Detection of glycated hemoglobin (HbA1c) based on impedance measurement with parallel electrodes integrated into a microfluidic device

被引:37
作者
Chuang, Ya-Chun
Lan, Kung-Chieh
Hsieh, Ku-Meng
Jang, Ling-Sheng
Chen, Ming-Kun
机构
[1] Natl Cheng Kung Univ, Dept Elect Engn, Tainan 70101, Taiwan
[2] Natl Cheng Kung Univ, Ctr Micro Nano Sci & Technol, Tainan 70101, Taiwan
关键词
Glycated hemoglobin; Microfluidics; Self-assembled monolayer (SAM); HbA1c biosensor; Boronic affinity; SELF-ASSEMBLED MONOLAYER; THIOCTIC ACID; GOLD SURFACE; DOUBLE-LAYER; BINDING; PROTEIN; RECOGNITION; BIOSENSOR;
D O I
10.1016/j.snb.2012.06.084
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper proposes a sensor for glycated hemoglobin (HbA1c) detection based on impedance measurement. The sensor contains a pair of parallel facing electrodes integrated into a microfluidic device. A pair of parallel facing electrodes is fabricated to characterize the impedance change and determine the concentration of HbA(1c). A self-assembled monolayer (SAM) of thiophene-3-boronic acid (T3BA) is used for modifying the gold electrodes. Non-faradaic electrochemical detection of surface-bound HbA(1c) with no redox indicator added is conducted by utilizing the capacitive property as a sensing signal. The performance of the HbA(1c) sensor was investigated by impedance analyzer and the impedance changes approximately logarithmically with the concentration of HbA1c in the range of 10-100 ng/mu L.The proposed detection method has potential for use in point-of-care diagnostics and can be used for the detection of various (c) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:1222 / 1230
页数:9
相关论文
共 31 条
[1]   FURTHER IDENTIFICATION OF NATURE AND LINKAGE OF CARBOHYDRATE IN HEMOGLOBIN-A1C [J].
BUNN, HF ;
HANEY, DN ;
GABBAY, KH ;
GALLOP, PM .
BIOCHEMICAL AND BIOPHYSICAL RESEARCH COMMUNICATIONS, 1975, 67 (01) :103-109
[2]   Protein adsorption on titanium dioxide: Effects on double layer and semiconductor space charge region studied by EIS [J].
Camara, Osvaldo R. ;
Avalle, Lucia B. ;
Oliva, Fabiana Y. .
ELECTROCHIMICA ACTA, 2010, 55 (15) :4519-4528
[3]  
Chandalia HB, 2002, CURR SCI INDIA, V83, P1522
[4]   Label-free impedance biosensors: Opportunities and challenges [J].
Daniels, Jonathan S. ;
Pourmand, Nader .
ELECTROANALYSIS, 2007, 19 (12) :1239-1257
[5]   A NEW METHOD FOR THE DETERMINATION OF THE PERMITTIVITY OF SMALL SAMPLES IN THE MICROWAVE RANGE AND ITS APPLICATION TO HEMOGLOBIN SINGLE-CRYSTALS [J].
DASELER, W ;
STEINHOFF, HJ ;
REDHARDT, A .
JOURNAL OF BIOCHEMICAL AND BIOPHYSICAL METHODS, 1991, 22 (01) :69-82
[6]   Simultaneous determination of glycated haemoglobin, a long term biomarker of diabetes mellitus, and total haemoglobin by isotope dilution and HPLC-ICP-MS [J].
del Castillo Busto, M. Estela ;
Montes-Bayon, Maria ;
Anon, Elena ;
Sanz-Medel, Alfredo .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2008, 23 (05) :758-764
[7]   Formation and electrochemical characterization of self-assembled monolayers of thioctic acid on polycrystalline gold electrodes in phosphate buffer pH 7.4 [J].
Dijksma, M ;
Kamp, B ;
Hoogvliet, JC ;
van Bennekom, WP .
LANGMUIR, 2000, 16 (08) :3852-3857
[8]   Effect of hexacyanoferrate(II/III) on self-assembled monolayers of thioctic acid and 11-mercaptoundecanoic acid on gold [J].
Dijksma, M ;
Boukamp, BA ;
Kamp, B ;
van Bennekom, WP .
LANGMUIR, 2002, 18 (08) :3105-3112
[9]   A single-use, disposable iridium-modified electrochemical biosensor for fructosyl valine for the glycoslated hemoglobin detection [J].
Fang, Lei ;
Li, Wei ;
Zhou, Yan ;
Liu, Chung-Chiun .
SENSORS AND ACTUATORS B-CHEMICAL, 2009, 137 (01) :235-238
[10]   Signal amplification in Immunoassays using Labeling via boronic acid binding to the sugar moiety of immunoglobulin G:: Proof of concept for glycated hemoglobin [J].
Halamek, J. ;
Wollenberger, U. ;
Stoecklein, W. F. M. ;
Warsinke, A. ;
Scheller, F. W. .
ANALYTICAL LETTERS, 2007, 40 (07) :1434-1444