Users' opinions on intensive care unit alarms - a survey of German intensive care units

被引:14
作者
Siebig, S. [1 ]
Sieben, W. [2 ]
Kollmann, F. [1 ]
Imhoff, M. [3 ]
Bruennler, T. [1 ]
Rockmann, F. [1 ]
Gather, U. [2 ]
Wrede, C. E. [1 ]
机构
[1] Univ Regensburg, Dept Internal Med 1, D-93042 Regensburg, Germany
[2] Univ Dortmund, Dept Stat, Dortmund, Germany
[3] Ruhr Univ Bochum, Dept Med Informat Biometr & Epidemiol, Bochum, Germany
关键词
intensive care unit; alarms; FALSE ALARMS; ICU; FREQUENCY;
D O I
10.1177/0310057X0903700116
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Monitoring of physiologic parameters in critically ill patients is associated with an enormous number of alarms, leading to reduced clinical value with high sensitivity but low specificity. To evaluate opinions of intensive care unit (ICU) staff on current monitoring we conducted. a survey of German ICUs. Furthermore, the survey aimed to assess requirements and requests for future alarm systems. The survey was conducted between May 2006 and June 2007 on a randomised sample of German ICUs. Questionnaires with 24 partly closed-ended partly open-ended questions were posted. Of 915 letters, 274 (30%) from 185 contacted ICUs were returned and evaluated. One hundred and sixty physicians, the majority (52%) working in a department of anaesthesiology, and 174 nurses returned the survey. Most responders (87%) estimated that less than 50% Of Current alarms result in clinical consequences (52% estimated less than 25%). We suggested trend alarms, smoothing of signals to reduce artefacts, generation of new combined alarms and integrative monitoring of different alarm systems as improvements of current ICU alarm systems, all of which were agreed to by the majority. Free text commentaries focused on the need for reducing alarms caused by artefacts and called for improvement of the monitor-user interfaces. Our survey demonstrates the dissatisfaction of clinical staff with the current alarm systems regarding alarm frequency and specificity in German ICUs, thereby confirming data raised in single institutions. ICU staffs acceptance for new alarm algorithms like signal extraction or detection of trends as a basis for smart monitoring appealed to the majority of users.
引用
收藏
页码:112 / 116
页数:5
相关论文
共 13 条
[1]   NOISE IN THE ICU [J].
BALOGH, D ;
KITTINGER, E ;
BENZER, A ;
HACKL, JM .
INTENSIVE CARE MEDICINE, 1993, 19 (06) :343-346
[2]   Multicentric study of monitoring alarms in the adult intensive care unit (ICU): a descriptive analysis [J].
Chambrin, MC ;
Ravaux, P ;
Calvelo-Aros, D ;
Jaborska, A ;
Chopin, C ;
Boniface, B .
INTENSIVE CARE MEDICINE, 1999, 25 (12) :1360-1366
[3]   Alarms in the intensive care unit: how can the number of false alarms be reduced? [J].
Chambrin, MC .
CRITICAL CARE, 2001, 5 (04) :184-188
[4]  
CHAMBRIN MC, 2003, P WORKSH INT DAT AN
[5]   On line extraction of temporal episodes from ICU high-frequency data: A visual support for signal interpretation [J].
Charbonnier, S .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2005, 78 (02) :115-132
[6]   NAME THAT TONE - THE PROLIFERATION OF ALARMS IN THE INTENSIVE-CARE UNIT [J].
CROPP, AJ ;
WOODS, LA ;
RANEY, D ;
BREDLE, DL .
CHEST, 1994, 105 (04) :1217-1220
[7]   Alarm algorithms in critical care monitoring [J].
Imhoff, M ;
Kuhls, S .
ANESTHESIA AND ANALGESIA, 2006, 102 (05) :1525-1537
[8]  
Korniewicz DM, 2008, AM J CRIT CARE, V17, P36
[9]   FREQUENCY AND RELIABILITY OF ALARMS IN THE MONITORING OF CARDIAC POSTOPERATIVE-PATIENTS [J].
KOSKI, EMJ ;
MAKIVIRTA, A ;
SUKUVAARA, T ;
KARI, A .
INTERNATIONAL JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1990, 7 (02) :129-133
[10]  
*LAND SADBUD, 2005, VERZ KRANK VERS REH