A multicentre comparison of quantitative 90Y PET/CT for dosimetric purposes after radioembolization with resin microspheres

被引:111
作者
Willowson, Kathy P. [1 ]
Tapner, Michael [2 ]
Bailey, Dale L. [3 ,4 ]
机构
[1] Univ Sydney, Sch Phys, Inst Med Phys, Sydney, NSW 2006, Australia
[2] Sirtex, Sydney, NSW 2060, Australia
[3] Royal N Shore Hosp, Dept Nucl Med, Sydney, NSW 2065, Australia
[4] Univ Sydney, Fac Hlth Sci, Lidcombe 2141, Australia
基金
澳大利亚研究理事会;
关键词
Yttrium-90; PET/CT; Quantitative; Radioembolization; STANDARDIZED UPTAKE VALUES; IMAGE-RECONSTRUCTION; INTERNAL RADIATION; DOSE-RESPONSE; FDG-PET/CT; TOF PET; PERFORMANCE; BIODISTRIBUTION; QUANTIFICATION; RESOLUTION;
D O I
10.1007/s00259-015-3059-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To investigate and compare the quantitative accuracy of Y-90 imaging across different generation PET/CT scanners, for the purpose of dosimetry after radioembolization with resin microspheres. Methods A strict experimental and imaging protocol was followed by 47 international sites using the NEMA 2007/IEC 2008 PET body phantom with an 8-to-1 sphere-to-background ratio of Y-90 solution. The phantom was imaged over a 7-day period (activity ranging from 0.5 to 3.0 GBq) and all reconstructed data were analysed at a core laboratory for consistent processing. Quantitative accuracy was assessed through measures of total phantom activity, activity concentration in background and hot spheres, misplaced counts in a nonradioactive insert, and background variability. Results Of the 69 scanners assessed, 37 had both time-of-flight (ToF) and resolution recovery (RR) capability. These current generation scanners from GE, Philips and Siemens could reconstruct background concentration measures to within 10 % of true values over the evaluated range, with greater deviations on the Philips systems at low count rates, and demonstrated typical partial volume effects on hot sphere recovery, which dominated spheres of diameter < 20 mm. For spheres > 20 mm in diameter, activity concentrations were consistently underestimated by about 20 %. Non-ToF scanners from GE Healthcare and Siemens were capable of producing accurate measures, but with inferior quantitative recovery compared with ToF systems. Conclusion Current generation ToF scanners can consistently reconstruct Y-90 activity concentrations, but they underestimate activity concentrations in small structures (a parts per thousand currency sign37 mm diameter) within a warm background due to partial volume effects and constraints of the reconstruction algorithm. At the highest count rates investigated, measures of background concentration (about 300 kBq/ml) could be estimated on average to within 1 %, 5 % and 2 % for GE Healthcare (all-pass filter, RR + ToF), Philips (4i8s ToF) and Siemens (2i21s all-pass filter, RR + ToF) ToF systems, respectively. Over the range of activities investigated, comparable performance between GE Healthcare and Siemens ToF systems suggests suitability for quantitative analysis in a scenario analogous to that of postradioembolization imaging for treatment of liver cancer.
引用
收藏
页码:1202 / 1222
页数:21
相关论文
共 37 条
[1]   Quantitative and Qualitative Assessment of Yttrium-90 PET/CT Imaging [J].
Attarwala, Ali Asgar ;
Molina-Duran, Flavia ;
Buesing, Karen-Anett ;
Schoenberg, Stefan O. ;
Bailey, Dale L. ;
Willowson, Kathy ;
Glatting, Gerhard .
PLOS ONE, 2014, 9 (11)
[2]   90Y-PET for the assessment of microsphere biodistribution after selective internal radiotherapy [J].
Bagni, Oreste ;
D'Arienzo, Marco ;
Chiaramida, Paola ;
Chiacchiararelli, Laura ;
Cannas, Pina ;
D'Agostini, Antonio ;
Cianni, Roberto ;
Salvatori, Rita ;
Scopinaro, Francesco .
