High prevalence of insulation failure with externalized cables in St. Jude Medical Riata family ICD leads: Fluoroscopic grading scale and correlation to extracted leads

被引:65
作者
Parvathaneni, Sunthosh V. [1 ]
Ellis, Christopher R. [1 ]
Rottman, Jeffrey N. [1 ]
机构
[1] Vanderbilt Univ, Med Ctr, Tennessee Valley Hlth Syst Vet Adm, Sch Med,MCE, Nashville, TN 37232 USA
关键词
Implantable cardioverter-defibrillator; defibrillation lead; Lead failure; Insulation failure; Fluoroscopy; Riata; Riata ST; Durata; Electrical failure; Mechanical failure; Perforation; OF-THE-LITERATURE; DEFIBRILLATOR LEAD;
D O I
10.1016/j.hrthm.2012.03.049
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Inside-out abrasion with externalization of sensing ring or high-voltage cables in St Jude Medical Riata implantable cardioverter-defibrillator leads has been reported. The prevalence of extruded cables, rate of electrical abnormalities, and predictors of failure in Riata leads are unknown. OBJECTIVES To estimate the incidence of lead failure in the St Jude Medical Riata implantable cardioverter-defibrillator leads and to propose a standard for the fluoroscopic assessment of insulation breakdown. METHODS Patients undergoing cine-fluoroscopy on Riata implantable cardioverter-defibrillator leads at our institution before January 25, 2012, were included (n = 87). Leads were graded as types 0-3 (0 = normal, 1 = abnormal conductor spacing, 2 <= 1 cm cable extrusion, 3 = >1 cm length extrusion). Comparison to extracted leads (n = 15) was documented. Device interrogation data were used for electrical analysis. RESULTS The mean time from implant was 5.9 +/- 3.45 years. Structural lead failure with externalized cables was seen in 33.3% (29 of 87) of the patients. Thirty-one percent (9 of 29) of the leads with exposed cables showed electrical failure, and 29.7% (19 of 64) of the leads with normal electrical data contained externalized cables. Time from implant >= 5 years predicted structural lead failure (P < 0.05). X-ray grade compared with extracted leads demonstrated a sensitivity and specificity of 86% and 100%, respectively. CONCLUSIONS Cine-fluoroscopy using a simple scale correlated with the structural integrity of extracted Riata leads. A high percentage of leads with extrusion showed electrical failure. Leads >= 5 years from implant showed a high rate of externalized cables. A large independent multicenter study to determine the prevalence and clinical sequelae of Riata lead failures is warranted.
引用
收藏
页码:1218 / 1224
页数:7
相关论文
共 9 条
  • [1] Timing of delayed perforation with the St. Jude Riata lead: A single-center experience and a review of the literature
    Danik, Stephan B.
    Mansour, Moussa
    Heist, E. Kevin
    Ellinor, Patrick
    Milan, David
    Singh, Jagmeet
    Das, Saumya
    Reddy, Vivek
    D'Avila, Andre
    Ruskin, Jeremy N.
    Mela, Theofanie
    [J]. HEART RHYTHM, 2008, 5 (12) : 1667 - 1672
  • [2] Increased rate of subacute lead complications with small-caliber implantable cardioverter-defibrillator leads
    Eltis, Christopher R.
    Rottman, Jeffrey N.
    [J]. HEART RHYTHM, 2009, 6 (05) : 619 - 624
  • [3] Performance of the St. Jude Medical Riata Leads
    Epstein, Andrew E.
    Baker, James H.
    Beau, Scott L.
    Deering, Thomas F.
    Greenberg, Steven M.
    Goldman, Daniel S.
    [J]. HEART RHYTHM, 2009, 6 (02) : 204 - 209
  • [4] Insulation Defects of Thin High-Voltage ICD Leads: An Underestimated Problem?
    Erkapic, Damir
    Duray, Gabor Z.
    Bauernfeind, Tamas
    De Rosa, Salvatore
    Hohnloser, Stefan H.
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2011, 22 (09) : 1018 - 1022
  • [5] Early failure of a small-diameter high-voltage implantable cardioverter-defibrillator lead
    Hauser, Robert G.
    Kallinen, Linda M.
    Almquist, Adrian K.
    Gornick, Charles C.
    Katsiyiannis, William T.
    [J]. HEART RHYTHM, 2007, 4 (07) : 892 - 896
  • [6] Kodoth V, 2011, EUR SOC CARDIOL
  • [7] Premature Failure of a Riata Defibrillator Lead Without Impedance Change or Inappropriate Sensing: A Case Report and Review of the Literature
    Krebsbach, Angela
    Alhumaid, Fawaz
    Henrikson, Charles A.
    Calkins, Hugh
    Berger, Ronald D.
    Cheng, Alan
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2011, 22 (09) : 1070 - 1072
  • [8] Clinical Performance of the St. Jude Medical Riata Defibrillation Lead in a Large Patient Population
    Porterfield, James G.
    Porterfield, Linda M.
    Kuck, Karl H.
    Corbisiero, Raffaele
    Greenberg, Steven M.
    Hindricks, Gerhard
    Wazni, Oussama
    Beau, Scott L.
    Herre, John M.
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2010, 21 (05) : 551 - 556
  • [9] t. Jude Medical Riata, PROD PERF REP NOV 20