Impeller inner diameter in a miniaturized centrifugal blood pump

被引:1
作者
Takano, T
Schulte-Eistrup, S
Kawahito, S
Maeda, T
Nonaka, K
Linneweber, J
Glueck, J
Fujisawa, A
Makinouchi, K
Yokokawa, M
Nosé, Y
机构
[1] Baylor Coll Med, Dept Surg, Houston, TX 77030 USA
[2] Kyocera Corp, Kyoto, Japan
关键词
centrifugal blood pump; impeller inner diameter; ventricular assist device; hemolysis; hydraulic performance; thrombogenicity;
D O I
10.1046/j.1525-1594.2002.6849_5.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To design a miniaturized centrifugal blood pump, the impeller internal diameter (ID), which is a circle diameter on the inner edge of the vane, is considered one of the important aspects. Hydraulic performance, hemolysis, and thrombogenicity were evaluated with different impeller IDs. Two impellers were fabricated with an outer diameter of 35 mm, of which 1 had an 8 mm ID impeller and the other had a 12 mm ID. These impellers were combined with 2 different housings in which the inlet port was eccentrically positioned 3.8 and 4.5 mm offset from the center. The hydraulic performance and hemolysis were evaluated in a mock circuit, and thrombogenicity was evaluated in a 2 day ex vivo study with each impeller housing combination. Both impellers required 3,000 rpm in the 3.8 mm offset inlet to attain 5 L/min against 100 mm Hg (left ventricular assist device condition). The 8 mm ID impeller required 3,200 rpm, and the 12 mm ID impeller required 3,100 rpm in the 4.5 mm offset housing. The normalized index of hemolysis was 0.0080 +/- 0.0048 g/100 L in the 8 mm ID impeller with the 3.8 mm offset and 0.022 +/- 0.018 g/100 L with 4.5 mm offset. The 12 mm ID impeller had 0.068 +/- 0.028 g/100 L with the 3.8 mm offset and 0.010 +/- 0.002 g/100 L with the 4.5 mm offset. After the 2 day ex vivo study, no blood clot was formed around the top bearing in all the pump heads. The 8 mm ID impeller with 3.8 mm offset inlet and the 12 mm ID impeller with the 4.5 mm offset had less hemolysis compared to the other pump heads that were subjected to 14 day ex vivo and 10 day in vivo studies. The 8 mm ID impeller with the 3.8 mm offset inlet had a blood clot around the top bearing after the 14 day ex vivo study. No thrombus was found around the top bearing of the 12 mm ID impeller with the 4.5 mm offset in the 10 day in vivo study. These results suggest that the ID does not greatly change the hydraulic performance of a small centrifugal blood pump. The proper combination of the impeller ID and inlet port offset obtains less hemolysis. The larger impeller ID is considered to have less thrombogenicity around the top bearing.
引用
收藏
页码:67 / 71
页数:5
相关论文
共 50 条
  • [21] Effects of gravity on flow rate estimations of a centrifugal blood pump using the eccentric position of a levitated impeller
    Shida, Shuya
    Masuzawa, Toru
    Osa, Masahiro
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2020, 43 (12) : 774 - 781
  • [22] Less blood damage in the impeller centrifugal pump: A comparative study with the roller pump in open heart surgery
    Nishinaka, T
    Nishida, H
    Endo, M
    Miyagishima, M
    Ohtsuka, G
    Koyanagi, H
    ARTIFICIAL ORGANS, 1996, 20 (06) : 707 - 710
  • [23] Numerical investigation on the effect of impeller axial position on hemodynamics of an extracorporeal centrifugal blood pump
    Lv, Shen
    He, Zhi-Peng
    Liu, Guang-Mao
    Hu, Sheng-Shou
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2024, 27 (13) : 1744 - 1755
  • [24] Flow visualization analysis of relative stagnation areas on the impeller surface of a centrifugal blood pump
    Aouidef A.
    Yamane T.
    Nishida M.
    Toyoda M.
    Tsukamoto Y.
    Ito K.
    Konishi Y.
    Masuzawa T.
    Tsukiya T.
    Taenaka Y.
    Journal of Artificial Organs, 2001, 4 (4) : 262 - 268
  • [25] Effects of tongue position and base circle diameter on the performance of a centrifugal blood pump
    Wong, Yew-Wah
    Chan, Weng-Kong
    Hu, Wei
    ARTIFICIAL ORGANS, 2007, 31 (08) : 639 - 645
  • [26] THE VIENNA IMPLANTABLE CENTRIFUGAL BLOOD PUMP
    SCHIMA, H
    TRUBEL, W
    WIESELTHALER, G
    SCHMIDT, C
    MULLER, MR
    SIEGL, H
    LOSERT, U
    WOLNER, E
    ARTIFICIAL ORGANS, 1994, 18 (07) : 500 - 505
  • [27] On the Optimization of a Centrifugal Maglev Blood Pump Through Design Variations
    Wu, Peng
    Huo, Jiadong
    Dai, Weifeng
    Wu, Wei-Tao
    Yin, Chengke
    Li, Shu
    FRONTIERS IN PHYSIOLOGY, 2021, 12
  • [28] Numerical study of blood pump weaning effects on hemocompatibility of centrifugal blood pump
    Xie N.
    Tang Y.
    Zhang Y.
    Liu Y.
    Beijing Hangkong Hangtian Daxue Xuebao/Journal of Beijing University of Aeronautics and Astronautics, 2023, 49 (07): : 1680 - 1688
  • [29] Development of a novel shrouded impeller pediatric blood pump
    Khan, Talha Irfan
    Zad, Haris Sheh
    Lazoglu, Ismail
    Yalcin, Ozlem
    JOURNAL OF ARTIFICIAL ORGANS, 2018, 21 (02) : 142 - 149
  • [30] A new blood pump for cardiopulmonary bypass: The HiFlow centrifugal pump
    Gobel, C
    Eilers, R
    Reul, H
    Schwindke, P
    Jorger, M
    Rau, G
    ARTIFICIAL ORGANS, 1997, 21 (07) : 841 - 845