Investigating the Effect of Column Geometry on Separation Efficiency using 3D Printed Liquid Chromatographic Columns Containing Polymer Monolithic Phases

被引:46
作者
Gupta, Vipul [1 ,2 ]
Beirne, Stephen [3 ]
Nesterenko, Pavel N. [1 ]
Paull, Brett [1 ,2 ]
机构
[1] Univ Tasmania, Sch Phys Sci, ACROSS, Hobart, Tas 7001, Australia
[2] Univ Tasmania, Sch Phys Sci, ARC Ctr Excellence Electromat Sci, Hobart, Tas 7001, Australia
[3] Univ Wollongong, Intelligent Polymer Res Inst, ARC Ctr Excellence Electromat Sci, Wollongong, NSW 2522, Australia
基金
澳大利亚研究理事会;
关键词
SUPERCRITICAL-FLUID CHROMATOGRAPHY; DEFORMED OPEN TUBES; GAS-CHROMATOGRAPHY; MASS-TRANSFER; STATIONARY PHASES; HPLC; PERFORMANCE; RESOLUTION; METAL; IDEAL;
D O I
10.1021/acs.analchem.7b03778
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Effect of column geometry on the liquid chromatographic separations using 3D printed liquid chromatographic columns with in-column polymerized monoliths has been studied. Three different liquid chromatographic columns were designed and 3D printed in titanium as 2D serpentine, 3D spiral, and 3D serpentine columns, of equal length and i.d. Successful in-column thermal polymerization of mechanically stable poly(BuMA-co-EDMA) monoliths was achieved within each design without any significant structural differences between phases. Van Deemter plots indicated higher efficiencies for the 3D serpentine chromatographic columns with higher aspect ratio turns at higher linear velocities and smaller analysis times as compared to their counterpart columns with lower aspect ratio turns. Computational fluid dynamic simulations of a basic monolithic structure indicated 44%, 90%, 100%, and 118% higher flow through narrow channels in the curved monolithic configuration as compared to the straight monolithic configuration at linear velocities of 1, 2.5, 5, and 10 mm s-1, respectively. Isocratic RPLC separations with the 3D serpentine column resulted in an average 23% and 245% (8 solutes) increase in the number of theoretical plates as compared to the 3D spiral and 2D serpentine columns, respectively. Gradient RPLC separations with the 3D serpentine column resulted in an average 15% and 82% (8 solutes) increase in the peak capacity as compared to the 3D spiral and 2D serpentine columns, respectively. Use of the 3D serpentine column at a higher flow rate, as compared to the 3D spiral column, provided a 58% reduction in the analysis time and 74% increase in the peak capacity for the isocratic separations of the small molecules and the gradient separations of proteins, respectively.
引用
收藏
页码:1186 / 1194
页数:9
相关论文
共 33 条
[1]   Fabrication and preliminary results for LiGA fabricated nickel micro gas chromatograph columns [J].
Bhushan, Abhinav ;
Yemane, Dawit ;
Overton, Edward B. ;
Goettert, Jost ;
Murphy, Michael C. .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2007, 16 (02) :383-393
[2]   Study of bed homogenity of methacrylate-based monolithic columns for micro-HPLC and CEC [J].
Buszewski, B ;
Szumski, M .
CHROMATOGRAPHIA, 2004, 60 (Suppl 1) :S261-S267
[3]   Silicon components for gas chromatograph [J].
Dziuban, J ;
Górecka-Drzazga, A ;
Malecki, K .
OPTOELECTRONIC AND ELCTRONIC SENSORS IV, 2001, 4516 :249-257
[4]   Tailoring the morphology of methacrylate ester-based monoliths for optimum efficiency in liquid chromatography [J].
Eeltink, S ;
Herrero-Martinez, JM ;
Rozing, GP ;
Schoenmakers, PJ ;
Kok, WT .
ANALYTICAL CHEMISTRY, 2005, 77 (22) :7342-7347
[5]   REACTION DETECTOR WITH 3-DIMENSIONAL COILED OPEN TUBES IN HPLC [J].
ENGELHARDT, H ;
NEUE, UD .
CHROMATOGRAPHIA, 1982, 15 (07) :403-408
[6]  
GIDDINGS JC, 1960, J CHROMATOGR, V3, P520
[7]   NATURE OF GAS PHASE MASS TRANSFER IN GAS CHROMATOGRAPHY [J].
GIDDINGS, JC .
ANALYTICAL CHEMISTRY, 1962, 34 (10) :1186-&
[8]   Monolithic columns in high-performance liquid chromatography [J].
Guiochon, Georges .
JOURNAL OF CHROMATOGRAPHY A, 2007, 1168 (1-2) :101-168
[9]   3D printed titanium micro-bore columns containing polymer monoliths for reversed-phase liquid chromatography [J].
Gupta, Vipul ;
Talebi, Mohammad ;
Deverell, Jeremy ;
Sandron, Sara ;
Nesterenko, Pavel N. ;
Heery, Brendan ;
Thompson, Fletcher ;
Beirne, Stephen ;
Wallace, Gordon G. ;
Paull, Brett .
ANALYTICA CHIMICA ACTA, 2016, 910 :84-94
[10]   MASS-TRANSFER IN IDEAL AND GEOMETRICALLY DEFORMED OPEN TUBES .3. DEFORMED METAL AND PLASTIC TUBES [J].
HOFMANN, K ;
HALASZ, I .
JOURNAL OF CHROMATOGRAPHY, 1980, 199 (OCT) :3-22