Spatial Resolution and Sensitivity of the Inveon Small-Animal PET Scanner

被引:161
|
作者
Visser, Eric P. [1 ]
Disselhorst, Jonathan A. [1 ]
Brom, Maarten [1 ]
Laverman, Peter [1 ]
Gotthardt, Martin [1 ]
Oyen, Wim J. G. [1 ]
Boerman, Otto C. [1 ]
机构
[1] Radboud Univ Nijmegen, Med Ctr, Dept Nucl Med, NL-6500 HB Nijmegen, Netherlands
关键词
positron emission tomography; small-animal; spatial resolution; sensitivity; image reconstruction; PERFORMANCE EVALUATION;
D O I
10.2967/jnumed.108.055152
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The Inveon small-animal PET scanner is characterized by a large, 127-mm axial length and a 161-mm crystal ring diameter. The associated high sensitivity is obtained by using all lines of response (LORs) up to the maximum ring difference (MRD) of 79, for which the most oblique LORs form acceptance angles of 38.3 degrees with transaxial planes. The result is 2 phenomena that are normally not encountered in PET scanners: a parallax or depth-of-interaction effect in the axial direction and the breakdown of Fourier rebinning (FORE). Both effects cause a deterioration of axial spatial resolution. Limiting the MRD to smaller values reduces this axial blurring at the cost of sensitivity. Alternatively, 3-dimensional (3D) reconstruction techniques can be used in which the rebinning step is absent. The aim of this study was to experimentally determine the spatial resolution and sensitivity of the Inveon for its whole field of view (FOV). Methods: Spatial resolution and sensitivity were measured using filtered backprojection (FBP) with FORE, FBP with LOR angle-weighted adapted FORE (AFORE), and 3D ordered-subset expectation maximization followed by maximum a posteriori reconstruction (OSEM3D/MAP). Results: Tangential and radial full width at half maximum (FWHM) showed almost no dependence on the MRD using FORE and FBP. Tangential FWHMs were 1.5 mm in the center of the FOV (CFOV) and 1.8 mm at the edge of the FOV (EFOV). Radial FWHMs were 1.5 and 3.0 mm in the CFOV and EFOV, respectively. In contrast, axial FWHMs increased with the MRD and ranged between 1.1 and 2.0 mm in the CFOV and between 1.5 and 2.7 mm in the EFOV for a MRD between 1 and 79. AFORE improved the axial resolution for a large part of the FOV, but image noise increased. OSEM3D/MAP yielded uniform spatial resolution in all directions, with an average FWHM of 1.65 +/- 0.06 mm. Sensitivity in the CFOV for the default energy and coincidence time window was 0.068; peak sensitivity was 0.111. Conclusion: The Inveon showed high spatial resolution and high sensitivity, both of which can be maintained using OSEM3D/MAP reconstruction instead of rebinning and 2D algorithms.
引用
收藏
页码:139 / 147
页数:9
相关论文
共 50 条
  • [41] A high spatial resolution CT scanner for small animal imaging
    Cicalini, E
    Baldazzi, G
    Belcari, N
    Del Guerra, A
    Gombia, M
    Motta, A
    Panetta, D
    NUCLEAR PHYSICS B-PROCEEDINGS SUPPLEMENTS, 2006, 150 : 313 - 316
  • [42] Characterization and optimization of image quality as a function of reconstruction algorithms and parameter settings in a Siemens Inveon small-animal PET scanner using the NEMA NU 4-2008 standards
    Visser, Eric P.
    Disselhorst, Jonathan A.
    van Lier, Monique G. J. T. B.
    Laverman, Peter
    de Jong, Gabie M.
    Oyen, Wim J. G.
    Boerman, Otto C.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 629 (01): : 357 - 367
  • [43] The ClearPET™:: A high resolution high sensitivity dual-layer phoswich small animal PET scanner
    Heinrichs, U
    Bruyndonckx, P
    Choi, Y
    Dietzel, G
    Korjik, M
    Lecoq, P
    Morel, C
    Pedrini, C
    Petrosyan, AG
    Pietrzyk, U
    Sappey-Marinier, D
    Streun, M
    Tavernier, S
    Ziemons, K
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2004, 31 : S400 - S400
  • [44] DigiPET: sub-millimeter spatial resolution small-animal PET imaging using thin monolithic scintillators
    Espana, Samuel
    Marcinkowski, Radoslaw
    Keereman, Vincent
    Vandenberghe, Stefaan
    Van Holen, Roel
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (13): : 3405 - 3420
  • [45] Achieving 0.4-mm FWHM Spatial Resolution with an RPC-Based Small-Animal PET Prototype
    Martins, Paulo
    Blanco, Alberto
    Crespo, Paulo
    Fatima Ferreira Marques, M.
    Marques, Rui Ferreira
    Gordo, Paulo M.
    Kajetanowicz, Marcin
    Korcyl, Grzegorz
    Lopes, Luis
    Michel, Jan
    Palka, Marek
    Traxler, Michael
    Fonte, Paulo
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [46] A sub-millimeter resolution detector module for small-animal PET applications
    Sacco, I.
    Dohle, R.
    Fischer, P.
    Gola, A.
    Piemonte, C.
    Ritzert, M.
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [47] Performance of a PET Insert for High-Resolution Small-Animal PET/MRI at 7 Tesla
    Stortz, Greg
    Thiessen, Jonathan D.
    Bishop, Daryl
    Khan, Muhammad Salman
    Kozlowski, Piotr
    Retiere, Fabrice
    Schellenberg, Graham
    Shams, Ehsan
    Zhang, Xuezhu
    Thompson, Christopher J.
    Goertzen, Andrew L.
    Sossi, Vesna
    JOURNAL OF NUCLEAR MEDICINE, 2018, 59 (03) : 536 - 542
  • [48] Effect of amphetamine on mouse striatal binding of [11C]Raclopride using the Inveon small animal PET scanner
    Arlott, Jessica
    Tang, Sac-Pham
    Wells, Lisa
    Huiban, Mickael
    Keat, Nicholas
    Salinas, Cristian
    Rabiner, Eugenii
    Ashworth, Sharon
    NEUROIMAGE, 2010, 52 : S184 - S184
  • [49] Performance characteristics of the Inveon micro-CT scanner in small animal imaging
    Khalil, Magdy Mohamed
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2018, 4 (01):
  • [50] Performance evaluation of an Inveon PET preclinical scanner
    Constantinescu, Cristian C.
    Mukherjee, Jogeshwar
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09): : 2885 - 2899