Effect of impact assembly on the fretting corrosion of modular hip tapers

被引:95
作者
Mroczkowski, ML [1 ]
Hertzler, JS [1 ]
Humphrey, SM [1 ]
Johnson, T [1 ]
Blanchard, CR [1 ]
机构
[1] Zimmer Corp Res, Warsaw, IN 46580 USA
关键词
fretting; corrosion; titanium alloy; cobalt alloy;
D O I
10.1002/jor.20048
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The goal of this study was to determine the effect of assembly load and local assembly environmental conditions on the fretting corrosion of modular femoral stem tapers. Femoral head/taper assemblies in both similar (CoCrMo/CoCrMo) and mixed (CoCrMo/Ti-6A1-4V) alloy combinations were evaluated using an electrochemical test method. Specimens were assembled under impact loading and by hand, in both wet and dry conditions. Incremental cyclic loads ranging from 89 to 5,340 N were applied at a frequency of 3 Hz in Ringer's solution at ambient temperature. During the test, both the open circuit potential (OCP) and fretting current (i(fret)) were measured using a saturated calomel electrode (SCE) and counter electrode, respectively. The results were comparable for both mixed and similar alloy couples. Decreases in OCP and increases in i(fret), (indicators of oxide film fracture and repassivation) were seen with increasing load magnitude, often occurring at loads well below those expected clinically. OCP at the 5,340 N cyclic load ranged from - 30.4 to - 103.7 mV versus SCE for similar alloy couples, and - 19.1 to - 181.4 mV versus SCE for mixed alloy couples. Mean peak fretting currents ranged from 0.84 to 1.42 pA and 1.06 to 3.12 J.LA for similar and mixed alloy couples, respectively. The larger current magnitudes and more negative shifts in OCP for mixed alloy couples indicate the difference in oxide film fracture behavior between titanium and cobalt alloys. The load at which OCP began to drop (onset of fretting) was dependent upon the assembly conditions for both material couples. Specimens assembled with impact loads in air showed the highest resistance to fretting. The results of this study indicate that the assembly load and the environment both play a role in the initial stability of modular hip taper connections. (c) 2005 Orthopaedic Research Society. Published by Wiley Periodicals, Inc.
引用
收藏
页码:271 / 279
页数:9
相关论文
共 22 条
[1]   Hip contact forces and gait patterns from routine activities [J].
Bergmann, G ;
Deuretzbacher, G ;
Heller, M ;
Graichen, F ;
Rohlmann, A ;
Strauss, J ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :859-871
[2]   Dissociation of modular humeral head components: A biomechanical and implant retrieval study [J].
Blevins, FT ;
Deng, XH ;
Torzilli, PA ;
Dines, D ;
Warren, RF .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 1997, 6 (02) :113-124
[3]   FRETTING CORROSION IN SALINE AND SERUM [J].
BROWN, SA ;
MERRITT, K .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1981, 15 (04) :479-488
[4]  
COLLIER JP, 1991, CLIN ORTHOP RELAT R, P305
[5]  
Gilbert J.L., 1996, MEDICAL APPL TITANIU, V1272, P199
[6]   IN-VIVO CORROSION OF MODULAR HIP-PROSTHESIS COMPONENTS IN MIXED AND SIMILAR METAL COMBINATIONS - THE EFFECT OF CREVICE, STRESS, MOTION, AND ALLOY COUPLING [J].
GILBERT, JL ;
BUCKLEY, CA ;
JACOBS, JJ .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1993, 27 (12) :1533-1544
[7]   The mechanical and electrochemical processes associated with taper fretting crevice corrosion: A review [J].
Gilbert, JL ;
Jacobs, JJ .
MODULARITY OF ORTHOPEDIC IMPLANTS, 1997, 1301 :45-59
[8]   INTERGRANULAR CORROSION-FATIGUE FAILURE OF COBALT-ALLOY FEMORAL STEMS - A FAILURE ANALYSIS OF 2 IMPLANTS [J].
GILBERT, JL ;
BUCKLEY, CA ;
JACOBS, JJ ;
BERTIN, KC ;
ZERNICH, MR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1994, 76A (01) :110-115
[9]   In vitro corrosion testing of modular hip tapers [J].
Goldberg, JR ;
Gilbert, JL .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2003, 64B (02) :78-93
[10]   The electrochemical and mechanical behavior of passivated and TiN/AlN-coated CoCrMo and Ti6Al4V alloys [J].
Goldberg, JR ;
Gilbert, JL .
BIOMATERIALS, 2004, 25 (05) :851-864