Clinical Evaluation of the New Rebound Tonometers Icare PRO and Icare ONE Compared With the Goldmann Tonometer

被引:37
作者
Moreno-Montanes, Javier [1 ]
Martinez-de-la-Casa, Jose M. [2 ,3 ,4 ]
Sabater, Alfonso L. [1 ]
Morales-Fernandez, Laura [2 ,3 ,4 ]
Saenz, Cristina [1 ]
Garcia-Feijoo, Julian [2 ,3 ,4 ]
机构
[1] Univ Navarra Clin, Dept Ophthalmol, Pamplona 31080, Spain
[2] Hosp Clin San Carlos, Serv Oftalmol, Madrid, Spain
[3] Univ Complutense Madrid, Fac Med, Dept Oftalmol, Madrid, Spain
[4] Hosp Clin San Carlos, Inst Invest Sanitaria, Madrid, Spain
关键词
intraocular pressure; rebound tonometer; autotonometry; Goldmann tonometer; Icare tonometer; glaucoma; MEASURING INTRAOCULAR-PRESSURE; APPLANATION TONOMETRY; SUSPECTED GLAUCOMA; FLUCTUATION; CHILDREN;
D O I
10.1097/IJG.0000000000000058
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose:To compare the participant-obtained intraocular pressures (IOPs) using Icare ONE and the clinician-obtained values using Icare PRO, both rebound tonometers, with Goldmann tonometry (GAT) values and analyze the ease of use of Icare ONE.Methods:One hundred fifty participants were included (60 normal controls, 90 patients with glaucoma/ocular hypertension). The participants measured the IOP 3 times using Icare ONE; a clinician measured the IOP once using Icare PRO. The instruments were used randomly. Clinical data were evaluated to analyze the difficulty of the technique, the effect on the results, and the ease of use of Icare ONE.Results:The mean IOPs with GAT, Icare ONE, and Icare PRO were 16.64.43, 17.5 +/- 5.42, and 16.6 +/- 4.77 mm Hg, respectively. The participant-measured IOP values were within +3 mm Hg of the GAT values in 67.1% of eyes with Icare ONE and in 79.6% with Icare PRO. The limits of agreement were higher with Icare ONE compared with Icare PRO. IOP value errors were found with Icare ONE in eyes with low and high GAT-IOP. The areas under the curve to detect IOPs of 21 mm Hg or higher (GAT) exceeded 0.80 with both tonometers. Young participants reported better ease of use with Icare ONE. No other factors were related to the results.Conclusions:Icare ONE may be useful for patients monitoring their IOP values; most individuals can use the device after a short training session. Icare PRO had better results compared with Icare ONE in all IOP ranges.
引用
收藏
页码:527 / 532
页数:6
相关论文
共 26 条
[11]   24-hour monitoring of intraocular pressure in glaucoma management: A retrospective review [J].
Hughes, E ;
Spry, P ;
Diamond, J .
JOURNAL OF GLAUCOMA, 2003, 12 (03) :232-236
[12]   Pascal, ICare and Goldmann applanation tonometry -: a comparative study [J].
Johannesson, Gauti ;
Hallberg, Per ;
Eklund, Anders ;
Linden, Christina .
ACTA OPHTHALMOLOGICA, 2008, 86 (06) :614-621
[13]  
Jonas Jost B, 2010, J Ophthalmic Vis Res, V5, P92
[14]   Fluctuation of Intraocular Pressure in 24-Hour Telemonitoring Compared to Tonometry during Normal Office Hours [J].
Juergens, C. ;
Antal, S. ;
Henrici, K. ;
Grossjohann, R. ;
Tost, F. H. .
KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2009, 226 (01) :54-59
[15]   Measuring intraocular pressure with the Pulsair 3000 and Rebound tonometers in elderly patients without an anesthetic [J].
Kontiola, A ;
Puska, P .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2004, 242 (01) :3-7
[16]   A new induction-based impact method for measuring intraocular pressure [J].
Kontiola, AI .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2000, 78 (02) :142-145
[17]   The induction/impact tonometer: A new instrument to measure intraocular pressure in the rat [J].
Kontiola, AI ;
Goldblum, D ;
Mittag, T ;
Danias, J .
EXPERIMENTAL EYE RESEARCH, 2001, 73 (06) :781-785
[18]   Reproducibility and clinical evaluation of rebound tonometry [J].
Martinez-de-la-Casa, JM ;
Garcia-Feijoo, J ;
Castillo, A ;
Garcia-Sanchez, J .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46 (12) :4578-4580
[19]   Effect of corneal thickness on dynamic contour, rebound, and Goldmann tonometry [J].
Martinez-de-la-Casa, Jose M. ;
Garcia-Feijoo, Julian ;
Vico, Eva ;
Fernandez-Vidal, Ana ;
del Castillo, Jose M. Benitez ;
Wasfi, Mohamed ;
Garcia-Sanchez, J. .
OPHTHALMOLOGY, 2006, 113 (12) :2156-2162
[20]   Correlations between corneal biomechanical properties measured with the Ocular Response Analyzer and ICare rebound tonometry [J].
Martins Jorge, Jorge Manuel ;
Gonzalez-Meijome, Jose M. ;
Queiros, Antonio ;
Fernandes, Paulo ;
Parafita, Manuel A. .
JOURNAL OF GLAUCOMA, 2008, 17 (06) :442-448