Spatial Resolution and Sensitivity of the Inveon Small-Animal PET Scanner

被引:161
|
作者
Visser, Eric P. [1 ]
Disselhorst, Jonathan A. [1 ]
Brom, Maarten [1 ]
Laverman, Peter [1 ]
Gotthardt, Martin [1 ]
Oyen, Wim J. G. [1 ]
Boerman, Otto C. [1 ]
机构
[1] Radboud Univ Nijmegen, Med Ctr, Dept Nucl Med, NL-6500 HB Nijmegen, Netherlands
关键词
positron emission tomography; small-animal; spatial resolution; sensitivity; image reconstruction; PERFORMANCE EVALUATION;
D O I
10.2967/jnumed.108.055152
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The Inveon small-animal PET scanner is characterized by a large, 127-mm axial length and a 161-mm crystal ring diameter. The associated high sensitivity is obtained by using all lines of response (LORs) up to the maximum ring difference (MRD) of 79, for which the most oblique LORs form acceptance angles of 38.3 degrees with transaxial planes. The result is 2 phenomena that are normally not encountered in PET scanners: a parallax or depth-of-interaction effect in the axial direction and the breakdown of Fourier rebinning (FORE). Both effects cause a deterioration of axial spatial resolution. Limiting the MRD to smaller values reduces this axial blurring at the cost of sensitivity. Alternatively, 3-dimensional (3D) reconstruction techniques can be used in which the rebinning step is absent. The aim of this study was to experimentally determine the spatial resolution and sensitivity of the Inveon for its whole field of view (FOV). Methods: Spatial resolution and sensitivity were measured using filtered backprojection (FBP) with FORE, FBP with LOR angle-weighted adapted FORE (AFORE), and 3D ordered-subset expectation maximization followed by maximum a posteriori reconstruction (OSEM3D/MAP). Results: Tangential and radial full width at half maximum (FWHM) showed almost no dependence on the MRD using FORE and FBP. Tangential FWHMs were 1.5 mm in the center of the FOV (CFOV) and 1.8 mm at the edge of the FOV (EFOV). Radial FWHMs were 1.5 and 3.0 mm in the CFOV and EFOV, respectively. In contrast, axial FWHMs increased with the MRD and ranged between 1.1 and 2.0 mm in the CFOV and between 1.5 and 2.7 mm in the EFOV for a MRD between 1 and 79. AFORE improved the axial resolution for a large part of the FOV, but image noise increased. OSEM3D/MAP yielded uniform spatial resolution in all directions, with an average FWHM of 1.65 +/- 0.06 mm. Sensitivity in the CFOV for the default energy and coincidence time window was 0.068; peak sensitivity was 0.111. Conclusion: The Inveon showed high spatial resolution and high sensitivity, both of which can be maintained using OSEM3D/MAP reconstruction instead of rebinning and 2D algorithms.
引用
收藏
页码:139 / 147
页数:9
相关论文
共 50 条
  • [1] Spatial resolution of the Inveon small-animal PET scanner for the entire field of view
    Disselhorst, J. A.
    Boerman, O. C.
    Oyen, W. J. G.
    Slump, C. H.
    Visser, E. P.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 615 (02): : 245 - 248
  • [2] Contribution of normalization to image noise for the Siemens Inveon small-animal PET scanner
    Visser, Eric P.
    Disselhorst, Jonathan A.
    Laverman, Peter
    Gotthardt, Martin
    Oyen, Wim J. G.
    Boerman, Otto C.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 605 (03): : 433 - 435
  • [3] Performance characterization of a high-sensitivity small-animal PET scanner
    Kao, Chien-Min
    Xie, Qingguo
    Dong, Yun
    Wan, Lu
    Chen, Chin-Tu
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 4176 - +
  • [4] A small animal PET scanner with millimeter spatial resolution.
    Del Guerra, A
    Di Domenico, G
    Scandola, M
    Zavattini, G
    Giganti, M
    Uccelli, L
    Piffanelli, A
    JOURNAL OF NUCLEAR MEDICINE, 1998, 39 (05) : 133P - 133P
  • [5] Temperature dependence of the LabPET small-animal PET scanner
    Keereman, Vincent
    Van Holen, Roel
    Vanhove, Christian
    Vandenberghe, Stefaan
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [6] Temperature Dependence of the LabPET Small-Animal PET Scanner
    Keereman, Vincent
    Vanhove, Christian
    Van Holen, Roel
    Vandenberghe, Stefaan
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 2896 - 2901
  • [7] A High-Sensitivity Small-Animal PET Scanner: Development and Initial Performance Measurements
    Kao, Chien-Min
    Xie, Qingguo
    Dong, Yun
    Wan, Lu
    Chen, Chin-Tu
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (05) : 2678 - 2688
  • [8] Continuously sampled digital pulse processing for Inveon small animal PET scanner
    McFarland, Aaron R.
    Siegel, Stefan
    Newport, Danny F.
    Mintzer, Robert
    Atkins, Blake
    Lenox, Mark
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 4262 - 4265
  • [9] A Monte Carlo investigation of the spatial resolution performance of a small-animal PET scanner designed for mouse brain imaging studies
    Rodriguez-Villafuerte, Mercedes
    Yang, Yongfeng
    Cherry, Simon R.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2014, 30 (01): : 76 - 85
  • [10] System Design Considerations for Collimation in a Small-Animal PET Scanner
    Li, Yusheng
    Matej, Samuel
    Karp, Joel S.
    Metzler, Scott D.
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 2838 - 2842