An Algorithm for Intelligent Sorting of CT-Related Dose Parameters

被引:4
作者
Cook, Tessa S. [1 ]
Zimmerman, Stefan L. [2 ]
Steingall, Scott R. [1 ]
Boonn, William W. [1 ]
Kim, Woojin [1 ]
机构
[1] Hosp Univ Penn, Dept Radiol, Philadelphia, PA 19104 USA
[2] Johns Hopkins Univ, Dept Radiol, Baltimore, MD USA
关键词
RADIANCE; Computed tomography; Dose monitoring; CT series separation; Radiation dose; Data extraction; Databases; RADIATION;
D O I
10.1007/s10278-011-9410-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Imaging centers nationwide are seeking innovative means to record and monitor computed tomography (CT)-related radiation dose in light of multiple instances of patient overexposure to medical radiation. As a solution, we have developed RADIANCE, an automated pipeline for extraction, archival, and reporting of CT-related dose parameters. Estimation of whole-body effective dose from CT dose length product (DLP)-an indirect estimate of radiation dose-requires anatomy-specific conversion factors that cannot be applied to total DLP, but instead necessitate individual anatomy-based DLPs. A challenge exists because the total DLP reported on a dose sheet often includes multiple separate examinations (e. g., chest CT followed by abdominopelvic CT). Furthermore, the individual reported series DLPs may not be clearly or consistently labeled. For example, "arterial" could refer to the arterial phase of the triple liver CT or the arterial phase of a CT angiogram. To address this problem, we have designed an intelligent algorithm to parse dose sheets for multi-series CT examinations and correctly separate the total DLP into its anatomic components. The algorithm uses information from the departmental PACS to determine how many distinct CT examinations were concurrently performed. Then, it matches the number of distinct accession numbers to the series that were acquired and anatomically matches individual series DLPs to their appropriate CT examinations. This algorithm allows for more accurate dose analytics, but there remain instances where automatic sorting is not feasible. To ultimately improve radiology patient care, we must standardize series names and exam names to unequivocally sort exams by anatomy and correctly estimate whole-body effective dose.
引用
收藏
页码:179 / 188
页数:10
相关论文
共 16 条
[1]  
Amis E Stephen Jr, 2007, J Am Coll Radiol, V4, P272, DOI 10.1016/j.jacr.2007.03.002
[2]   Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[3]   Radiation risk to children from computed tomography [J].
Brody, Alan S. ;
Frush, Donald P. ;
Huda, Walter ;
Brent, Robert L. .
PEDIATRICS, 2007, 120 (03) :677-682
[4]   Estimating Effective Dose for CT Using Dose-Length Product Compared With Using Organ Doses: Consequences of Adopting International Commission on Radiological Protection Publication 103 or Dual-Energy Scanning [J].
Christner, Jodie A. ;
Kofler, James M. ;
McCollough, Cynthia H. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (04) :881-889
[5]   Automated Extraction of Radiation Dose Information for CT Examinations [J].
Cook, Tessa S. ;
Zimmerman, Stefan ;
Maidment, Andrew D. A. ;
Kim, Woojin ;
Boonn, William W. .
JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2010, 7 (11) :871-877
[6]   Projected Cancer Risks From Computed Tomographic Scans Performed in the United States in 2007 [J].
de Gonzalez, Amy Berrington ;
Mahesh, Mahadevappa ;
Kim, Kwang-Pyo ;
Bhargavan, Mythreyi ;
Lewis, Rebecca ;
Mettler, Fred ;
Land, Charles .
ARCHIVES OF INTERNAL MEDICINE, 2009, 169 (22) :2071-2077
[7]  
DICOM Standards Committee, 2007, DICOM STAND S127
[8]  
DICOM Standards Committee, 2008, DICOM STAND 16
[9]   The 'Image Gently' campaign: increasing CT radiation dose awareness through a national education and awareness program [J].
Goske, Marilyn J. ;
Applegate, Kimberly E. ;
Boylan, Jennifer ;
Butler, Penny F. ;
Callahan, Michael J. ;
Coley, Brian D. ;
Farley, Shawn ;
Frush, Donald P. ;
Hernanz-Schulman, Marta ;
Jaramillo, Diego ;
Johnson, Neil D. ;
Kaste, Sue C. ;
Morrison, Gregory ;
Strauss, Keith J. ;
Tuggle, Nora .
PEDIATRIC RADIOLOGY, 2008, 38 (03) :265-269
[10]  
Kase KR, 2009, 160 NAT COUNC RAD PR