Comparison of Electroanatomical Mapping Systems: Accuracy in Left Atrial Mapping

被引:3
作者
Heist, E. Kevin
Perna, Francesco
Chalhoub, Fadi
Danik, Stephan
Barrett, Conor
Houghtaling, Christopher
Tondo, Claudio
Mahapatra, Srijoy
Ruskin, Jeremy
Mansour, Moussa [1 ]
机构
[1] Massachusetts Gen Hosp, Cardiac Arrhythmia Serv, Boston, MA 02114 USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2013年 / 36卷 / 05期
关键词
ablation; atrial fibrillation; electrophysiologyclinical; mapping; CATHETER ABLATION; IMAGE INTEGRATION; FIBRILLATION;
D O I
10.1111/pace.12095
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Current 3D mapping systems have difficulty rendering complex cardiac structures. Different electroanatomical mapping software has been recently developed which uses a mathematical algorithm to improve interpolation between mapped points and delineation of closely spaced structures. This study tested the feasibility and accuracy of this software in comparison to traditional software. Methods In vivo 3D impedance-based mapping using a multielectrode catheter with a single geometry point cloud was performed in the left atria and pulmonary veins (PV) in 23 patients undergoing catheter ablation for atrial fibrillation. The maps were analyzed with traditional (NavX, St. Jude Medical, Minnetonka, MN, USA), either with or without multichamber mapping versus St. Jude OneModel software and dimensions of cardiac chambers in human studies were compared to preprocedural computed tomographic (CT) or magnetic resonance (MR) scans to determine the relative accuracy of the maps. Results Maps created by the OneModel software provided greater detail of complex cardiac structures compared to traditional software. Comparison of the left atrial/pulmonary vein electroanatomical maps with the CT and MR scans as reference standard demonstrated significantly less error in measurement of all PV ostial long- and short-axis dimensions, inter-PV distance, and ridge width (left PV to left atrial appendage) with the OneModel versus traditional software (P < 0.001 for all dimensions measured). Conclusions The OneModel software produces maps that are more accurate in rendering complex cardiac structures compared to traditional software.
引用
收藏
页码:626 / 631
页数:6
相关论文
共 13 条
  • [1] Image integration increases efficacy of paroxysmal atrial fibrillation catheter ablation: results from the CartoMerge™ Italian Registry
    Bertaglia, Emanuele
    Della Bella, Paolo
    Tondo, Claudio
    Proclemer, Alessandro
    Bottoni, Nicola
    De Ponti, Roberto
    Landolina, Maurizio
    Bongiorni, Maria Grazia
    Coro, Leonardo
    Stabile, Giuseppe
    Dello Russo, Antonio
    Verlato, Roberto
    Mantica, Massimo
    Zoppo, Franco
    [J]. EUROPACE, 2009, 11 (08): : 1004 - 1010
  • [2] Calkins H, 2012, HEART RHYTHM
  • [3] EnSite Velocity™ cardiac mapping system: a new platform for 3D mapping of cardiac arrhythmias
    Eitel, Charlotte
    Hindricks, Gerhard
    Dagres, Nikolaos
    Sommer, Philipp
    Piorkowski, Christopher
    [J]. EXPERT REVIEW OF MEDICAL DEVICES, 2010, 7 (02) : 185 - 192
  • [4] A randomised comparison of Cartomerge vs. NavX fusion in the catheter ablation of atrial fibrillation: The CAVERN Trial
    Finlay, Malcolm C.
    Hunter, Ross J.
    Baker, Victoria
    Richmond, Laura
    Goromonzi, Farai
    Thomas, Glyn
    Rajappan, Kim
    Duncan, Edward
    Tayebjee, Muzahir
    Dhinoja, Mehul
    Sporton, Simon
    Earley, Mark J.
    Schilling, Richard J.
    [J]. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2012, 33 (02) : 161 - 169
  • [5] Factors affecting error in integration of electroanatomic mapping with CT and MR imaging during catheter ablation of atrial fibrillation
    Heist, E. Kevin
    Chevalier, Jianping
    Holmvang, Godtfred
    Singh, Jagmeet P.
    Ellinor, Patrick T.
    Milan, David J.
    D'Avila, Andre
    Mela, Theofanie
    Ruskin, Jeremy N.
    Mansour, Moussa
    [J]. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2006, 17 (01) : 21 - 27
  • [6] Impact of variant pulmonary vein anatomy and image integration on long-term outcome after catheter ablation for atrial fibrillation
    Hunter, Ross J.
    Ginks, Matthew
    Ang, Richard
    Diab, Ihab
    Goromonzi, Farai C.
    Page, Stephen
    Baker, Victoria
    Richmond, Laura
    Tayebjee, Muzahir
    Sporton, Simon
    Earley, Mark J.
    Schilling, Richard J.
    [J]. EUROPACE, 2010, 12 (12): : 1691 - 1697
  • [7] CARTO-guided vs. NavX-guided pulmonary vein antrum isolation and pulmonary vein antrum isolation performed without 3-D mapping: effect of the 3-D mapping system on procedure duration and fluoroscopy time
    Khaykin, Yaariv
    Oosthuizen, Richard
    Zarnett, Lauren
    Wulffhart, Zaev A.
    Whaley, Bonnie
    Hill, Carol
    Giewercer, David
    Verma, Atul
    [J]. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2011, 30 (03) : 233 - 240
  • [8] The impact of image integration on catheter ablation of atrial fibrillation using electroanatomic mapping: a prospective randomized study
    Kistler, Peter M.
    Rajappan, Kim
    Harris, Stuart
    Earley, Mark J.
    Richmond, Laura
    Sporton, Simon C.
    Schilling, Richard J.
    [J]. EUROPEAN HEART JOURNAL, 2008, 29 (24) : 3029 - 3036
  • [9] Accuracy of left atrial anatomical maps acquired with a multielectrode catheter during catheter ablation for atrial fibrillation
    Koruth, Jacob S.
    Heist, E. Kevin
    Danik, Stephan
    Barrett, Conor D.
    Kabra, Rajesh
    Blendea, Dan
    Ruskin, Jeremy
    Mansour, Moussa
    [J]. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2011, 32 (01) : 45 - 51
  • [10] Three-dimensional anatomy of the left atrium by magnetic resonance angiography: Implications for catheter ablation for atrial fibrillation
    Mansour, M
    Refaat, M
    Heist, EK
    Mela, T
    Cury, R
    Holmvang, G
    Ruskin, JN
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2006, 17 (07) : 719 - 723