Evaluation the Clinitek status™ automated dipstick analysis device for semiquantitative testing of canine urine

被引:24
作者
Bauer, N. [1 ]
Rettig, S. [1 ]
Moritz, A. [1 ]
机构
[1] Univ Giessen, Dept Vet Clin Sci Clin Pathol & Clin Pathophysiol, D-35392 Giessen, Germany
关键词
Dog; Laboratory information system; Sediment; Sensitivity; specificity; Ketone;
D O I
10.1016/j.rvsc.2008.02.003
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
The aim of this prospective study was to evaluate the Clinitek status (TM) analyser using Multistix10SG (TM)/Microalbustix (TM) dipsticks (all: Siemens Dx) for canine urine (n = 101) compared to reference methods: visual reading (Combur9 dipstick, Roche), refractometry, microscopy, and quantitative protein/creatinine analysis (Pentra400, AxonLab). The automated analyses were done twice and visual tests were performed by two examiners. An excellent to good concordance was demonstrated between the first/second analysis with the Multistix10SG and the Combur9 dipstick, respectively with Cohen's K-values ranging from 0.776 to 1.000. Agreement between both dipsticks was good for glucose (kappa = 0.753), blood (kappa = 0.793), protein (kappa = 0.788), and moderate for bilirubin (kappa = 0.431) and ketones (kappa = 0.540). In 6/101 specimens, false positive ketone reactions were obtained with the Multistix10SG (TM). Multistix10SG (TM) could not be used for determination of pyuria or specific gravity. Semiquantitative/quantitative protein results correlated well (rho = 0.90) and creatinine measurements moderately (rho = 0.76). Due to automated data transmission to the laboratory information system, the Clinitek status (TM) is of advantage in veterinary laboratories/clinics. (C) 2008 Elsevier Ltd. All rights reserved.
引用
收藏
页码:467 / 472
页数:6
相关论文
共 28 条
[21]   Urinary acetoacetate or capillary β-hydroxybutyrate for the diagnosis of ketoacidosis in the Emergency Department setting [J].
Taboulet, Pierre ;
Haas, Laurent ;
Porcher, Raphael ;
Manamani, Jaffar ;
Fontaine, Jean-Paul ;
Feugeas, Jean-Paul ;
Gautier, Jean-Francois .
EUROPEAN JOURNAL OF EMERGENCY MEDICINE, 2004, 11 (05) :251-258
[22]  
Tighe P, 1999, BRIT J BIOMED SCI, V56, P6
[23]  
VAIL DM, 1986, J AM VET MED ASSOC, V189, P1451
[24]   Comparison of multistix PRO dipsticks with other biochemical assays for determining urine protein (UP), urine creatinine (UC), and UP:UC ratio in dogs and cats [J].
Welles, EG ;
Whatley, EM ;
Hall, AS ;
Wright, JC .
VETERINARY CLINICAL PATHOLOGY, 2006, 35 (01) :31-36
[25]  
Yuen S. F., 2001, Hong Kong Medical Journal, V7, P5
[26]  
2007, MANUAL HANDREFRACTOM
[27]  
2004, COMBUR9 TEST R
[28]  