Repeatability and reproducibility of corneal thickness using SOCT Copernicus HR

被引:18
作者
Vidal, Silvia [1 ]
Viqueira, Valentin [1 ]
Mas, David [1 ]
Domenech, Begona [1 ]
机构
[1] Univ Alicante, Dept Opt, E-03080 Alicante, Spain
关键词
cornea; corneal thickness; Fourier domain; optical coherence tomography; pachymetry; OPTICAL COHERENCE TOMOGRAPHY; TIME-DOMAIN; PACHYMETRY; AGREEMENT;
D O I
10.1111/cxo.12002
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose The aim of this study is to determine the reliability of corneal thickness measurements derived from SOCT Copernicus HR (Fourier domain OCT). Methods Thirty healthy eyes of 30 subjects were evaluated. One eye of each patient was chosen randomly. Images were obtained of the central (up to 2.0mm from the corneal apex) and paracentral (2.0 to 4.0mm) cornea. We assessed corneal thickness (central and paracentral) and epithelium thickness. The intra-observer repeatability data were analysed using the intra-class correlation coefficient (ICC) for a range of 95 per cent within-subject standard deviation (SW) and the within-subject coefficient of variation (CW). The level of agreement by BlandAltman analysis was also represented for the study of the reproducibility between observers and agreement between methods of measurement (automatic versus manual). Results The mean value of the central corneal thickness (CCT) was 542.4 +/- 30.1m (SD). There was a high intra-observer agreement, finding the best result in the central sector with an intra-class correlation coefficient of 0.99, 95 per cent CI (0.989 to 0.997) and the worst, in the minimum corneal thickness, with an intra-class correlation coefficient of 0.672, 95 per cent CI (0.417 to 0.829). Reproducibility between observers was very high. The best result was found in the central sector thickness obtained both manually and automatically with an intra-class correlation coefficient of 0.990 in both cases and the worst result in the maximum corneal thickness with an intra-class correlation coefficient of 0.827. The agreement between measurement methods was also very high with intra-class correlation coefficient greater than 0.91. On the other hand the repeatability and reproducibility for epithelial measurements was poor. Conclusion Pachymetric mapping with SOCT Copernicus HR was found to be highly repeatable and reproducible. We found that the device lacks an appropriate ergonomic design as proper focusing of the laser beam onto the cornea for anterior segment scanning required that patients were positioned slightly farther away from the machine head-rest than in the setup for retinal imaging.
引用
收藏
页码:278 / 285
页数:8
相关论文
共 19 条
[1]   Reproducibility-Repeatability of Choroidal Thickness Calculation Using Optical Coherence Tomography [J].
Benavente-Perez, Alexandra ;
Hosking, Sarah L. ;
Logan, Nicola S. ;
Bansal, Dheeraj .
OPTOMETRY AND VISION SCIENCE, 2010, 87 (11) :867-872
[2]   Value of optical coherence tomography for anterior segment surgery [J].
Doors, Muriel ;
Berendschot, Tos T. J. M. ;
de Brabander, John ;
Webers, Carroll A. B. ;
Nuijts, Rudy M. M. A. .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (07) :1213-1229
[3]  
FERMANIAN J, 1984, REV EPIDEMIOL SANTE, V32, P408
[4]   An evidence-based assessment of risk factors for the progression of ocular hypertension and glaucoma [J].
Friedman, DS ;
Wilson, MR ;
Liebmann, JM ;
Fechtner, RD ;
Weinreb, RN .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2004, 138 (03) :S19-S31
[5]   Repeatability and reproducibility of anterior ocular biometric measurements with 2-dimensional and 3-dimensional optical coherence tomography [J].
Fukuda, Shinichi ;
Kawana, Keisuke ;
Yasuno, Yoshiaki ;
Oshika, Tetsuro .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (11) :1867-1873
[6]   The role of blood pressure in glaucoma [J].
He, Zheng ;
Vingrys, Algis J. ;
Armitage, James A. ;
Bui, Bang V. .
CLINICAL AND EXPERIMENTAL OPTOMETRY, 2011, 94 (02) :133-149
[7]   Intra-examiner repeatability and agreement of corneal pachymetry map measurement by time-domain and Fourier-domain optical coherence tomography [J].
Huang, Jehn-Yu ;
Pekmezci, Melike ;
Yaplee, Stephanie ;
Lin, Shan .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2010, 248 (11) :1647-1656
[8]   A Comprehensive Comparison of Central Corneal Thickness Measurement [J].
Huang, Jinhai ;
Pesudovs, Konrad ;
Yu, Ayong ;
Wright, Thomas ;
Wen, Daizong ;
Li, Ming ;
Yu, Ye ;
Wang, Qinmei .
OPTOMETRY AND VISION SCIENCE, 2011, 88 (08) :940-949
[9]   Comparison of central corneal thickness using anterior segment optical coherence tomography vs ultrasound pachymetry [J].
Kim, Hanna Y. ;
Budenz, Donald L. ;
Lee, Pak S. ;
Feuer, William J. ;
Barton, Keith .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2008, 145 (02) :228-232
[10]   Comparative study of central corneal thickness measurement with slit-lamp optical coherence tomography and visante optical coherence tomography [J].
Li, Haitao ;
Leung, Christopher Kai Shun ;
Wong, Lee ;
Cheung, Carol Yim Lui ;
Pang, Chi Pui ;
Weinreb, Robert Neal ;
Lam, Dennis Shun Chiu .
OPHTHALMOLOGY, 2008, 115 (05) :796-801