Method Validation of 16 Types of Structural Amino Acids using an Automated Amino Acid Analyzer

被引:31
|
作者
Shim, You-Shin [1 ]
Yoon, Won-Jin [1 ,2 ]
Ha, Jaeho [1 ]
Seo, Dongwon [1 ]
Lee, Kwang-Won [2 ]
Lee, Woo-Young [3 ]
Kwon, Kwang-Il [4 ]
Kang, Tae-Seok [5 ]
Lee, Jin-Ha [5 ]
Kim, Hyun-Jeong [6 ]
Kwak, Hae-Jung [6 ]
Lee, Sam-Pin [6 ]
Kim, So-Jung [7 ]
Yun, Won-Kap [7 ]
Lee, Junsoo [8 ]
Hwang, Jin-Bong [1 ]
机构
[1] Korea Food Res Inst, Food Anal Ctr, Gyeonggi 463746, South Korea
[2] Korea Univ, Coll Life Sci & Biotechnol, Seoul 136701, South Korea
[3] Nutr Safety Policy Div, Food Nutr & Dietary Safety Bur, Minist Food & Drug safety, Chungbuk 363700, South Korea
[4] Dietary Life Safety Div, Food Nutr & Dietary Safety Bur, Minist Food & Drug safety, Chungbuk 363700, South Korea
[5] Natl Inst Food & Drug safety Evaluat, Food safety Evaluat Dept, Chungbuk 363700, South Korea
[6] Keimyung Univ, Ctr Tradit Microorganism Resource, Taegu 704701, South Korea
[7] Gyeongbuk Inst Marine Bioind, Gyeongbuk 767813, South Korea
[8] Chungbuk Natl Univ, Dept Food Sci & Technol, Chungbuk 361763, South Korea
关键词
amino acids; automated amino acid analyzer; analytical method; validation; post column reaction; CHROMATOGRAPHY-MASS-SPECTROMETRY; PRECOLUMN DERIVATIZATION; PERFORMANCE; DERIVATIVES; MS/MS; FLUIDS; SERUM;
D O I
10.1007/s10068-013-0252-0
中图分类号
TS2 [食品工业];
学科分类号
0832 ;
摘要
An analytical method for the simultaneous determination of 16 types of structural amino acids in infant formula using an automated amino acid analyzer was validated through tests of precision, accuracy and linearity. The automated analysis of the amino acids was performed on an ion-exchange packed column with a visible detector. The certificated reference material (CRM) 1546 from NIST was used as the test sample to determine the precision and accuracy of the analytical method. The regression analyses revealed good correlations [correlation coefficient (r(2))] that were greater than 0.99. The recovery values of the amino acids were ranged from 87.18 to 118.08%. The limits of detection and limits of quantification were less than 0.059 mg/100 g and less than 0.198 mg/100 g, respectively. The infra- and inter-day precisions were less than 14.62% in the CRM sample.