Hip endoprosthesis for in vivo measurement of joint force and temperature

被引:75
作者
Graichen, F [1 ]
Bergmann, G [1 ]
Rohlmann, A [1 ]
机构
[1] Free Univ Berlin, Orthopaed Hosp, Biomech Lab, D-14195 Berlin, Germany
关键词
telemetry; hip joint force; hip joint temperature; in vivo measurement; biomechanics; instrumented hip prosthesis;
D O I
10.1016/S0021-9290(99)00110-4
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Friction between the prosthetic head and acetabular cup increases the temperature in hip implants during activities like walking. A hip endoprosthesis was instrumented with sensors to measure the joint contact forces and the temperature distribution along the entire length of the titanium implant. Sensors and two inductively powered telemetry units are placed inside the hip implant and hermetically sealed against body fluids. Each telemetry unit contains an integrated 8-channel telemetry chip and a radio frequency transmitter. Force, temperature and power supply data are transmitted at different frequencies by two antennas to an external twin receiver. The inductive power supply is controlled by a personal computer. Force and temperature are monitored in real time and all data are stored on a video tape together with the patient's images. This paper describes the design and accuracy of the instrumented implant and the principal function of the external system components. (C) 1999 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:1113 / 1117
页数:5
相关论文
共 11 条
[1]   HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :969-990
[2]   INFLUENCE OF SHOES AND HEEL STRIKE ON THE LOADING OF THE HIP-JOINT [J].
BERGMANN, G ;
KNIGGENDORF, H ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1995, 28 (07) :817-827
[3]   MULTICHANNEL STRAIN-GAUGE TELEMETRY FOR ORTHOPEDIC IMPLANTS [J].
BERGMANN, G ;
GRAICHEN, F ;
SIRAKY, J ;
JENDRZYNSKI, H ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1988, 21 (02) :169-176
[4]   IS STAIRCASE WALKING A RISK FOR THE FIXATION OF HIP IMPLANTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1995, 28 (05) :535-553
[5]  
Bergmann G, 1990, IMPLANTABLE TELEMETR, P35
[6]  
BERGMANN G, 1991, 37 ANN M ORTH RES SO
[7]  
BERGMANN G, 1997, IN VIVO MESSUNG BELA
[8]   Patient monitoring system for load measurement with spinal fixation devices [J].
Graichen, F ;
Bergmann, G ;
Rohlmann, A .
MEDICAL ENGINEERING & PHYSICS, 1996, 18 (02) :167-174
[9]   4-CHANNEL TELEMETRY SYSTEM FOR INVIVO MEASUREMENT OF HIP-JOINT FORCES [J].
GRAICHEN, F ;
BERGMANN, G .
JOURNAL OF BIOMEDICAL ENGINEERING, 1991, 13 (05) :370-375
[10]  
GRAICHEN F, 1990, IMPLANTIERBARES TELE