A Comparison of the Corrected Intraocular Pressure Obtained by the Corvis ST and Reichert 7CR Tonometers in Glaucoma Patients

被引:12
作者
Nakao, Yoshitaka [1 ]
Kiuchi, Yoshiaki [1 ]
Okimoto, Satoshi [1 ]
机构
[1] Hiroshima Univ, Ophthalmol & Visual Sci Dept, Hiroshima, Japan
来源
PLOS ONE | 2017年 / 12卷 / 01期
关键词
OCULAR RESPONSE ANALYZER; CENTRAL CORNEAL THICKNESS; GOLDMANN APPLANATION TONOMETRY; OPEN-ANGLE GLAUCOMA; BIOMECHANICAL PROPERTIES; NONCONTACT TONOMETRY; DYNAMIC CONTOUR; EYES; REPRODUCIBILITY; CURVATURE;
D O I
10.1371/journal.pone.0170206
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
The purpose of the study was to investigate the accuracy of two corrected intraocular pressure (IOP) measurements by Corvis Scheimpflug Technology (CST)-IOPpachy and by corneal compensated lOP (IOPcc) using the Reichert 7CR (7CR) tonometers. We also investigated the effects of corneal anatomical and structural parameters on the lOP measurements. The participants included 90 primary open-angle glaucoma patients. We assessed the lOP measurements, obtained by the CST, 7CR, and Goldmann applanation tonometer (GAT), using a paired t-test with Bonferroni correction, Bland-Altman plots, and multiple regression analyses. The 7CR-IOPcc gave the highest value (15.5 +/- 2.7 mmHg), followed by the 7CR-IOPg (13.7 3.1 mmHg), GAT-IOP (13.6 +/- 2.2 mmHg), CST-IOP (10.3 +/- 2.6 mmHg), and CST-IOPpachy (9.7 +/- 2.5 mmHg). The values of CST-IOPpachy were significantly lower than those obtained by the other lOP measurement methods (all, p < 0.01). The values of 7CR-IOPcc were significantly higher than those obtained by the other lOP measurement methods (all, p < 0.01). Bland-Altman plots showed a mean difference between the GAT-IOP and the other lOP measurements (CST-IOP, CST-IOPpachy, 7CR-IOPg, and 7CR-IOPcc), which were -3.20, -3.82, 0.14, and 2.00 mmHg, respectively. The widths of the 95% limits of agreement between all pairs of IOP measurements were greater than 3 mmHg. With the exception of the 7CR-IOPcc, all of the lOP variations were explained by regression coefficients involving gender, average corneal curvature, and central corneal thickness. The lOP values obtained by the GAT, CST, and 7CR were not interchangeable. Each new IOP measurement device that was corrected for ocular structure had its own limitations.
引用
收藏
页数:12
相关论文
共 33 条
  • [21] Comparison of additional intraocular pressure-lowering effects of latanoprost vs brimonidine in primary open-angle glaucoma patients with intraocular pressure uncontrolled by timolol–dorzolamide combination
    A Akman
    A Cetinkaya
    Y A Akova
    A Ertan
    Eye, 2005, 19 : 145 - 151
  • [22] Comparison of Visual Field Progression Between Relatively Low and High Intraocular Pressure Groups in Normal Tension Glaucoma Patients
    Lee, Jeongmin
    Kong, Mingui
    Kim, Jaeryung
    Kee, Changwon
    JOURNAL OF GLAUCOMA, 2014, 23 (08) : 553 - 560
  • [23] Comparison of additional intraocular pressure-lowering effects of latanoprost vs brimonidine in primary open-angle glaucoma patients with intraocular pressure uncontrolled by timolol dorzolamide-combination
    Akman, A
    Cetinkaya, A
    Akova, Y
    Ertan, A
    EYE, 2005, 19 (02) : 145 - 151
  • [24] Comparison of self-measured diurnal intraocular pressure profiles using rebound tonometry between primary angle closure glaucoma and primary open angle glaucoma patients
    Tan, Shaoying
    Baig, Nafees
    Hansapinyo, Linda
    Jhanji, Vishal
    Wei, Shihui
    Tham, Clement C.
    PLOS ONE, 2017, 12 (03):
  • [25] Pattern of intraocular pressure reduction following laser trabeculoplasty in open-angle glaucoma patients: comparison between selective and nonselective treatment
    Almeida Junior, Eglailson Dantas
    Pinto, Luciano Moreira
    Brant Fernandes, Rodrigo Antonio
    Prata, Tiago Santos
    CLINICAL OPHTHALMOLOGY, 2011, 5 : 933 - 936
  • [26] Comparison of localized retinal nerve fiber layer defects between a low-teen Intraocular pressure group and a high-teen intraocular pressure group in normal-tension glaucoma patients
    Kim, Dong Myung
    Seo, Je Hyun
    Kim, Seok Hwan
    Hwang, Seung-Sik
    JOURNAL OF GLAUCOMA, 2007, 16 (03) : 293 - 296
  • [27] Comparison of the Intraocular Pressure Variation Provoked by Postural Change and by the Water Drinking Test in Primary Open-angle Glaucoma and Normal Patients
    Hatanaka, Marcelo
    Sakata, Lisandro M.
    Susanna, Remo, Jr.
    Nascimento, Luisa T. F.
    Vessani, Roberto M.
    JOURNAL OF GLAUCOMA, 2016, 25 (11) : 914 - 918
  • [28] A comparison of the effects on intraocular pressure of latanoprost 0.005% and the fixed combination of dorzolamide 2% and timolol 0.5% in patients with open-angle glaucoma
    Honrubia, FM
    Larsson, LI
    Spiegel, D
    ACTA OPHTHALMOLOGICA SCANDINAVICA, 2002, 80 (06): : 635 - 641
  • [29] Comparison between provocative test-based and long-term intraocular pressure parameters in patients with stable open-angle glaucoma
    Almeida, Izabela
    Scoralick, Ana Luiza B.
    Dias, Diego T.
    Ushida, Michele
    Dorairaj, Syril
    Gracitelli, Carolina P. B.
    Paranhos Jr, Augusto
    Kanadani, Fabio N.
    Prata, Tiago S.
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2021, 31 (02) : 453 - 459
  • [30] Comparison of Intraocular Pressure Changes During the Water Drinking Test Between Different Fluid Volumes in Patients With Primary Open-angle Glaucoma
    Susanna, Carolina N.
    Susanna, Remo, Jr.
    Hatanaka, Marcelo
    Susanna, Bianca N.
    Susanna, Fernanda N.
    De Moraes, Carlos G.
    JOURNAL OF GLAUCOMA, 2018, 27 (11) : 950 - 956