Evaluation of Computer-aided Diagnosis (CAD) Software for the Detection of Lung Nodules on Multidetector Row Computed Tomography (MDCT): JAFROC Study for the Improvement in Radiologists' Diagnostic Accuracy

被引:32
作者
Hirose, Tomohiro [1 ]
Nitta, Norihisa [1 ]
Shiraishi, Junji [2 ]
Nagatani, Yukihiro [1 ]
Takahashi, Masashi [1 ]
Murata, Kiyoshi [1 ]
机构
[1] Shiga Univ Med Sci, Dept Radiol, Otsu, Shiga 5202192, Japan
[2] Univ Chicago, Dept Radiol, Chicago, IL 60637 USA
关键词
Computer-aided diagnosis; CAD; lung nodule; jackknife free-response receiver operating characteristic analysis; JAFROC; multidetector-row computed tomography; MDCT;
D O I
10.1016/j.acra.2008.06.009
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. The aim of this study was to evaluate the usefulness of computer-aided diagnosis (CAD) software for the detection of lung nodules on multidetector-row computed tomography (MDCT) in terms of improvement in radiologists' diagnostic accuracy in detecting lung nodules, using jackknife free-response receiver-operating characteristic (JAFROC) analysis. Materials and Methods. Twenty-one patients (6 without and 15 with lung nodules) were selected randomly from 120 consecutive thoracic computed tomographic examinations. The gold standard for the presence or absence of nodules in the observer study was determined by consensus of two radiologists. Six expert radiologists participated in a free-response receiver operating characteristic study for the detection of lung nodules on MDCT, in which cases were interpreted first without and then with the output of CAD software. Radiologists were asked to indicate the locations of lung nodule candidates on the monitor with their confidence ratings for the presence of lung nodules. Results. The performance of the CAD software indicated that the sensitivity in detecting lung nodules was 71.4%, with 0.95 false-positive results per case. When radiologists used the CAD software, the average sensitivity improved from 39.5% to 81.0%, with an increase in the average number of false-positive results from 0.14 to 0.89 per case. The average figure-of-merit values for the six radiologists were 0.390 without and 0.845 with the output of the CAD software, and there was a statistically significant difference (P <.0001) using the JAFROC analysis. Conclusion. The CAD software for the detection of lung nodules on MDCT has the potential to assist radiologists by increasing their accuracy.
引用
收藏
页码:1505 / 1512
页数:8
相关论文
共 27 条
[1]   Computerized scheme for automated detection of lung nodules in low-dose computed tomography images for lung cancer screening [J].
Arimura, H ;
Katsuragawa, S ;
Suzuki, K ;
Li, F ;
Shiraishi, J ;
Sone, S ;
Doi, K .
ACADEMIC RADIOLOGY, 2004, 11 (06) :617-629
[2]   Computerized detection of pulmonary nodules on CT scans [J].
Armato, SG ;
Giger, ML ;
Moran, CJ ;
Blackburn, JT ;
Doi, K ;
MacMahon, H .
RADIOGRAPHICS, 1999, 19 (05) :1303-1311
[3]   Automated detection of lung nodules in CT scans: Preliminary results [J].
Armato, SG ;
Giger, ML ;
MacMahon, H .
MEDICAL PHYSICS, 2001, 28 (08) :1552-1561
[4]   MISSED BRONCHOGENIC-CARCINOMA - RADIOGRAPHIC FINDINGS IN 27 PATIENTS WITH A POTENTIALLY RESECTABLE LESION EVIDENT IN RETROSPECT [J].
AUSTIN, JHM ;
ROMNEY, BM ;
GOLDSMITH, LS .
RADIOLOGY, 1992, 182 (01) :115-122
[5]   Pulmonary nodules at chest CT: Effect of computer-aided diagnosis on radiologists' detection performance [J].
Awai, K ;
Murao, K ;
Ozawa, A ;
Komi, M ;
Hayakawa, H ;
Hori, S ;
Nishimura, Y .
RADIOLOGY, 2004, 230 (02) :347-352
[6]   Pulmonary nodules: Automated detection on CT images with morphologic matching algorithm preliminary results [J].
Bae, KT ;
Kim, JS ;
Na, YH ;
Kim, KG ;
Kim, JH .
RADIOLOGY, 2005, 236 (01) :286-294
[7]   Analysis of location specific observer performance data: Validated extensions of the jackknife free-response (JAFROC) method [J].
Chakraborty, Dev P. .
ACADEMIC RADIOLOGY, 2006, 13 (10) :1187-1193
[8]   Observer studies involving detection and localization: Modeling, analysis, and validation [J].
Chakraborty, DP ;
Berbaum, KS .
MEDICAL PHYSICS, 2004, 31 (08) :2313-2330
[9]   MAXIMUM-LIKELIHOOD ANALYSIS OF FREE-RESPONSE RECEIVER OPERATING CHARACTERISTIC (FROC) DATA [J].
CHAKRABORTY, DP .
MEDICAL PHYSICS, 1989, 16 (04) :561-568
[10]   Small pulmonary nodules:: Effect of two computer-aided detection systems on radiologist performance [J].
Das, Marco ;
Muehlenbruch, Georg ;
Mahnken, Andreas H. ;
Flohr, Thomas G. ;
Guendel, Lutz ;
Stanzel, Sven ;
Kraus, Thomas ;
Guenther, Rolf W. ;
Wildberger, Joachim E. .
RADIOLOGY, 2006, 241 (02) :564-571