On the Fundamental Limits of the Position Resolution of Position-Sensitive Scintillation Detector Using Continuous Crystal

被引:0
作者
Gal, J. [1 ]
Imrek, J. [1 ]
Kalinka, G. [1 ]
Kiraly, B. [1 ]
Nagy, F. [1 ]
Nyako, B. M. [1 ]
Molnar, J. [1 ]
Valastyan, I. [1 ]
Balkay, L. [2 ]
Kis, S. A. [2 ]
机构
[1] Hungarian Acad Sci, Inst Nucl Res, Bern Ter 18-C, Debrecen, Hungary
[2] Univ Debrecen, Inst Nucl Med, Debrecen, Hungary
来源
2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC) | 2011年
关键词
DEPTH; APD;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The factors determining the position resolution of continuous crystal based position sensitive scintillation gamma detectors are discussed. The key factors are the position functions which give the relation between the coordinates of the point of interaction (POI) and the responses of the photodetectors reading out of the scintillator. The physical phenomena limiting the position resolution are the statistical fluctuation of the detected photons and the electronic noise. For the standard deviation of the POI coordinates simple relations are deduced. It has been established that the contributions of both the statistical fluctuation and the noise to the standard deviation of the POI coordinates are inversely proportional to the square root of the detected photon numbers and the measure of these contributions can be characterised with the position resolution coefficients deduced from the position functions. A detailed discussion of the effect of the scintillator and photodetector characteristics is also given. As an example, calculations for the resolution coefficients in the case of the solid angle method with a chosen scintillator photodetector arrangement are given. The results show that the resolution coefficients as a function of the POI coordinates vary in a wide range. With realistic practical data and using almost the largest value for the resolution coefficients similar or better position resolution can be reached with continuous crystal than with a pixellated one.
引用
收藏
页码:3684 / 3689
页数:6
相关论文
共 24 条
  • [21] Design and performance evaluation of a 4π-view gamma camera with mosaic-patterned 3D position-sensitive scintillators
    Hu, Yifan
    Fan, Peng
    Lyu, Zhenlei
    Huang, Jianwei
    Wang, Shi
    Xia, Yan
    Liu, Yaqiang
    Ma, Tianyu
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2022, 1023
  • [22] Design of linear dynamic range and high sensitivity matrix quadrant APDs ROIC for position sensitive detector application
    Zheng, Hao
    Ma, Rui
    Zhu, Zhangming
    MICROELECTRONICS JOURNAL, 2017, 63 : 49 - 57
  • [23] Event Localization in Continuous Crystal Scintillation Cameras using Distribution Matching Neural Networks
    Rodriguez Colmeiro, Ramiro
    Verrastro, Claudio
    Minsky, Daniel
    Grosges, Thomas
    2021 ARGENTINE CONFERENCE ON ELECTRONICS (CAE 2021), 2021, : 7 - 12
  • [24] The "X'tal Cube" PET Detector: 3D Scintillation Photon Detection of a 3D Crystal Array Using MPPCs
    Yazaki, Yujiro
    Murayama, Hideo
    Inadama, Naoko
    Osada, Hiroto
    Nishikido, Fumihiko
    Shibuya, Kengo
    Yamaya, Taiga
    Yoshida, Eiji
    Suga, Mikio
    Moriya, Takahiro
    Watanabe, Mitsuo
    Yamashita, Takaji
    Kawai, Hideyuki
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3822 - +