De-bundled single-walled carbon nanotube-modified sensors for simultaneous differential pulse voltammetric determination of ascorbic acid, dopamine, and uric acid

被引:28
|
作者
Thirumalai, Dinakaran [1 ]
Subramani, Devaraju [2 ]
Yoon, Jang-Hee [3 ]
Lee, Jaewon [4 ]
Paik, Hyun-jong [2 ]
Chang, Seung-Cheol [5 ]
机构
[1] Pusan Natl Univ, Grad Dept Chem Mat, Busan 46241, South Korea
[2] Pusan Natl Univ, Dept Polymer Sci & Engn, Busan 46241, South Korea
[3] Korea Basic Sci Inst, Busan Ctr, Busan 46742, South Korea
[4] Pusan Natl Univ, Mol Inflammat Res Ctr Aging Intervent, Coll Pharm, Busan 46241, South Korea
[5] Pusan Natl Univ, Inst Biophysio Sensor Technol, Busan 46241, South Korea
基金
新加坡国家研究基金会;
关键词
TRANSPARENT CONDUCTIVE FILMS; GRAPHENE OXIDE; MODIFIED ELECTRODE; FABRICATION; NANOPARTICLES; NANOMATERIALS; TRYPTOPHAN; DISPERSION; BIOSENSORS; OXIDATION;
D O I
10.1039/c7nj04371f
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
A new approach based on differential pulse voltammetry (DPV) was developed for the simultaneous determination of ascorbic acid (AA), dopamine (DA), and uric acid (UA) using a modified glassy carbon electrode (GCE). The sensor was constructed by a simple "one-step" technique, wherein de-bundled single-walled carbon nanotubes (SWCNTs) were drop-cast onto the GCE. Without de-bundling, the SWCNTs were poorly dispersed in aqueous solution and were ineffective for the one-step procedure. De-bundling of the SWCNTs was achieved using a small amount (0.1 wt%) of the synthesized polymer dispersant, sulfonated poly(ether sulfone) (SPES); the de-bundled SWCNTs had a high aspect ratio (length = 2.5 +/- 1.0 mu m; height = 2 +/- 1 nm, as determined using transmission electron microscopy and atomic force microscopy). The de-bundled SWCNTs also led to enhanced electrocatalytic activity and selectivity of the modified sensor for the simultaneous determination of AA, DA, and UA in DPV measurements: the peak-to-peak separation values were 221, 119, and 340 mV (vs. Ag/AgCl) for DA-AA, UA-DA, and AA-UA, respectively. The dynamic linear ranges for AA, DA, and UA were 0.2-1.6 mM, 5.0-50 mu M, and 5.0-60 mu M, and the detection limits were 10.6 mu M, 15 nM, and 113 nM (S/N = 3), respectively. The analytical performance of the developed sensor was demonstrated in the determination of AA and DA in commercial pharmaceutical samples (vitamin C tablets and DA injection). The characteristics of the modified sensor make it promising for the individual or simultaneous determination of AA, DA, and UA.
引用
收藏
页码:2432 / 2438
页数:7
相关论文
共 50 条
  • [1] A Metal-Free, Non-Enzymatic Electrochemical Glucose Sensor with a de-Bundled Single-Walled Carbon Nanotube-Modified Electrode
    Thirumalai, Dinakaran
    Subramani, Devaraju
    Shin, Bosung
    Paik, Hyun-jong
    Chang, Seung-Cheol
    BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2018, 39 (02): : 141 - 145
  • [2] Differential pulse voltammetric simultaneous determination of ascorbic acid, dopamine and uric acid on a glassy carbon electrode modified with electroreduced graphene oxide and imidazolium groups
    Wu, Feng
    Huang, Ting
    Hu, Yangjian
    Yang, Xin
    Ouyang, Yuejun
    Xie, Qingji
    MICROCHIMICA ACTA, 2016, 183 (09) : 2539 - 2546
  • [3] Simultaneous Determination of Ascorbic Acid, Dopamine, and Uric Acid by Differential Pulse Voltammetry using Tiron Modified Glassy Carbon Electrode
    Ensafi, Ali A.
    Taei, M.
    Khayamian, Taghi
    INTERNATIONAL JOURNAL OF ELECTROCHEMICAL SCIENCE, 2010, 5 (01): : 116 - 130
  • [4] Carbon-nanotube-modified glassy carbon electrode for simultaneous determination of dopamine, ascorbic acid and uric acid: The effect of functional groups
    Bi, Huaqing
    Li, Yanhui
    Liu, Shufeng
    Guo, Peizhi
    Wei, Zhongbin
    Lv, Chunxiao
    Zhang, Jizhen
    Zhao, X. S.
    SENSORS AND ACTUATORS B-CHEMICAL, 2012, 171 : 1132 - 1140
  • [5] Simultaneous determination of ascorbic acid, dopamine and uric acid by use of a MWCNT modified carbon-ceramic electrode and differential pulse voltammetry
    Habibi, Biuck
    Pournaghi-Azar, Mohammad Hossien
    ELECTROCHIMICA ACTA, 2010, 55 (19) : 5492 - 5498
  • [6] Simultaneous determination of ascorbic acid, dopamine and uric acid using a glassy carbon electrode modified with the nickel(II)-bis(1,10-phenanthroline) complex and single-walled carbon nanotubes
    Yan, Songling
    Li, Xi
    Xiong, Yan
    Wang, Mengmeng
    Yang, Linlin
    Liu, Xin
    Li, Xiaoyu
    Alshahrani, Lina Abdullah M.
    Liu, Peng
    Zhang, Chaocan
    MICROCHIMICA ACTA, 2016, 183 (04) : 1401 - 1408
  • [7] Electrocatalytic detection of dopamine in the presence of ascorbic acid and uric acid using single-walled carbon nanotubes modified electrode
    Li, Yaya
    Du, Jie
    Yang, Jiandong
    Liu, Dong
    Lu, Xiaoquan
    COLLOIDS AND SURFACES B-BIOINTERFACES, 2012, 97 : 32 - 36
  • [8] Carbon Black-Carbon Nanotube Co-Doped Polyimide Sensors for Simultaneous Determination of Ascorbic Acid, Uric Acid, and Dopamine
    Wang, Yue
    Yang, Tian
    Hasebe, Yasushi
    Zhang, Zhiqiang
    Tao, Dongping
    MATERIALS, 2018, 11 (09)
  • [9] Screen-printed carbon electrode modified with de-bundled single-walled carbon nanotubes for voltammetric determination of norepinephrine in ex vivo rat tissue
    Rajarathinam, Thenmozhi
    Thirumalai, Dinakaran
    Kwon, Minho
    Lee, Seulah
    Jayaraman, Sivaguru
    Paik, Hyun-jong
    Lee, Jaewon
    Chang, Seung-Cheol
    BIOELECTROCHEMISTRY, 2022, 146
  • [10] CTAB functionalized graphene oxide/multiwalled carbon nanotube composite modified electrode for the simultaneous determination of ascorbic acid, dopamine, uric acid and nitrite
    Yang, Yu Jun
    Li, Weikun
    BIOSENSORS & BIOELECTRONICS, 2014, 56 : 300 - 306