Development and Evaluation of Quality Control Dried Blood Spot Materials in Newborn Screening for Lysosomal Storage Disorders

被引:100
|
作者
De Jesus, Victor R. [1 ]
Zhang, X. Kate [2 ]
Keutzer, Joan [2 ]
Bodamer, Olaf A. [3 ]
Muehl, Adolf [3 ]
Orsini, Joseph J. [4 ]
Caggana, Michele [4 ]
Vogt, Robert F. [1 ]
Hannon, W. Harry [1 ]
机构
[1] Ctr Dis Control & Prevent, Newborn Screening & Mol Biol Branch, Atlanta, GA 30341 USA
[2] Genzyme, Framingham, MA USA
[3] Univ Childrens Hosp, Div Biochem & Paediat Genet, Vienna, Austria
[4] Wadsworth Ctr, New York Dept Hlth, Albany, NY USA
关键词
TANDEM MASS-SPECTROMETRY; ENZYMATIC DIAGNOSIS; FILTER-PAPER; DISEASE;
D O I
10.1373/clinchem.2008.111864
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
BACKGROUND: Lysosomal storage disorders (LSDs) comprise more than 40 genetic diseases that result in the accumulation of products that would normally be degraded by lysosomal enzymes. A tandem mass spectrometry (MS/MS)-based method is available for newborn screening for 5 LSDs, and many laboratories are initiating pilot studies to evaluate the incorporation of this method into their screening panels. We developed and evaluated dried blood spot (DBS) QC materials for LSDs and used the MS/MS method to investigate their suitability for LSD QC monitoring. METHODS: We incubated 3.2-mm punches from DBS controls for 20-24 h with assay cocktails containing substrate and internal standard. Using MS/MS, we quantified the resulting product and internal standard. Samples were run in triplicate for 3 consecutive days, and results were reported as product-to-internal standard ratios and enzyme activity units (mu mol/L/h). RESULTS: Enzyme activity interday imprecision (CV) for the high, medium, and low series were 3.4%-14.3% for galactocerebroside a-galactosidase, 6.8%-24.6% for acid a-galactosidase A, 7.36%-22.1% for acid sphingomyelinase, 6.2%-26.2% for acid alpha-glucocere-brosidase, and 7.0%-24.8% for lysosomal acid a-glucosidase (n = 9). In addition, DBS stored at -20 degrees and 4 degrees C showed minimal enzyme activity loss over a 187-d period. DBS stored at 37 degrees and 45 degrees C had lower activity values over the 187-day evaluation time. CONCLUSIONS: Suitable QC materials for newborn screening of LSDs were developed for laboratories performing DBS LSD screening. Good material linearity was observed, with goodness-of-fit values of 0.953 and higher. The QC materials may be used by screening laboratories that perform LSD analysis by MS and/or more conventional fluorescence-based screening methods. (C) 2008 American Association for Clinical Chemistry
引用
收藏
页码:158 / 164
页数:7
相关论文
共 48 条
  • [31] The stability of markers in dried-blood spots for recommended newborn screening disorders in the United States
    Adam, B. W.
    Hall, E. M.
    Sternberg, M.
    Lim, T. H.
    Flores, S. R.
    O'Brien, S.
    Simms, D.
    Li, L. X.
    De Jesus, V. R.
    Hannon, W. H.
    CLINICAL BIOCHEMISTRY, 2011, 44 (17-18) : 1445 - 1450
  • [32] Is Our Newborn Screening Working Well? A Literature Review of Quality Requirements for Newborn Blood Spot Screening (NBS) Infrastructure and Procedures
    Odenwald, Birgit
    Brockow, Inken
    Hanauer, Marianne
    Lueders, Anja
    Nennstiel, Uta
    INTERNATIONAL JOURNAL OF NEONATAL SCREENING, 2023, 9 (03)
  • [33] Lysosomal storage disorder 4+1 multiplex assay for newborn screening using tandem mass spectrometry: Application to a small-scale population study for five lysosomal storage disorders
    Orsini, Joseph J.
    Martin, Monica M.
    Showers, Amanda L.
    Bodamer, Olaf A.
    Zhang, X. Kate
    Gelb, Michael H.
    Caggana, Michele
    CLINICA CHIMICA ACTA, 2012, 413 (15-16) : 1270 - 1273
  • [34] Effect of blood volume on analytical bias in dried blood spots prepared for newborn screening external quality assurance
    Moat, Stuart J.
    Dibden, Catherine
    Tetlow, Lesley
    Griffith, Caroline
    Chilcott, Jim
    George, Roanna
    Hamilton, Laura
    Wu, Teresa H. Y.
    MacKenzie, Finlay
    Hall, Susan Kate
    BIOANALYSIS, 2020, 12 (02) : 99 - 109
  • [35] Harmonization of distributed multi-center analysis based on dried blood spot reference materials supporting the screening of neonatal inherited metabolic disorders
    Qu, Shou-Fang
    Tao, Hao-Ran
    Qin, Liu-Ji
    Zhang, Wen-Xin
    Han, Shan
    Zhang, Shen-Yan
    Huang, Jie
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2023, 37 (19-20)
  • [36] Short-incubation mass spectrometry assay for lysosomal storage disorders in newborn and high-risk population screening
    Mechtler, Thomas P.
    Metz, Thomas F.
    Mueller, Hannes G.
    Ostermann, Katharina
    Ratschmann, Rene
    De Jesus, Victor R.
    Shushan, Bori
    Di Bussolo, Joseph M.
    Herman, Joseph L.
    Herkner, Kurt R.
    Kasper, David C.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2012, 908 : 9 - 17
  • [37] What is the right blood hematocrit preparation procedure for standards and quality control samples for dried blood spot analysis?
    Koster, Remco A.
    Alffenaar, Jan-Willem C.
    Botma, Rixt
    Greijdanus, Ben
    Touw, Daan J.
    Uges, Donald R. A.
    Kosterink, Jos G. W.
    BIOANALYSIS, 2015, 7 (03) : 345 - 351
  • [38] Storage stability of five steroids and in dried blood spots for newborn screening and retrospective diagnosis of congenital adrenal hyperplasia
    Grecso, Nora
    Zadori, Anita
    Szecsi, Ilona
    Barath, Akos
    Galla, Zsolt
    Bereczki, Csaba
    Monostori, Peter
    PLOS ONE, 2020, 15 (05):
  • [39] Easy method for newborn screening of six lysosomal storage disorders using online solid-phase extraction with mass spectrometry
    Serdar, Muhittin
    Lay, Incilay
    Coskun, Julide
    Aslan, Berna
    Aslan, Huseyin
    Coskun, Abdurrahman
    Serteser, Mustafa
    Unsal, Ibrahim
    Ozpinar, Aysel
    TURKISH JOURNAL OF BIOCHEMISTRY-TURK BIYOKIMYA DERGISI, 2016, 41 (04): : 302 - 307
  • [40] Bio-Plex immunoassay measuring the quantity of lysosomal N-acetylgalactosamine-6-sulfatase protein in dried blood spots for the screening of mucopolysaccharidosis IVA in newborn: a pilot study
    Chuang, Chih-Kuang
    Lin, Hsiang-Yu
    Wang, Tuan-Jen
    Huang, Sung-Fa
    Lin, Shuan-Pei
    BMJ OPEN, 2017, 7 (07):