Dependence of the LR-115 radon detector calibration factor on track density

被引:4
作者
De Cicco, F. [1 ,3 ]
Pugliese, M. [2 ,3 ]
Roca, V. [2 ,3 ]
Sabbarese, C. [1 ,3 ]
机构
[1] Univ Naples 2, Dipartimento Sci Ambientali, Caserta, Italy
[2] Univ Naples Federico II, Dipartimento Sci Fis, Naples, Italy
[3] Ist Nazl Fis Nucl, Sez Napoli, I-80125 Naples, Italy
关键词
LR-115 calibration factor; Track density saturation; Track counting correction; Indoor radon; In soil gas radon; PARAMETERS;
D O I
10.1016/j.apradiso.2013.03.028
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
The reliability and accuracy of the methodology based on using LR-115 track detectors for radon measurements have been studied by determining the dependence of their calibration factors on radon exposure at levels reaching 13 MBq m(-3) h. This factor results not constant and demonstrated a decreasing exponential trend vs. exposure that has been explained in terms of the saturation effect and verified using a numerical simulation. This dependence does not affect the parameter that normalizes track density vs. film thickness. This parameter results constant and equal to - 0.30 +/- 0.02 cm(-2)/mu m in the 300-8000 kBq h m(-3) exposure range. (c) 2013 Published by Elsevier Ltd.
引用
收藏
页码:108 / 112
页数:5
相关论文
共 14 条
  • [1] AZIMIGARAKANI D, 1988, RADIAT PROT DOSIM, V24, P267
  • [2] Annual average and seasonal variations of residential radon concentration for all the Italian Regions
    Bochicchio, F
    Campos-Venuti, G
    Piermattei, S
    Nuccetelli, C
    Risica, S
    Tommasino, L
    Torri, G
    Magnoni, M
    Agnesod, G
    Sgorbati, G
    Bonomi, M
    Minach, L
    Trotti, F
    Malisan, MR
    Maggiolo, S
    Gaidolfi, L
    Giannardi, C
    Rongoni, A
    Lombardi, M
    Cherubini, G
    D'Ostilio, S
    Cristofaro, C
    Pugliese, M
    Martucci, V
    Crispino, A
    Cuzzocrea, P
    Santamaria, AS
    Cappai, M
    [J]. RADIATION MEASUREMENTS, 2005, 40 (2-6) : 686 - 694
  • [3] Uncertainty evaluation of radon measurements with LR115 detector and spark counter
    Campi, F
    Caresana, M
    Ferrarini, M
    Garlati, L
    Palermo, M
    Rusconi, R
    [J]. RADIATION PROTECTION DOSIMETRY, 2004, 111 (01) : 59 - 64
  • [4] Campos Venuti G., 1987, RADIAT PROT DOSIM, V24, P379
  • [5] Evaluation of etching correction factor for LR115 cellulose nitrate films from track parameters
    Caresana, M
    Campi, F
    Ferrarini, M
    [J]. RADIATION PROTECTION DOSIMETRY, 2005, 113 (04) : 354 - 358
  • [6] CHEROUATI DE, 1988, NUCL TRACKS RAD MEAS, V15, P583
  • [7] Calibration factors for LR-115 (type-II) based radon thoron discriminating dosimeter
    Eappen, KP
    Mayya, YS
    [J]. RADIATION MEASUREMENTS, 2004, 38 (01) : 5 - 17
  • [8] On radon surveys: Design and data interpretation
    Font, Ll.
    [J]. RADIATION MEASUREMENTS, 2009, 44 (9-10) : 964 - 968
  • [9] Evaluation of Efficiency Calibration Parameters of the LR-115 Radon Detector
    Mireles, F.
    Davila, J. I.
    Garcia-Betancourt, M. L.
    Pinedo, J. L.
    Lopez, H.
    [J]. HEALTH PHYSICS, 2010, 98 (05): : S63 - S68
  • [10] Radon as an earthquake precursor
    Planinic, J
    Radolic, V
    Vukovic, B
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 530 (03) : 568 - 574