Needle Tip Visibility in 3D Ultrasound Images

被引:25
作者
Arif, Muhammad [1 ,2 ]
Moelker, Adriaan [2 ]
van Walsum, Theo [1 ,2 ]
机构
[1] Univ Med Ctr Rotterdam, Erasmus MC, Dept Med Informat, Wytemaweg 80,Room 2506 Erasmus MC, NL-3015 CN Rotterdam, Netherlands
[2] Univ Med Ctr Rotterdam, Erasmus MC, Dept Radiol & Nucl Med, Rotterdam, Netherlands
关键词
Needle tip; Ultrasound; Visibility; Contrast-to-noise-ratio (CNR); Image guidance; ULTRASONOGRAPHIC ECHOGENICITY; IN-VITRO; VISUALIZATION; STANDARD;
D O I
10.1007/s00270-017-1798-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Needle visibility is crucial for effective and safe ultrasound-guided interventional procedures. Several studies have investigated needle visibility in 2D ultrasound imaging, but less information is available for 3D ultrasound imaging, a modality that has great potential for image guidance interventions. We performed a prospective study, to quantitatively compare the echogenicity of various commercially available needles in 3D ultrasound images used in clinical practice under freehand needle introduction. A set of seven needles, containing biopsy needles, a TIPS needle, an ablation needle and a puncture needle, were included in the study. A liver-mimicking phantom and cow liver were punctured by each needle. 3D sweeps and real-time 3D data were acquired at three different angles (20A degrees, 55A degrees and 90A degrees). Needle visibility was quantified by calculating contrast-to-noise ratio. In the liver-mimicking phantom, all needles showed better visibility than in the cow liver. At large angles, contrast-to-noise ratio and needle visibility were almost similar in both cases, but at lower angles differences in visibility were observed with different types of needles. The contrast-to-noise ratio increased with the increase in angle of insonation. The difference in visibility of different needles is more pronounced at 20A degrees angle. The echogenic properties of inhomogeneous cow liver tissues make the needles visibility worse as compared to a homogenous phantom. The needle visibility becomes worse in 3D real-time data as compared to 3D ultrasound sweeps.
引用
收藏
页码:145 / 152
页数:8
相关论文
共 12 条
[1]  
[Anonymous], THESIS
[2]   ESTIMATION OF URINARY FLOW VELOCITY IN MODELS OF OBSTRUCTED AND UNOBSTRUCTED URETHRAS BY DECORRELATION OF ULTRASOUND RADIOFREQUENCY SIGNALS [J].
Arif, Muhammad ;
Idzenga, Tim ;
van Mastrigt, Ron ;
de Korte, Chris L. .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2014, 40 (05) :938-946
[3]   4D Ultrasound Tracking of Liver and its Verification for TIPS Guidance [J].
Banerjee, Jyotirmoy ;
Klink, Camiel ;
Niessen, Wiro J. ;
Moelker, Adriaan ;
van Walsum, Theo .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2016, 35 (01) :52-62
[4]   An in-vitro study to understand successful free-hand ultrasound guided intervention [J].
Bradley, MJ .
CLINICAL RADIOLOGY, 2001, 56 (06) :495-498
[5]   Visualisation of needle position using ultrasonography [J].
Chapman, GA ;
Johnson, D ;
Bodenham, AR .
ANAESTHESIA, 2006, 61 (02) :148-158
[6]   Improvement of Sonographic needle visibility in cirrhotic livers during Transjugular intrahepatic Portosystemic Stent-shunt procedures with use of real-time compound Imaging [J].
Cohnen, M ;
Saleh, A ;
Lüthen, R ;
Bode, J ;
Mödder, U .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2003, 14 (01) :103-106
[7]   Relative ultrasonographic echogenicity of standard, dimpled, and polymeric-coated needles [J].
Culp, WC ;
McCowan, TC ;
Goertzen, TC ;
Habbe, TG ;
Hummel, MM ;
LeVeen, RF ;
Anderson, JC .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2000, 11 (03) :351-358
[8]   In-vitro visualization of biopsy needles with ultrasound:: A comparative study of standard and echogenic needles using an ultrasound phantom [J].
Hopkins, RE ;
Bradley, M .
CLINICAL RADIOLOGY, 2001, 56 (06) :499-502
[9]  
Jones CD, 1997, J ULTRAS MED, V16, P269
[10]   Evaluation of vascular puncture needles with specific modifications for enhanced ultrasound visibility: In vitro study [J].
Kawai, Nobuyuki ;
Minamiguchi, Hiroki ;
Sato, Morio ;
Nakai, Motoki ;
Sanda, Hiroki ;
Tanaka, Takami ;
Ikoma, Akira ;
Nakata, Kouhei ;
Shirai, Shintaro ;
Sonomura, Tetsuo .
WORLD JOURNAL OF RADIOLOGY, 2012, 4 (06) :273-277