Evaluation of the Fourth-Generation FloTrac/Vigileo System in Comparison With the Intermittent Bolus Thermodilution Method in Patients Undergoing Cardiac Surgery

被引:24
作者
Kusaka, Yusuke [1 ]
Ohchi, Fumihiro [1 ]
Minami, Toshiaki [1 ]
机构
[1] Osaka Med Coll, Dept Anesthesiol, Daigaku Machi 2-7, Takatsuki, Osaka 5698686, Japan
关键词
FloTrac/Vigileo system; cardiac output; pulse pressure; systemic vascular resistance index; pulmonary artery catheter; PULSE CONTOUR ANALYSIS; VIGILEO-FLOTRAC; OUTPUT; RELIABILITY; IMPACT;
D O I
10.1053/j.jvca.2018.06.017
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Objectives. The aim of this study was to evaluate the accuracy, precision, and trending ability of the fourth-generation FloTrac/Vigileo system (version 4.00; Edwards Lifesciences, Irvine, CA) by comparing cardiac output derived from FloTracNigileo system (COAP) with that measured by a pulmonary artery catheter (COTD), and to determine the effects of hemodynamic variables on the bias between COTD and COAP. Design: A prospective study. Setting: University hospital. Participants: Thirty patients undergoing elective cardiac surgery using cardiopulmonary bypass. Interventions: Including hemodynamic variables, COTD and COAP were measured simultaneously at the following 10 time points: after the induction of anesthesia, at the start of operation, after sternotomy, before and after the administration of heparin, before and after the administration of protamine, at the start of sternal closure, at the end of operation, and on arrival to intensive care unit. Measurements and Main Results: In total, 280 pairs of datasets were obtained. Bland-Altman analysis showed a bias of -0.41 L/min, a precision of 0.72 L/min, and limits of agreement of -1.85 and 1.03 L/min, with a percentage error of 37.1%. The concordance rate determined by 4-quadrant plot analysis and the polar concordance rate were 76% and 79%, respectively. The linear mixed-effect model revealed that the bias was influenced strongly by the difference in pulse pressure between the radial and femoral artery (p < 0.001), and the systemic vascular resistance index (p < 0.001). Conclusion: The fourth-generation FloTrac/Vigileo system still lacks accuracy and trending ability in cardiac surgery, and the discrepancy in cardiac output measurement depends on the peripheral vascular tone. Further improvement of this system is needed. (C) 2018 Elsevier Inc. All rights reserved.
引用
收藏
页码:953 / 960
页数:8
相关论文
共 16 条
  • [1] Am Soc Anesthesiologists Task Forc, 2003, ANESTHESIOLOGY, V99, P988
  • [2] Agreement between methods of measurement with multiple observations per individual
    Bland, J. Martin
    Altman, Douglas G.
    [J]. JOURNAL OF BIOPHARMACEUTICAL STATISTICS, 2007, 17 (04) : 571 - 582
  • [3] Bench-to-bedside review: The importance of the precision of the reference technique in method comparison studies - with specific reference to the measurement of cardiac output
    Cecconi, Maurizio
    Rhodes, Andrew
    Poloniecki, Jan
    Della Rocca, Giorgio
    Grounds, R. Michael
    [J]. CRITICAL CARE, 2009, 13 (01):
  • [4] Comparison of cardiac output measures by transpulmonary thermodilution, pulse contour analysis, and pulmonary artery thermodilution during off-pump coronary artery bypass surgery: a subgroup analysis of the cardiovascular anaesthesia registry at a single tertiary centre
    Cho, Youn Joung
    Koo, Chang-Hoon
    Kim, Tae Kyong
    Hong, Deok Man
    Jeon, Yunseok
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2016, 30 (06) : 771 - 782
  • [5] A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques
    Critchley, LAH
    Critchley, JAJH
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) : 85 - 91
  • [6] Assessment of Trending Ability of Cardiac Output Monitors by Polar Plot Methodology
    Critchley, Lester A.
    Yang, Xiao X.
    Lee, Anna
    [J]. JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2011, 25 (03) : 536 - 546
  • [7] Goal-directed Intraoperative therapy reduces morbidity and length of hospital stay in high-risk surgical patients
    Donati, Abele
    Loggi, Silvia
    Preiser, Jean-Charles
    Orsetti, Giovanni
    Muench, Cristopher
    Gabbanelli, Vincenzo
    Pelaia, Paolo
    Pietropaoli, Paolo
    [J]. CHEST, 2007, 132 (06) : 1817 - 1824
  • [8] Reliability of a new 4th generation FloTrac algorithm to track cardiac output changes in patients receiving phenylephrine
    Ji, Fuhai
    Li, Jian
    Fleming, Neal
    Rose, David
    Liu, Hong
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2015, 29 (04) : 467 - 473
  • [9] Early On-Cardiopulmonary Bypass Hypotension and Other Factors Associated With Vasoplegic Syndrome
    Levin, Matthew A.
    Lin, Hung-Mo
    Castillo, Javier G.
    Adams, David H.
    Reich, David L.
    Fischer, Gregory W.
    [J]. CIRCULATION, 2009, 120 (17) : 1664 - 1671
  • [10] Uncalibrated arterial pulse contour analysis versus continuous thermodilution technique: Effects of alterations in arterial waveform
    Lorsomradee, Suraphong
    Lorsomradee, Sratwadee
    Cromheecke, Stefanie
    De Hert, Stefan G.
    [J]. JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2007, 21 (05) : 636 - 643