Glucose; Temperature sensors; Temperature measurement; Humidity; Blood; Permittivity; Glucose sensor; microwave resonator; non-invasive detection; environmental effect elimination; switching circuit; COMPLEX PERMITTIVITY; HYPERGLYCEMIA; MANAGEMENT;
D O I
10.1109/JSEN.2020.2984779
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this paper, a microwave fluidic glucose sensor based on dual complementary split ring resonators (CSRRs) with a switching circuit is proposed for non-invasive and continuous glucose concentration detection. A CSRR has been designed to detect subtle changes in the dielectric constant of glucose solutions, and the electric field distribution is utilized to determine the effective sensing region. In a well-controlled environment, the glucose concentration can be detected by tracking the transmission coefficient of the designed CSRR. However, because the variations in dielectric constant with the glucose concentration are very small, the environmental factors (i.e., ambient temperature and relative humidity) can be dominant compared to the glucose concentration. To qualify the effect of the environmental factors, the variation in the transmission coefficient of the CSRR has been measured when the ambient temperature condition is varied from 293 K to 313 K and the relative humidity is varied from 30% to 90% inside temperature- and humidity-controlled chamber. Moreover, the environmental effects are calibrated out using two identical CSRRs with the switching circuit: one detects changes in the glucose concentration, while the other operates as a reference for environmental factors. After the elimination of the environmental effects, glucose concentrations in the range of 0 mg/dL to 400 mg/dL have been measured, and the sensitivity is 0.008dB/(100mg/dL). In addition, the reproducibility of the proposed sensor is verified by repeated measurements. It is demonstrated that the proposed sensor can detect glucose concentrations under practical environmental conditions.
引用
收藏
页码:8520 / 8527
页数:8
相关论文
共 38 条
  • [1] Sensitivity Enhanced Vital Sign Detection Based on Antenna Reflection Coefficient Variation
    An, Yong-Jun
    Yun, Gi-Ho
    Yook, Jong-Gwan
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2016, 10 (02) : 319 - 327
  • [2] Real-Time Noninvasive Measurement of Glucose Concentration Using a Microwave Biosensor
    Bababjanyan, Arsen
    Melikyan, Harutyun
    Kim, Seungwan
    Kim, Jongchel
    Lee, Kiejin
    Friedman, Barry
    [J]. JOURNAL OF SENSORS, 2010, 2010
  • [3] Feasibility of measuring blood glucose concentration by near-infrared Raman spectroscopy
    Berger, AJ
    Itzkan, I
    Feld, MS
    [J]. SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 1997, 53 (02) : 287 - 292
  • [4] High-Resolution, Far-Field, and Passive Temperature Sensing up to 700 °C Using an Isolated ZST Microwave Dielectric Resonator
    Boccard, Jean-Michel
    Aftab, Taimur
    Hoppe, Joachim
    Yousaf, Adnan
    Huetter, Ruediger
    Reindl, Leonhard M.
    [J]. IEEE SENSORS JOURNAL, 2016, 16 (03) : 715 - 722
  • [5] Material Characterization Using Complementary Split-Ring Resonators
    Boybay, Muhammed Said
    Ramahi, Omar M.
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2012, 61 (11) : 3039 - 3046
  • [6] Glucose Sensing for Diabetes Monitoring: Recent Developments
    Bruen, Danielle
    Delaney, Colm
    Florea, Larisa
    Diamond, Dermot
    [J]. SENSORS, 2017, 17 (08)
  • [7] Choi H, 2014, DIVINE DOMESTICITIES: CHRISTIAN PARADOXES IN ASIA AND THE PACIFIC, P1
  • [8] CHOROM JANG, 2018, [The Journal Of Korean Institute of Electromagnetic Engineering and Science, 한국전자파학회 논문지], V29, P247, DOI 10.5515/KJKIEES.2018.29.4.247
  • [9] Microwave-Based Microfluidic Sensor for Non-Destructive and Quantitative Glucose Monitoring in Aqueous Solution
    Chretiennot, Thomas
    Dubuc, David
    Grenier, Katia
    [J]. SENSORS, 2016, 16 (10)
  • [10] Microwave Sensor for Liquid Dielectric Characterization Based on Metamaterial Complementary Split Ring Resonator
    Chuma, Euclides Lourenco
    Iano, Yuzo
    Fontgalland, Glauco
    Bravo Roger, Leonardo Lorenzo
    [J]. IEEE SENSORS JOURNAL, 2018, 18 (24) : 9978 - 9983