A portable multi-sensor system for non-invasive measurement of biometrical data

被引:3
作者
Depari, A. [1 ]
Flammini, A. [1 ]
Rinaldi, S. [1 ]
Vezzolia, A. [1 ]
机构
[1] Univ Brescia, Dept Informat Engn, I-25123 Brescia, Italy
来源
26TH EUROPEAN CONFERENCE ON SOLID-STATE TRANSDUCERS, EUROSENSOR 2012 | 2012年 / 47卷
关键词
Photoplethysmographic sensor; body impedance measurement; non-invasive measurement;
D O I
10.1016/j.proeng.2012.09.399
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In the last years, the availability of performing personal computing devices, like smartphones, opens new scenarios to the so called Body Area Network, i.e., sensor networks used to monitor physical conditions of the user. Generally speaking, the sensors adopted to monitor physical parameters, like heartbeat or exertion level, are not suitable for run-time measurements. However, the physical performance of athletes may be improved by a continuous monitoring of the body conditions. In this work, a simple, portable and low-cost system for non-invasive measurement of physical parameters is proposed and experimentally characterized. It is based on two different sensors combined in a unique structure to be applied in a clip-like fashion to the earlobe or analogue body part of a person. A photoplethysmographic sensor is used to optically measure the change in volume of blood in the arteries, thus to acquire information about heart beating, blood pressure, and arrhythmias. The second sensor is composed of two electrodes, thus an impedance measurement of the tissue is conducted. The impedance estimation can be used to monitor application-specific parameters, such as blood pH. The structure is compact in size and low-power, making it suitable for portable and battery-operated systems. The experimental characterization of the prototype demonstrates the feasibility of the proposed approach. (C) 2012 Elsevier Ltd....Selection and/or peer-review under responsibility of the Symposium Cracoviense Sp. z.o.o.
引用
收藏
页码:1323 / 1326
页数:4
相关论文
共 8 条
[1]  
[Anonymous], P INSTR MEAS TECHN C
[2]  
[Anonymous], MED INSTRUMENTATION
[3]  
De Dominicis C. M., 2011, P IEEE SENS APPL S S, P120
[4]   LED-TRANSISTOR PHOTOPLETHYSMOGRAPH [J].
LEE, AL ;
TAHMOUSH, AJ ;
JENNINGS, JR .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1975, BM22 (03) :248-250
[5]   NONINVASIVE PULSE OXIMETRY UTILIZING SKIN REFLECTANCE PHOTOPLETHYSMOGRAPHY [J].
MENDELSON, Y ;
OCHS, BD .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (10) :798-805
[6]   Identification of high-risk acute coronary syndromes by spectral analysis of ear photoplethysmographic waveform variability [J].
Middleton, Paul M. ;
Chan, Gregory S. H. ;
Marr, Sebastian ;
Celler, Branko G. ;
Lovell, Nigel H. .
PHYSIOLOGICAL MEASUREMENT, 2011, 32 (08) :1181-1192
[7]  
Miklaveie D., 2006, WILEY ENCY BIOMEDICA, P1, DOI DOI 10.1002/9780471740360.EBS0403
[8]  
Vishay Semiconductor, TCRT1000 DAT SHEET