Optimizing stent expansion with new stent delivery systems

被引:25
作者
Takano, Y [1 ]
Yeatman, LA [1 ]
Higgins, JR [1 ]
Currier, JW [1 ]
Ascencio, E [1 ]
Kopelson, KA [1 ]
Tobis, JM [1 ]
机构
[1] Univ Calif Los Angeles, Hlth Sci Ctr, Los Angeles, CA 90024 USA
关键词
D O I
10.1016/S0735-1097(01)01618-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES The purpose of this study was to assess whether the newer stent delivery systems provide a stented lumen cross-sectional area (CSA) that is equal to the delivery balloon nominal dimensions. BACKGROUND First generation stents were often not adequately expanded with their delivery system and frequently required higher pressure or a larger balloon after deployment. Newer stents were designed to optimize expansion with noncompliant, high-pressure balloons provided as the delivery systems. METHODS Intravascular ultrasound (IVUS) was used to evaluate 38 stents in 32 patients after deployment at 14 to 16 atm with their delivery balloon system. Minimum stent lumen CSA and stent minimum lumen diameter (MLD) were measured by IVUS imaging. The manufacturer's expected stent diameter was defined as the balloon diameter measured by the company at the maximum pressure used. The manufacturer's expected stent area was calculated based on the manufacturer's expected stent diameter. RESULTS The MLD (2.5 +/- 0.5 mm) and minimum stent CSA (6.0 +/- 1.7 mm(2)) by, IVUS were significantly smaller than the manufacturer's expected stent diameter (3.5 +/- 0.4 mm) and area (9.5 +/- 1.9 MM2) (p < 0.0001, respectively). The mean MLD by IVUS was 72 +/- 8% of the expected stent diameter, and the mean minimum stent CSA by IVUS was 62 +/- 10% of the expected stent area. CONCLUSIONS Despite moderately high-pressure inflations, the mean minimum stent CSA actually achieved was, on average, only 62% of the manufacturer's expected stent area. To optimize stent deployment, these IVUS observations should be considered during coronary artery stenting. (C) 2001 by the American College of Cardiology.
引用
收藏
页码:1622 / 1627
页数:6
相关论文
共 21 条
  • [1] Comparison of immediate and intermediate-term results of intravascular ultrasound versus angiography-guided Palmaz-Schatz stent implantation in matched lesions
    Albiero, R
    Rau, T
    Schluter, M
    DiMario, C
    Reimers, B
    Mathey, DG
    Tobis, JM
    Schofer, J
    Colombo, A
    [J]. CIRCULATION, 1997, 96 (09) : 2997 - 3005
  • [2] Mechanisms of residual lumen stenosis after high-pressure stent implantation -: A quantitative coronary angiography and intravascular ultrasound study
    Bermejo, J
    Botas, J
    García, E
    Elízaga, J
    Osende, J
    Soriano, J
    Abeytua, M
    Delcán, JL
    [J]. CIRCULATION, 1998, 98 (02) : 112 - 118
  • [3] Blasini R, 1997, CATHETER CARDIO DIAG, V42, P113, DOI 10.1002/(SICI)1097-0304(199710)42:2<113::AID-CCD2>3.0.CO
  • [4] 2-G
  • [5] Incomplete expansion of Palmaz-Schatz stents despite high-pressure implantation technique:: Impact on target lesion revascularization
    Blessing, E
    Hausmann, D
    Sturm, M
    Wolpers, HG
    Amende, I
    Mügge, A
    [J]. CARDIOLOGY, 1999, 91 (02) : 102 - 108
  • [6] Quantitative coronary angiographic and intravascular ultrasound assessment of a new nonarticulated stent: Report from the advanced cardiovascular systems MultiLink stent pilot study
    Carrozza, JP
    Hermiller, JB
    Linnemeier, TJ
    Popma, JJ
    Yock, PG
    Roubin, GS
    Dean, LS
    Kuntz, RE
    Robertson, L
    Ho, KKL
    Cutlip, DE
    Baim, DS
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (01) : 50 - 56
  • [7] INTRACORONARY STENTING WITHOUT ANTICOAGULATION ACCOMPLISHED WITH INTRAVASCULAR ULTRASOUND GUIDANCE
    COLOMBO, A
    HALL, P
    NAKAMURA, S
    ALMAGOR, Y
    MAIELLO, L
    MARTINI, G
    GAGLIONE, A
    GOLDBERG, SL
    TOBIS, JM
    [J]. CIRCULATION, 1995, 91 (06) : 1676 - 1688
  • [8] Reference chart derived from post-stent-implantation intravascular ultrasound predictors of 6-month expected restenosis on quantitative coronary angiography
    de Feyter, PJ
    Kay, P
    Disco, C
    Serruys, PW
    [J]. CIRCULATION, 1999, 100 (17) : 1777 - 1783
  • [9] Intravascular ultrasound-guided optimized stent deployment - Immediate and 6 months clinical and angiographic results from the Multicenter Ultrasound Stenting in Coronaries Study (MUSIC Study)
    de Jaegere, P
    Mudra, H
    Figulla, H
    Almagor, Y
    Doucet, S
    Penn, I
    Colombo, A
    Hamm, C
    Bartorelli, A
    Rothman, M
    Nobuyoshi, M
    Yamaguchi, T
    Voudris, V
    DiMario, C
    Makovski, S
    Hausmann, D
    Rowe, S
    Rabinovich, S
    Sunamura, M
    van Es, GA
    [J]. EUROPEAN HEART JOURNAL, 1998, 19 (08) : 1214 - 1223
  • [10] INTRAVASCULAR ULTRASOUND VERSUS ANGIOGRAPHY FOR MEASUREMENT OD LUMINAL DIAMETERS IN NORMAL AND DISEASED CORONARY-ARTERIES
    DESCHEERDER, I
    DEMAN, F
    HERREGODS, MC
    WILCZEK, K
    BARRIOS, L
    RAYMENANTS, E
    DESMET, W
    DEGEEST, H
    PIESSENS, J
    [J]. AMERICAN HEART JOURNAL, 1994, 127 (02) : 243 - 251