Evaluation of the Stryker S2™ IM Nail Distal Targeting Device for reduction of radiation exposure: A case series study

被引:19
作者
Anastopoulos, George [1 ]
Ntagiopoulos, Panagiotis G. [1 ]
Chissas, Dionisios [1 ]
Loupasis, George [1 ]
Asimakopoulos, Antonios [1 ]
Athanaselis, Eustratios [2 ]
Megas, Panagiotis [2 ]
机构
[1] G Gennimatas Gen Hosp Athens, Dept Orthopaed & Trauma Surg 2, Athens, Greece
[2] Univ Hosp Patras, Dept Orthopaed & Trauma Surg, Patras, Greece
来源
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED | 2008年 / 39卷 / 10期
关键词
Intramedullary nailing; Distal Locking; Radiation exposure; Targeting devices;
D O I
10.1016/j.injury.2008.04.020
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Distal locking is one challenging step during intramedullary nailing of femoral shaft fractures that can lead to an increase of radiation exposure. In the present study, the authors describe a technique for the distal Locking of femoral nails, implementing a new targeting device in an attempt to reduce radiation exposure and operational time. Over a 2-year period, 127 consecutive cases of femoral shaft fractures were included in the study. ALL cases were treated with nailing of femoral shaft fractures with an unslotted reamed antegrade femoral nail and distal Locking was performed with the use of a proximally mounted aiming device. Mean duration of the procedure was 63.5 +/- 18.1 min while the duration for distal Locking was 6.6 +/- 12.6 min. In all successful cases, exposure from intraoperative fluoroscopy was 17.2 +/- 7.4 s for the whole operative procedure, and for distal locking was 2 shots, 1.35 s (range, 0.9-2.2 s) and 1.9 mGy (range, 1.1-2.9 mGy). Five cases (3.9%) were unsuccessful, but overall no intraoperative complications were encountered from the application of this technique. The ability of the device to correspond to the Level of nail deformation and to property identify the distal holes, reduced exposure to radiation compared to other published reports, and should be considered as a valuable toot for distal. locking of femoral fractures. (c) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1210 / 1215
页数:6
相关论文
共 30 条
  • [1] ANASTOPOULOS G, 2006, PRACTICE INTRAMEDULL, P271
  • [2] Intramedullary nailing of the lower extremity: Biomechanics and biology
    Bong, Matthew R.
    Kummer, Frederick J.
    Koval, Kenneth J.
    Egol, Kenneth A.
    [J]. JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2007, 15 (02) : 97 - 106
  • [3] CONLAN D P, 1990, Journal of the Royal College of Surgeons of Edinburgh, V35, P61
  • [4] Devalia Kailash L, 2006, Acta Orthop Belg, V72, P615
  • [5] AN IMAGE INTENSIFIER LASER GUIDANCE-SYSTEM FOR THE DISTAL LOCKING OF AN INTRAMEDULLARY NAIL
    GOODALL, JD
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1991, 22 (04): : 339 - 339
  • [6] GOULET JA, 1992, CLIN ORTHOP RELAT R, P199
  • [7] A SIMPLE JIG TO EASE THE INSERTION OF DISTAL SCREWS IN INTRAMEDULLARY LOCKING NAILS
    HASHEMINEJAD, A
    GARLICK, N
    GODDARD, NJ
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1994, 25 (06): : 407 - 408
  • [8] LOCKING NAILING - AN AID TO DISTAL TARGETTING
    HUDSON, I
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1989, 20 (03): : 129 - 130
  • [9] HUTSON JJ, 1995, CLIN ORTHOP RELAT R, P129
  • [10] A SIMPLE TECHNIQUE FOR INSERTION OF DISTAL SCREWS INTO INTERLOCKING NAILS
    KELLEY, SS
    BONAR, S
    HUSSAMY, OD
    MORRISON, JA
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 1995, 9 (03) : 227 - 230