Diagnostic accuracy of rest/stress ECG-gated Rb-82 myocardial perfusion PET: Comparison with ECG-gated Tc-99m sestamibi SPECT

被引:340
|
作者
Bateman, TM
Heller, GV
McGhie, AI
Friedman, JD
Case, JA
Bryngelson, JR
Hertenstein, GK
Moutray, KL
Reid, K
Cullom, SJ
机构
[1] Cardiovasc Consultants, PC, Kansas City, MO 64111 USA
[2] Cardiovasc Imaging Technol LLC, Kansas City, MO USA
[3] Mid Amer Heart Inst, Kansas City, MO USA
[4] Hartford Hosp, Henry Low Heart Ctr, Div Cardiol, Hartford, CT 06115 USA
[5] Cedars Sinai Med Ctr, Los Angeles, CA 90048 USA
关键词
single photon emission computed tomography; positron emission tomography; myocardial perfusion imaging;
D O I
10.1016/j.nuclcard.2005.12.004
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Although single photon emission computed tomography (SPECT) and positron emission tomography (PET) myocardial perfusion imaging (MPI) have evolved considerably over the last decade, there is no recent comparison of diagnostic performance. This study was designed to assess relative image quality, interpretive confidence, and diagnostic accuracy by use of contemporary technology and protocols. Methods and Results. By consensus and without clinical information, 4 experienced nuclear cardiologists interpreted 112 SPECT technetium-99m sestamibi and 112 PET rubidium-82 MPI electrocardiography (ECG)-gated rest/pharmacologic stress studies in patient populations matched by gender, body mass index, and presence and extent of coronary disease. The patients were categorized as having a low likelihood for coronary artery disease (27 in each group) or had coronary angiography within 60 days. SPECT scans were acquired on a Cardio-60 system and PET scans on an ECAT ACCEL scanner. Image quality was excellent for 78% and 79% of rest and stress PET scans, respectively, versus 62% and 62% of respective SPECT scans (both p <.05). An equal percent of PET and SPECT gated images were rated excellent in quality. Interpretations were definitely normal or abnormal for 96% of PET scans versus 81% of SPECT scans (p =.001). Diagnostic accuracy was higher for PET for both stenosis severity thresholds of 70% (89% vs 79%,p =.03) and 50% (87% vs 71%,p =.003) and was higher in men and women, in obese and nonobese patients, and for correct identification of multivessel coronary artery disease. Conclusion. In a large population of matched pharmacologic stress patients, myocardial perfusion PET was superior to SPECT in image quality, interpretive certainty, and diagnostic accuracy.
引用
收藏
页码:24 / 33
页数:10
相关论文
共 50 条
  • [41] Assessment of contractile response to dobutamine stress by means of ECG-gated myocardial SPECT: Comparison with myocardial perfusion and fatty acid metabolism
    Kumita, S
    Cho, K
    Nakajo, H
    Toba, M
    Fukushima, Y
    Mizumura, S
    Sano, J
    Takano, T
    Kumazaki, T
    ANNALS OF NUCLEAR MEDICINE, 2005, 19 (05) : 379 - 386
  • [42] Comparison of Tc-99m sestamibi and Tl-201 gated perfusion SPECT
    E. Gordon DePuey
    Steven Parmett
    Munir Ghesani
    Alan Rozanski
    Kenneth Nichols
    Helen Salensky
    Journal of Nuclear Cardiology, 1999, 6 : 278 - 285
  • [43] Comparison of Tc-99m sestamibi and Tl-201 gated perfusion SPECT
    DePuey, EG
    Parmett, S
    Ghesani, M
    Rozanski, A
    Nichols, K
    Salensky, H
    JOURNAL OF NUCLEAR CARDIOLOGY, 1999, 6 (03) : 278 - 285
  • [44] Assessment of contractile response to dobutamine stress by means of ECG-gated myocardial SPECT: Comparison with myocardial perfusion and fatty acid metabolism
    Shin-ichiro Kumita
    Keiichi Cho
    Hidenobu Nakajo
    Masahiro Toba
    Yoshimitsu Fukushima
    Sunao Mizumura
    Junko Sano
    Teruo Takano
    Tatsuo Kumazaki
    Annals of Nuclear Medicine, 2005, 19 : 379 - 386
  • [45] Early and delayed technetium-99m labeled sestamibi myocardial ECG-gated SPECT by QGS program in normal volunteers.
    Fujino, S
    Masuyama, K
    Kanayama, S
    Miura, S
    Ohkubo, S
    Kanemitsu, S
    Tsugawa, H
    Matsui, S
    Takekoshi, N
    JOURNAL OF NUCLEAR MEDICINE, 1999, 40 (05) : 180P - 180P
  • [46] Evaluation of cardiac resynchronization therapy in drug-resistant dilated-phase hypertrophic cardiomyopathy by means of Tc-99m sestamibi ECG-gated SPECT
    Shinro Matsuo
    Yuichi Sato
    Ichiro Nakae
    Daisuke Masuda
    Naoya Matsumoto
    Minoru Horie
    Annals of Nuclear Medicine, 2006, 20
  • [47] Evaluation of cardiac resynchronization therapy in drug-resistant dilated-phase hypertrophic cardiomyopathy by means of Tc-99m sestamibi ECG-gated SPECT
    Matsuo, Shinro
    Sato, Yuichi
    Nakae, Ichiro
    Masuda, Daisuke
    Matsumoto, Naoya
    Horie, Minoru
    ANNALS OF NUCLEAR MEDICINE, 2006, 20 (09) : 643 - 647
  • [48] Evaluation of a new automatic algorithm for quantification of ECG-gated 82Rb cardiac PET
    Slomka, Piotr
    Germano, Guido
    Kavanagh, Paul
    Javadi, Mehrbod
    Berman, Daniel
    Bengel, Frank
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50
  • [49] Assessment of regional and global left ventricular function by reinjection Tl-201 and rest Tc-99m sestamibi ECG-gated SPECT - Comparison with three-dimensional magnetic resonance imaging
    Tadamura, E
    Kudoh, T
    Motooka, M
    Inubushi, M
    Shirakawa, S
    Hattori, N
    Okada, T
    Matsuda, T
    Koshiji, T
    Nishimura, K
    Matsuda, K
    Konishi, J
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1999, 33 (04) : 991 - 997
  • [50] Feasibility of 2-hour rest/stress protocol of Tc-99m sestamibi gated myocardial SPECT
    Song, HC
    Bom, HS
    Min, JJ
    Jeong, HJ
    Kim, JY
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1999, 26 (09): : 1068 - 1068