Sub-millisecond time-resolved SAXS using a continuous-flow mixer and X-ray microbeam

被引:52
作者
Graceffa, Rita [1 ,2 ]
Nobrega, R. Paul [3 ]
Barrea, Raul A. [1 ,2 ]
Kathuria, Sagar V. [3 ]
Chakravarthy, Srinivas [1 ,2 ]
Bilsel, Osman [3 ]
Irving, Thomas C. [1 ,2 ]
机构
[1] IIT, CSRRI, BioCAT, Chicago, IL 60616 USA
[2] IIT, Dept BCS, Chicago, IL 60616 USA
[3] Univ Massachusetts, Sch Med, Dept Biochem & Mol Pharmacol, Worcester, MA 01605 USA
基金
美国国家卫生研究院;
关键词
micro-SAXS; time-resolved; protein folding; PROTEIN-FOLDING KINETICS; CYTOCHROME-C; EARLY EVENTS; MICROSECOND; SCATTERING; SPECTROSCOPY; LANDSCAPE;
D O I
10.1107/S0909049513021833
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Small-angle X-ray scattering (SAXS) is a well established technique to probe the nanoscale structure and interactions in soft matter. It allows one to study the structure of native particles in near physiological environments and to analyze structural changes in response to variations in external conditions. The combination of microfluidics and SAXS provides a powerful tool to investigate dynamic processes on a molecular level with sub-millisecond time resolution. Reaction kinetics in the sub-millisecond time range has been achieved using continuous-flow mixers manufactured using micromachining techniques. The time resolution of these devices has previously been limited, in part, by the X-ray beam sizes delivered by typical SAXS beamlines. These limitations can be overcome using optics to focus X-rays to the micrometer size range providing that beam divergence and photon flux suitable for performing SAXS experiments can be maintained. Such micro-SAXS in combination with microfluidic devices would be an attractive probe for time-resolved studies. Here, the development of a high-duty-cycle scanning microsecond-time-resolution SAXS capability, built around the Kirkpatrick-Baez mirror-based microbeam system at the Biophysics Collaborative Access Team (BioCAT) beamline 18ID at the Advanced Photon Source, Argonne National Laboratory, is reported. A detailed description of the microbeam small-angle-scattering instrument, the turbulent flow mixer, as well as the data acquisition and control and analysis software is provided. Results are presented where this apparatus was used to study the folding of cytochrome c. Future prospects for this technique are discussed.
引用
收藏
页码:820 / 825
页数:6
相关论文
共 31 条
  • [1] Conformational landscape of cytochrome c folding studied by microsecond-resolved small-angle x-ray scattering
    Akiyama, S
    Takahashi, S
    Kimura, T
    Ishimori, K
    Morishima, I
    Nishikawa, Y
    Fujisawa, T
    [J]. PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2002, 99 (03) : 1329 - 1334
  • [2] Microsecond hydrophobic collapse in the folding of Escherichia coli dihydrofolate reductase, an α/β-type protein
    Arai, Munehito
    Kondrashkina, Elena
    Kayatekin, Can
    Matthews, C. Robert
    Iwakura, Masahiro
    Bilsel, Osman
    [J]. JOURNAL OF MOLECULAR BIOLOGY, 2007, 368 (01) : 219 - 229
  • [3] Fast-scanning high-flux microprobe for biological X-ray fluorescence microscopy and microXAS
    Barrea, R. A.
    Gore, D.
    Kujala, N.
    Karanfil, C.
    Kozyrenko, S.
    Heurich, R.
    Vukonich, M.
    Huang, R.
    Paunesku, T.
    Woloschak, G.
    Irving, T. C.
    [J]. JOURNAL OF SYNCHROTRON RADIATION, 2010, 17 : 522 - 529
  • [4] BARREA RA, 2006, IPAP C SERIES, V7, P230
  • [5] A microchannel solution mixer for studying microsecond protein folding reactions
    Bilsel, O
    Kayatekin, C
    Wallace, LA
    Matthews, CR
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2005, 76 (01)
  • [6] Submillisecond protein folding kinetics studied by ultrarapid mixing
    Chan, CK
    Hu, Y
    Takahashi, S
    Rousseau, DL
    Eaton, WA
    Hofrichter, J
    [J]. PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1997, 94 (05) : 1779 - 1784
  • [7] Measurement of microsecond dynamic motion in the intestinal fatty acid binding protein by using fluorescence correlation spectroscopy
    Chattopadhyay, K
    Saffarian, S
    Elson, EL
    Frieden, C
    [J]. PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2002, 99 (22) : 14171 - 14176
  • [8] SecA folding kinetics: A large dimeric protein rapidly forms multiple native states
    Doyle, SM
    Bilsel, O
    Teschke, CM
    [J]. JOURNAL OF MOLECULAR BIOLOGY, 2004, 341 (01) : 199 - 214
  • [9] THE RADIUS OF GYRATION OF AN APOMYOGLOBIN FOLDING INTERMEDIATE
    ELIEZER, D
    JENNINGS, PA
    WRIGHT, PE
    DONIACH, S
    HODGSON, KO
    TSURUTA, H
    [J]. SCIENCE, 1995, 270 (5235) : 487 - 488
  • [10] Dynamically figured Kirkpatrick Baez x-ray micro-focusing optics
    Eng, PJ
    Newville, M
    Rivers, ML