Quality of myocardial perfusion single-photon emission tomography imaging: multicentre evaluation with a cardiac phantom

被引:8
作者
Heikkinen, J
Ahonen, A
Kuikka, JT
Rautio, P
机构
[1] Mikkeli Cent Hosp, Etela Savo Hosp Dist, Dept Nucl Med, FIN-50100 Mikkeli, Finland
[2] Oulu Univ Hosp, Dept Nucl Med, FIN-90220 Oulu, Finland
[3] Kuopio Univ Hosp, Dept Clin Physiol, FIN-70210 Kuopio, Finland
[4] Niuvanniemi Hosp, FIN-70210 Kuopio, Finland
[5] N Karelia Cent Hosp, Dept Clin Physiol, FIN-80210 Joensuu, Finland
关键词
myocardial perfusion; single-photon emission tomography; external quality assurance;
D O I
10.1007/s002590050585
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of the study was to evaluate quality of myocardial perfusion single-photon emission tomography (SPET) imaging in Finnish hospitals. Nineteen nuclear medicine departments participated in the study. A myocardial phantom simulating clinical stress and rest conditions was filled with routinely used isotope solution (technetium-99m or thallium-201). The cardiac insert included three reversible defects (simulating ischaemia): 30x30x14 mm(3) septal (90% recovery at rest), 30x20x14 mm(3) posterobasal (full recovery) and 20x20x14 mm(3) lateral (full recovery). There were two fixed defects (simulating infarct): 30x20x14 mm(3) postero-apical and 10x10x6 mm(3) apical. The phantom was imaged and interpreted as a myocardial perfusion patient. Reconstruction, printout and reporting were performed according to the clinical routine of each centre. Three nuclear medicine specialists anonymously evaluated the quality of the image sets. The visual scores of the experts were ranked from 1 to 5. Additionally, points from 0 to 8 were given to research reports according to how well perfusion defects were detected. Quantitative points were calculated by comparing background-subtracted and -normalized counts from 12 regions of interest between stress and rest images. Results for technetium studies (12 departments) were better than those for thallium (7 departments), The average visual scores of the experts were 3.7+/-0.9 for all image sets, 3.2+/-0.5 for thallium users and 3.9+/-0.6 for technetium users (P=0.003). Five laboratories received a low score which, according to the specialists, is barely sufficient for limited clinical use. Average points for the reports were 5.6+/-2.1, 4.9+/-1.5 and 6.5+/-1.7 (P=0.051), and for the quantitation 8.2+/-1.0, 7.9+/-0.4 and 8.4+/-1.1 (P=0.185), respectively. Seven out of 22 interpreters did not detect the lateral 20x20x14 mm(3) defect; five of them used thallium. This study demonstrated the heterogeneity of myocardial perfusion SPET in Finland. The participating laboratories used a wide scale of methods and, sometimes, inappropriate imaging protocols. The need for quality assurance in nuclear cardiology, correct use of SPET instrumentation and objective comparison of clinical studies is evident. The method described is suitable for external quality assurance and quality improvement of myocardial SPET imaging, and is recommended for regular use in nuclear medicine.
引用
收藏
页码:1289 / 1297
页数:9
相关论文
共 19 条
[1]   QUALITY ASSURANCE AND HARMONIZATION OF NUCLEAR-MEDICINE INVESTIGATIONS IN EUROPE [J].
BERGMANN, H ;
BUSEMANNSOKOLE, E ;
HORTON, PW .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1995, 22 (05) :477-480
[2]  
BONOW RO, 1992, J NUCL MED, V33, P1434
[3]   PROTOCOLS FOR SELECTION OF CARDIAC RADIONUCLIDE STUDIES FOR USE AS A DATA-BASE OF NORMAL STUDIES AND TYPICAL PATTERNS OF DISEASES - COST-B2-WORKING-GROUP-II, IN ASSOCIATION WITH RELEVANT WORKING TASK GROUPS OF THE EUROPEAN-ASSOCIATION-OF-NUCLEAR-MEDICINE AND THE EUROPEAN-SOCIETY-OF-CARDIOLOGY [J].
BOURGUIGNON, MH ;
SOKOLE, EB ;
JONES, B ;
VANDERWALL, E .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1993, 20 (01) :59-65
[4]  
EISNER RL, 1986, J NUCL MED, V27, P1717
[5]   INDUSTRIAL PERSPECTIVE - FROM IND TO NDA [J].
GIERING, LP ;
SMITH, JE .
JOURNAL OF NUCLEAR CARDIOLOGY, 1995, 2 (01) :66-70
[6]   A FINNISH MULTICENTER QUALITY ASSURANCE PROJECT IN BONE-SCINTIGRAPHY AND BRAIN SPET - A PHANTOM STUDY [J].
HEIKKINEN, J ;
KUIKKA, JT ;
AHONEN, A ;
KURVELIN, J ;
HARTIKAINEN, K ;
KVIST, G .
NUCLEAR MEDICINE COMMUNICATIONS, 1994, 15 (10) :795-805
[7]   Quality of brain perfusion single-photon emission tomography images: multicentre evaluation using an anatomically accurate three-dimensional phantom [J].
Heikkinen, J ;
Kuikka, JT ;
Ahonen, A ;
Rautio, P .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1998, 25 (10) :1415-1422
[8]  
HEIKKINEN JO, 1999, KUOPIO U PUBLICATI C, V89, P25
[9]   What is the current status of quantification and nuclear medicine in cardiology? [J].
Hor, G .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1996, 23 (07) :815-851
[10]  
LEPPO JA, 1991, J NUCL MED, V32, P2012