Impeller inner diameter in a miniaturized centrifugal blood pump

被引:1
作者
Takano, T
Schulte-Eistrup, S
Kawahito, S
Maeda, T
Nonaka, K
Linneweber, J
Glueck, J
Fujisawa, A
Makinouchi, K
Yokokawa, M
Nosé, Y
机构
[1] Baylor Coll Med, Dept Surg, Houston, TX 77030 USA
[2] Kyocera Corp, Kyoto, Japan
关键词
centrifugal blood pump; impeller inner diameter; ventricular assist device; hemolysis; hydraulic performance; thrombogenicity;
D O I
10.1046/j.1525-1594.2002.6849_5.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To design a miniaturized centrifugal blood pump, the impeller internal diameter (ID), which is a circle diameter on the inner edge of the vane, is considered one of the important aspects. Hydraulic performance, hemolysis, and thrombogenicity were evaluated with different impeller IDs. Two impellers were fabricated with an outer diameter of 35 mm, of which 1 had an 8 mm ID impeller and the other had a 12 mm ID. These impellers were combined with 2 different housings in which the inlet port was eccentrically positioned 3.8 and 4.5 mm offset from the center. The hydraulic performance and hemolysis were evaluated in a mock circuit, and thrombogenicity was evaluated in a 2 day ex vivo study with each impeller housing combination. Both impellers required 3,000 rpm in the 3.8 mm offset inlet to attain 5 L/min against 100 mm Hg (left ventricular assist device condition). The 8 mm ID impeller required 3,200 rpm, and the 12 mm ID impeller required 3,100 rpm in the 4.5 mm offset housing. The normalized index of hemolysis was 0.0080 +/- 0.0048 g/100 L in the 8 mm ID impeller with the 3.8 mm offset and 0.022 +/- 0.018 g/100 L with 4.5 mm offset. The 12 mm ID impeller had 0.068 +/- 0.028 g/100 L with the 3.8 mm offset and 0.010 +/- 0.002 g/100 L with the 4.5 mm offset. After the 2 day ex vivo study, no blood clot was formed around the top bearing in all the pump heads. The 8 mm ID impeller with 3.8 mm offset inlet and the 12 mm ID impeller with the 4.5 mm offset had less hemolysis compared to the other pump heads that were subjected to 14 day ex vivo and 10 day in vivo studies. The 8 mm ID impeller with the 3.8 mm offset inlet had a blood clot around the top bearing after the 14 day ex vivo study. No thrombus was found around the top bearing of the 12 mm ID impeller with the 4.5 mm offset in the 10 day in vivo study. These results suggest that the ID does not greatly change the hydraulic performance of a small centrifugal blood pump. The proper combination of the impeller ID and inlet port offset obtains less hemolysis. The larger impeller ID is considered to have less thrombogenicity around the top bearing.
引用
收藏
页码:67 / 71
页数:5
相关论文
共 50 条
[41]   Flow rate estimation of a centrifugal blood pump using the passively stabilized eccentric position of a magnetically levitated impeller [J].
Shidai, Shuya ;
Masuzawa, Toru ;
Osa, Masahiro .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2019, 42 (06) :291-298
[42]   Design of a Centrifugal Blood Pump: Heart Turcica Centrifugal [J].
Demir, Onur ;
Biyikli, Emre ;
Lazoglu, Ismail ;
Kucukaksu, Suha .
ARTIFICIAL ORGANS, 2011, 35 (07) :720-U138
[43]   A magnetically suspended and hydrostatically stabilized centrifugal blood pump [J].
Hart, RM ;
Filipenco, VG ;
Kung, RTV .
ARTIFICIAL ORGANS, 1996, 20 (06) :591-596
[44]   Design of a Small Centrifugal Blood Pump With Magnetic Bearings [J].
Jahanmir, Said ;
Hunsberger, Andrew Z. ;
Ren, Zhaohui ;
Heshmat, Hooshang ;
Heshmat, Crystal ;
Tomaszewski, Michael J. ;
Walton, James F. .
ARTIFICIAL ORGANS, 2009, 33 (09) :714-726
[45]   Study of a Centrifugal Blood Pump in a Mock Loop System [J].
Uebelhart, Beatriz ;
da Silva, Bruno Utiyama ;
Fonseca, Jeison ;
Bock, Eduardo ;
Leme, Juliana ;
da Silva, Cibele ;
Leao, Tarcisio ;
Andrade, Aron .
ARTIFICIAL ORGANS, 2013, 37 (11) :946-949
[46]   Computational modeling of the Food and Drug Administration's benchmark centrifugal blood pump [J].
Good, Bryan C. ;
Manning, Keefe B. .
ARTIFICIAL ORGANS, 2020, 44 (07) :E263-E276
[47]   Impact of Ventricular Volume Waveform on Centrifugal Blood Pump for the Children with Cardiac Failure [J].
Han, Lu ;
Liu, Jinlong ;
Yu, Xiaoqing ;
Zhang, Haibo ;
Wang, Wei ;
Yang, Ming .
2014 7TH INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING AND INFORMATICS (BMEI 2014), 2014, :546-550
[48]   NUMERICAL INVESTIGATIONS OF A CENTRIFUGAL BLOOD PUMP [J].
Chan, W. K. ;
Wong, Y. W. ;
Ding, Y. .
JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2004, 4 (03) :237-255
[49]   Optimization of the hydraulic performance of a centrifugal pump impeller based on the blade load distribution [J].
Jiang, Wei ;
Li, Guojun ;
Zhang, Xinsheng .
Harbin Gongcheng Daxue Xuebao/Journal of Harbin Engineering University, 2015, 36 (04) :505-510
[50]   Particle image velocimetry in the impeller of a centrifugal pump: A POD-based analysis [J].
Fonseca, William D. P. ;
Cerqueira, Rafael F. L. ;
Perissinotto, Rodolfo M. ;
Verde, William Monte ;
Castro, Marcelo S. ;
Franklin, Erick M. .
FLOW MEASUREMENT AND INSTRUMENTATION, 2023, 94