Wear at the taper-trunnion junction of contemporary ceramic-on-ceramic hips shown in a multistation hip simulator

被引:14
|
作者
Bhalekar, Rohan M. [1 ]
Smith, Simon L. [1 ]
Joyce, Thomas J. [1 ]
机构
[1] Newcastle Univ, Sch Engn, Newcastle Upon Tyne NE1 7RU, Tyne & Wear, England
关键词
total hip replacement; ceramic-on-ceramic; hip simulator; taper-trunnion junction; wear; TITANIUM-ALLOY; IN-VITRO; METAL; FAILURE; HEAD; PROSTHESES; UHMWPE; DEBRIS; MICROSEPARATION; BEARINGS;
D O I
10.1002/jbm.b.34213
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Ceramic-on-ceramic (CoC) total hip replacements (THR) have a substantially lower wear rate than metal-on-polyethylene (MoP) hips, as shown by hip simulator testing. However, the revision rates of CoC and MoP hips are comparable. To try and explain this discrepancy the wear, at both the bearing surfaces and taper-trunnion interface of 36 mm BIOLOXdelta CoC THRs, mounted on 12/14 titanium (Ti6Al4V) trunnions was investigated using a hip simulator (n = 3) and a dynamically loaded CoC sample in a separate test station. Wear was assessed gravimetrically and surface roughness measurements of the articulating and taper surfaces taken at regular intervals. Scanning electron microscopy, surface roughness, and gravimetric measurements of trunnions was performed. After 5 million cycles, the mean total wear from the ceramic articulating surfaces was 0.25 mm(3) from the hip simulator test, and that from the titanium trunnions was 0.29 mm(3). This metal wear may provide an explanation for adverse reaction to metal debris found in contemporary CoC hip joints. It is therefore vital to consider taper-trunnion wear in preclinical testing of artificial hip joints. (c) 2018 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 107B: 1199-1209, 2019.
引用
收藏
页码:1199 / 1209
页数:11
相关论文
共 35 条
  • [11] Concomitant evolution of wear and squeaking in dual-severity, lubricated wear testing of ceramic-on-ceramic hip prostheses
    Sanders, Anthony
    Tibbitts, Ira
    Brannon, Rebecca
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2012, 30 (09) : 1377 - 1383
  • [12] Incidence of Ceramic Fracture in Contemporary Ceramic-on-Ceramic Total Hip Arthroplasty: A Meta-analysis of Proportions
    Yoon, Byung-Ho
    Park, Jung-Wee
    Cha, Yong-Han
    Won, Suk-Hyung
    Lee, Young-Kyun
    Ha, Yong-Chan
    Koo, Kyung-Hoi
    JOURNAL OF ARTHROPLASTY, 2020, 35 (05) : 1437 - +
  • [13] Precision and accuracy in ceramic-on-ceramic wear analyses: influence of simulator test duration
    Richardson, HA
    Clarke, IC
    Williams, P
    Donaldson, T
    Oonishi, H
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2005, 219 (H6) : 401 - 405
  • [14] Combined wear and ageing of ceramic-on-ceramic bearings in total hip replacement under edge loading conditions
    Al-Hajjar, Mazen
    Gremillard, Laurent
    Begand, Sabine
    Oberbach, Thomas
    Hans, Karen
    Delfosse, Daniel
    Chevalier, Jerome
    Jennings, Louise M.
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2019, 98 : 40 - 47
  • [15] Optimal wear performance of ceramic-on-ceramic bearings for total hip replacement
    Clarke, IC
    Oonishi, H
    JOINT ARTHROPLASTY, 1999, : 140 - 151
  • [16] Improved friction and wear performance of micro dimpled ceramic-on-ceramic interface for hip joint arthroplasty
    Roy, Taposh
    Choudhury, Dipankar
    Ghosh, Subir
    Bin Maniat, Azuddin
    Pingguan-Murphy, Belinda
    CERAMICS INTERNATIONAL, 2015, 41 (01) : 681 - 690
  • [17] Severe Wear and multiple Pseudotumor formation due to revision for ceramic head breakage after ceramic-on-ceramic Total hip arthroplasty: a case report
    Xing, Dan
    Yang, Chaolei
    Li, Rujun
    Hou, Yunfei
    Kou, Bolong
    Li, Hu
    Lin, Jianhao
    BMC MUSCULOSKELETAL DISORDERS, 2019, 20 (1)
  • [18] Characterisation of wear debris from UHMWPE on zirconia ceramic, metal-on-metal and alumina ceramic-on-ceramic hip prostheses generated in a physiological anatomical hip joint simulator
    Tipper, JL
    Firkins, PJ
    Besong, AA
    Barbour, PSM
    Nevelos, J
    Stone, MH
    Ingham, E
    Fisher, J
    WEAR, 2001, 250 : 120 - 128
  • [19] The incidence of noise with a new ceramic-on-ceramic bearing in hip replacement : a prospective multicenter analysis of 142 hips
    Ilchmann, Thomas
    Jovanovic, Milan
    Ranke, Thomas-Peter
    Clauss, Martin
    ACTA ORTHOPAEDICA BELGICA, 2021, 87 (ES1): : 8 - 13
  • [20] The wear of the taper junction of the stem trunnion with the femoral head of a total modular hip prosthesis
    Capitanu, Lucian
    Badita, Liliana-Laura
    Tiganesteanu, Constantin
    Florescu, Virgil
    JURNAL TRIBOLOGI, 2019, 23 : 61 - 75