Quantitative evaluation of analyte transport on microfluidic paper-based analytical devices (μPADs)

被引:36
作者
Ota, Riki [1 ]
Yamada, Kentaro [1 ]
Suzuki, Koji [1 ]
Citterio, Daniel [1 ]
机构
[1] Keio Univ, Dept Appl Chem, Kohoku Ku, 3-14-1 Hiyoshi, Yokohama, Kanagawa 2238522, Japan
基金
日本学术振兴会;
关键词
DIRECT COLORIMETRIC ASSAY; CAPILLARY-FLOW; LOW-COST; POROUS-MEDIA; WASTE-WATER; NITRO-PAPS; METAL-IONS; ADSORPTION; CELLULOSE; LACTOFERRIN;
D O I
10.1039/c7an01702b
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The transport efficiency during capillary flow-driven sample transport on microfluidic paper-based analytical devices (mu PADs) made from filter paper has been investigated for a selection of model analytes (Ni2+, Zn2+, Cu2+, PO43-, bovine serum albumin, sulforhodamine B, amaranth) representing metal cations, complex anions, proteins and anionic molecules. For the first time, the transport of the analytical target compounds rather than the sample liquid, has been quantitatively evaluated by means of colorimetry and absorption spectrometry-based methods. The experiments have revealed that small paperfluidic channel dimensions, additional user operation steps (e.g. control of sample volume, sample dilution, washing step) as well as the introduction of sample liquid wicking areas allow to increase analyte transport efficiency. It is also shown that the interaction of analytes with the negatively charged cellulosic paper substrate surface is strongly influenced by the physico-chemical properties of the model analyte and can in some cases (Cu2+) result in nearly complete analyte depletion during sample transport. The quantitative information gained through these experiments is expected to contribute to the development of more sensitive mu PADs.
引用
收藏
页码:643 / 653
页数:11
相关论文
共 58 条
[21]   Biomolecule immobilization techniques for bioactive paper fabrication [J].
Kong, Fanzhi ;
Hu, Yim Fun .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2012, 403 (01) :7-13
[22]   Paper-Based Analytical Device for Zinc Ion Quantification in Water Samples with Power-Free Analyte Concentration [J].
Kudo, Hiroko ;
Yamada, Kentaro ;
Watanabe, Daiki ;
Suzuki, Koji ;
Citterio, Daniel .
MICROMACHINES, 2017, 8 (04)
[23]   Raman Characterization of Nanoparticle Transport in Microfluidic Paper-Based Analytical Devices (μPADs) [J].
Lahr, Rebecca Halvorson ;
Wallace, Grant C. ;
Vikesland, Peter J. .
ACS APPLIED MATERIALS & INTERFACES, 2015, 7 (17) :9139-9146
[24]   Detection of heavy metal by paper-based microfluidics [J].
Lin, Yang ;
Gritsenko, Dmitry ;
Feng, Shaolong ;
Teh, Yi Chen ;
Lu, Xiaonan ;
Xu, Jie .
BIOSENSORS & BIOELECTRONICS, 2016, 83 :256-266
[25]   Paper-based sensors and assays: a success of the engineering design and the convergence of knowledge areas [J].
Lopez-Marzo, Adaris M. ;
Merkoci, Arben .
LAB ON A CHIP, 2016, 16 (17) :3150-3176
[26]   Rapid prototyping of paper-based microfluidics with wax for low-cost, portable bioassay [J].
Lu, Yao ;
Shi, Weiwei ;
Jiang, Lei ;
Qin, Jianhua ;
Lin, Bingcheng .
ELECTROPHORESIS, 2009, 30 (09) :1497-1500
[27]   A SENSITIVE, DIRECT COLORIMETRIC ASSAY OF SERUM ZINC USING NITRO-PAPS AND MICROWELL PLATES [J].
MAKINO, T .
CLINICA CHIMICA ACTA, 1991, 197 (03) :209-220
[28]   A SENSITIVE, DIRECT COLORIMETRIC ASSAY OF SERUM IRON USING THE CHROMOGEN, NITRO-PAPS [J].
MAKINO, T ;
KIYONAGA, M ;
KINA, K .
CLINICA CHIMICA ACTA, 1988, 171 (01) :19-28
[29]   Patterned paper as a platform for inexpensive, low-volume, portable bioassays [J].
Martinez, Andres W. ;
Phillips, Scott T. ;
Butte, Manish J. ;
Whitesides, George M. .
ANGEWANDTE CHEMIE-INTERNATIONAL EDITION, 2007, 46 (08) :1318-1320
[30]   Diagnostics for the Developing World: Microfluidic Paper-Based Analytical Devices [J].
Martinez, Andres W. ;
Phillips, Scott T. ;
Whitesides, George M. ;
Carrilho, Emanuel .
ANALYTICAL CHEMISTRY, 2010, 82 (01) :3-10