NUCLEAR MEDICINE COMMUNICATIONS, 2012, 33 (02) :198-204
[3]   Physical Performance of the new hybrid PET/CT Discovery-690 [J].
Bettinardi, V. ;
Presotto, L. ;
Rapisarda, E. ;
Picchio, M. ;
Gianolli, L. ;
Gilardi, M. C. .
MEDICAL PHYSICS, 2011, 38 (10) :5394-5411
[4]   FDG PET and PET/CT: EANM procedure guidelines for tumour PET imaging: version 1.0 [J].
Boellaard, Ronald ;
O'Doherty, Mike J. ;
Weber, Wolfgang A. ;
Mottaghy, Felix M. ;
Lonsdale, Markus N. ;
Stroobants, Sigrid G. ;
Oyen, Wim J. G. ;
Kotzerke, Joerg ;
Hoekstra, Otto S. ;
Pruim, Jan ;
Marsden, Paul K. ;
Tatsch, Klaus ;
Hoekstra, Corneline J. ;
Visser, Eric P. ;
Arends, Bertjan ;
Verzijlbergen, Fred J. ;
Zijlstra, Josee M. ;
Comans, Emile F. I. ;
Lammertsma, Adriaan A. ;
Paans, Anne M. ;
Willemsen, Antoon T. ;
Beyer, Thomas ;
Bockisch, Andreas ;
Schaefer-Prokop, Cornelia ;
Delbeke, Dominique ;
Baum, Richard P. ;
Chiti, Arturo ;
Krause, Bernd J. .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 (01) :181-200
[5]   Standards for PET Image Acquisition and Quantitative Data Analysis [J].
Boellaard, Ronald .
JOURNAL OF NUCLEAR MEDICINE, 2009, 50 :11S-20S
[6]   The Netherlands protocol for standardisation and quantification of FDG whole body PET studies in multi-centre trials [J].
Boellaard, Ronald ;
Oyen, Wim J. G. ;
Hoekstra, Corneline J. ;
Hoekstra, Otto S. ;
Visser, Eric P. ;
Willemsen, Antoon T. ;
Arends, Bertjan ;
Verzijlbergen, Fred J. ;
Zijlstra, Josee ;
Paans, Anne M. ;
Comans, Emile F. I. ;
Pruim, Jan .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2008, 35 (12) :2320-2333
[7]   EARLY DOSE RESPONSE TO YTTRIUM-90 MICROSPHERE TREATMENT OF METASTATIC LIVER CANCER BY A PATIENT-SPECIFIC METHOD USING SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY AND POSITRON EMISSION TOMOGRAPHY [J].
Campbell, Janice M. ;
Wong, C. Oliver ;
Muzik, Otto ;
Marples, Brian ;
Joiner, Michael ;
Burmeister, Jay .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 74 (01) :313-320
[8]   Assessment of acquisition protocols for routine imaging of Y-90 using PET/CT [J].
Carlier, Thomas ;
Eugene, Thomas ;
Bodet-Milin, Caroline ;
Garin, Etienne ;
Ansquer, Catherine ;
Rousseau, Caroline ;
Ferrer, Ludovic ;
Barbet, Jacques ;
Schoenahl, Frederic ;
Kraeber-Bodere, Francoise .
EJNMMI RESEARCH, 2013, 3 :1-12
[9]   Treatment Modification of Yttrium-90 Radioembolization Based on Quantitative Positron Emission Tomography/CT Imaging [J].
Chang, Ted T. ;
Bourgeois, Austin C. ;
Balius, Anastasia M. ;
Pasciak, Alexander S. .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2013, 24 (03) :333-337
[10]   Effect of randoms on signal-to-noise ratio in TOF PET [J].
Conti, Maurizio .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (03) :1188-1193