisher, J .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2002, 216 (H1) :51-61