A study of lateral fall-off (penumbra) optimisation for pencil beam scanning (PBS) proton therapy

被引:43
作者
Winterhalter, C. [1 ]
Lomax, A. [1 ]
Oxley, D. [1 ]
Weber, D. C. [1 ,2 ,3 ]
Safai, S. [1 ]
机构
[1] Paul Scherrer Inst, Ctr Proton Therapy, CH-5232 Villigen, Switzerland
[2] Univ Hosp Bern, Radiat Oncol Dept, Bern, Switzerland
[3] Univ Hosp Zurich, Radiat Oncol Dept, Zurich, Switzerland
关键词
proton therapy; pencil beam scanning; penumbra; collimation; optimisation; pre-absorber; MONTE-CARLO; DOSE CALCULATION; SPOT SIZE; COLLIMATION; DELIVERY; APERTURE; SYSTEM;
D O I
10.1088/1361-6560/aaa2ad
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The lateral fall-off is crucial for sparing organs at risk in proton therapy. It is therefore of high importance to minimize the penumbra for pencil beam scanning (PBS). Three optimisation approaches are investigated: edge-collimated uniformly weighted spots (collimation), pencil beam optimisation of uncollimated pencil beams (edge-enhancement) and the optimisation of edge collimated pencil beams (collimated edge-enhancement). To deliver energies below 70 MeV, these strategies are evaluated in combination with the following pre-absorber methods: field specific fixed thickness pre-absorption (fixed), range specific, fixed thickness pre-absorption (automatic) and range specific, variable thickness pre-absorption (variable). All techniques are evaluated by Monte Carlo simulated square fields in a water tank. For a typical air gap of 10 cm, without pre-absorber collimation reduces the penumbra only for water equivalent ranges between 4-11 cm by up to 2.2 mm. The sharpest lateral fall-off is achieved through collimated edge-enhancement, which lowers the penumbra down to 2.8 mm. When using a pre-absorber, the sharpest fall-offs are obtained when combining collimated edge-enhancement with a variable pre-absorber. For edge-enhancement and large air gaps, it is crucial to minimize the amount of material in the beam. For small air gaps however, the superior phase space of higher energetic beams can be employed when more material is used. In conclusion, collimated edge-enhancement combined with the variable pre-absorber is the recommended setting to minimize the lateral penumbra for PBS. Without collimator, it would be favourable to use a variable pre-absorber for large air gaps and an automatic pre-absorber for small air gaps.
引用
收藏
页数:14
相关论文
共 29 条
[11]   Characterizing a proton beam scanning system for Monte Carlo dose calculation in patients [J].
Grassberger, C. ;
Lomax, Anthony ;
Paganetti, H. .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (02) :633-645
[12]   A Monte Carlo pencil beam scanning model for proton treatment plan simulation using GATE/GEANT4 [J].
Grevillot, L. ;
Bertrand, D. ;
Dessy, F. ;
Freud, N. ;
Sarrut, D. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (16) :5203-5219
[13]   Effects of spot size and spot spacing on lateral penumbra reduction when using a dynamic collimation system for spot scanning proton therapy [J].
Hyer, Daniel E. ;
Hill, Patrick M. ;
Wang, Dongxu ;
Smith, Blake R. ;
Flynn, Ryan T. .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (22) :N187-N196
[14]   Physics settings for using the Geant4 toolkit in proton therapy [J].
Jarlskog, Christina Zacharatou ;
Paganetti, Harald .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (03) :1018-1025
[15]   Intensity modulation methods for proton radiotherapy [J].
Lomax, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (01) :185-205
[16]   Contour scanning for penumbra improvement in pencil beam scanned proton therapy [J].
Meier, G. ;
Leiser, D. ;
Besson, R. ;
Mayor, A. ;
Safai, S. ;
Weber, D. C. ;
Lomax, A. J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (06) :2398-2416
[17]   Impact of Spot Size and Beam-Shaping Devices on the Treatment Plan Quality for Pencil Beam Scanning Proton Therapy [J].
Moteabbed, Maryam ;
Yock, Torunn I. ;
Depauw, Nicolas ;
Madden, Thomas M. ;
Kooy, Hanne M. ;
Paganetti, Harald .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 95 (01) :190-198
[18]   Relative biological effectiveness (RBE) values for proton beam therapy [J].
Paganetti, H ;
Niemierko, A ;
Ancukiewicz, M ;
Gerweck, LE ;
Goitein, M ;
Loeffler, JS ;
Suit, HD .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (02) :407-421
[19]   Pencil beam characteristics of the next-generation proton scanning gantry of PSI: design issues and initial commissioning results [J].
Pedroni, E. ;
Meer, D. ;
Bula, C. ;
Safai, S. ;
Zenklusen, S. .
EUROPEAN PHYSICAL JOURNAL PLUS, 2011, 126 (07) :1-27
[20]   THE 200-MEV PROTON THERAPY PROJECT AT THE PAUL-SCHERRER-INSTITUTE - CONCEPTUAL DESIGN AND PRACTICAL REALIZATION [J].
PEDRONI, E ;
BACHER, R ;
BLATTMANN, H ;
BOHRINGER, T ;
CORAY, A ;
LOMAX, A ;
LIN, SX ;
MUNKEL, G ;
SCHEIB, S ;
SCHNEIDER, U ;
TOUROVSKY, A .
MEDICAL PHYSICS, 1995, 22 (01) :37-53