Accuracy of a narrow-diameter battery-powered endoscope in sedated and unsedated patients

被引:26
作者
Catanzaro, A [1 ]
Faulx, A [1 ]
Pfau, PR [1 ]
Cooper, G [1 ]
Isenberg, G [1 ]
Wong, RCK [1 ]
Sivak, MV [1 ]
Chak, A [1 ]
机构
[1] Case Western Reserve Univ, Univ Hosp Cleveland, Div Gastroenterol, Cleveland, OH 44106 USA
关键词
D O I
10.1067/mge.2002.122576
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Background. Esophagoscopy with a portable battery-powered endoscope could provide a safe, inexpensive, and minimally invasive way to screen for Barrett's esophagus or esophageal varices. The use of such an instrument in an unsedated fashion has not been previously evaluated. Methods: Patients referred for an EGD were recruited to undergo an additional examination with the battery-powered endoscope before EGD. In phase 1, (n = 42) patients received conscious sedation before the battery-powered endoscopic examination. In phase 2, (n = 56) patients were not sedated and were given the option of a peroral (n = 43) or transnasal (n = 13) endoscopy. Examiners were blinded to patient history and procedure indications. Esophageal findings, ease of intubation, optical quality, and patient comfort for the battery-powered endoscope and standard EGD were recorded by the endoscopist. Results: Ninety-eight patients (60 men, 38 women, mean age 53 years) were recruited. The sensitivity for detecting Barrett's esophagus, esophageal tumors, and esophageal varices was 54.5%, 66.7%, and 80%, respectively. Ease of intubation and patient comfort as perceived by the endoscopist were not significantly different between the battery-powered endoscope and EGD. Optical quality was ranked as less than 4 (on a 5-point scale with 5 = standard EGD and 1 = poor) in 42% of battery-powered endoscopic examinations. There were no complications. Conclusion: The accuracy of esophageal examination with a 3.1-mm endoscope is substantially inferior to standard EGD. Thus, the battery-powered endoscope would not be useful for screening patients for Barrett's esophagus or varices unless improvements in optical quality and visualization are made.
引用
收藏
页码:484 / 487
页数:4
相关论文
共 14 条
  • [1] BELL GD, 1990, ALIMENT PHARM THERAP, V4, P103
  • [2] A comparative study of unsedated transnasal esophagogastroduodenoscopy and conventional EGD
    Dean, R
    Dua, K
    Massey, B
    Berger, W
    Hogan, WJ
    Shaker, R
    [J]. GASTROINTESTINAL ENDOSCOPY, 1996, 44 (04) : 422 - 424
  • [3] Peroral ultrathin endoscopy in adult patients
    DeGregorio, BT
    Poorman, JC
    Katon, RM
    [J]. GASTROINTESTINAL ENDOSCOPY, 1997, 45 (03) : 303 - 306
  • [4] THE MEASUREMENT OF CLINICAL PAIN INTENSITY - A COMPARISON OF 6 METHODS
    JENSEN, MP
    KAROLY, P
    BRAVER, S
    [J]. PAIN, 1986, 27 (01) : 117 - 126
  • [5] CARDIOPULMONARY RISK OF ESOPHAGOGASTRODUODENOSCOPY - ROLE OF ENDOSCOPE DIAMETER AND SYSTEMIC SEDATION
    LIEBERMAN, DA
    WUERKER, CK
    KATON, RM
    [J]. GASTROENTEROLOGY, 1985, 88 (02) : 468 - 472
  • [6] O'CONNOR K W, 1990, Gastrointestinal Endoscopy, V36, pS2
  • [7] Prospective comparison of nasal versus oral insertion of a thin video endoscope in healthy volunteers
    Rey, JF
    Duforest, D
    Marek, TA
    [J]. ENDOSCOPY, 1996, 28 (05) : 422 - 424
  • [8] HYPOXIA DURING UPPER GASTROINTESTINAL ENDOSCOPY IS CAUSED BY SEDATION
    SCOTTCOOMBES, DM
    THOMPSON, JN
    [J]. ENDOSCOPY, 1993, 25 (04) : 308 - 309
  • [9] UNSEDATED TRANS-NASAL PHARYNGOESOPHAGOGASTRODUODENOSCOPY (T-EGD) - TECHNIQUE
    SHAKER, R
    [J]. GASTROINTESTINAL ENDOSCOPY, 1994, 40 (03) : 346 - 348
  • [10] A wake-up call? Unsedated versus conventional esophagogastroduodenoscopy
    Shaker, R