Real-time frequency-domain combined temperature and oxygen sensor using a single optical fiber

被引:0
|
作者
Liao, SC [1 ]
Xu, Z [1 ]
Izatt, JA [1 ]
Alcala, JR [1 ]
机构
[1] Case Western Reserve Univ, Dept Biomed Engn, Cleveland, OH 44106 USA
关键词
fiber sensors; oxygen; temperature; lifetime; phosphorescence; frequency-domain;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The combined excited state phosphorescence lifetimes of alexandrite and platinum tetraphenylporphyrin Pt(TPP) were used to remotely monitor temperature and oxygen concentration simultaneously using a single-optical-fiber probe. Temperature and oxygen concentration were successfully monitored in the physiological range from 15-45 degrees C and 0-50% O-2 with precisions of 0.24 degrees C and 0.15% O-2, and accuracies of 0.28 degrees C, 0.2% O-2. The dual-parameter sensor comprised a 500 mu m cubic alexandrite crystal bound to the distal end of a 750 mu m diameter plastic optical fiber core which was circumferentially coated with a 1 cm length of Pt(TPP), Sensor luminescence was excited using pulsed super bright blue LED light, and a frequency-domain representation of the time-domain phosphorescence decay was obtained by using fast Fourier transform algorithms. At least four log-spaced harmonic phases were averaged before decoding the two lifetimes of the temperature and oxygen phosphorescent sensors. A component of zero lifetime has introduced to account for the excitation backscatter leakage. Linear and second order empirical polynomials were employed to compute the temperatures and oxygen concentrations from the inverse lifetimes. The system drift was 0.24 degrees C for the temperature measurement and 0.59% for the oxygen concentration measurement over 30 hours of continuous operation. The instrumentation and methods allow for 6-s update times and 90-s full response times.
引用
收藏
页码:2333 / 2336
页数:4
相关论文
共 50 条
  • [31] Comparison of delay compensation methods for real-time hybrid simulation using frequency-domain evaluation index
    Xu Weijie
    Tong, Guo
    Cheng, Chen
    EARTHQUAKE ENGINEERING AND ENGINEERING VIBRATION, 2016, 15 (01) : 129 - 143
  • [32] Comparison of delay compensation methods for real-time hybrid simulation using frequency-domain evaluation index
    Xu Weijie
    Guo Tong
    Chen Cheng
    Earthquake Engineering and Engineering Vibration, 2016, 15 (01) : 129 - 143
  • [33] Real-Time Dynamic Assessment Of Airway Behavior Using Optical Frequency Domain Imaging
    Suter, M. J.
    Kelly, V. J.
    Chee, A.
    Applegate, M. B.
    Tan, K. M.
    Hariri, L. P.
    Musch, G.
    Harris, R. S.
    Winkler, T.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2013, 187
  • [34] Fiber-optic distributed temperature sensor using incoherent optical frequency domain reflectometry
    Karamehmedovic, E
    Glombitza, U
    EMERGING OPTOELECTRONIC APPLICATIONS, 2004, 5363 : 107 - 115
  • [35] Optical frequency-domain reflectometry (OFDR) using an integrated fiber tunable filter
    Zhao, YH
    Chen, ZP
    de Boer, JF
    Nelson, JS
    COHERENCE DOMAIN OPTICAL METHODS IN BIOMEDICAL SCIENCE AND CLINICAL APPLICATIONS III, PROCEEDINGS OF, 1999, 3598 : 56 - 60
  • [36] Brillouin optical-fiber frequency-domain analysis for distributed temperature and strain measurements
    Garus, D
    Gogolla, T
    Krebber, K
    Schliep, F
    JOURNAL OF LIGHTWAVE TECHNOLOGY, 1997, 15 (04) : 654 - 662
  • [37] Brillouin optical-fiber frequency-domain analysis for distributed temperature and strain measurements
    Ruhr-Universität Bochum, Bochum D-44780, Germany
    不详
    不详
    不详
    不详
    不详
    J Lightwave Technol, 4 (654-662):
  • [38] Frequency-domain single-shot optical frequency comb tomography using VIPA
    Miyaoka, Takumi
    Shioda, Tatsutoshi
    PHOTONIC INSTRUMENTATION ENGINEERING III, 2016, 9754
  • [39] Physical significance of frequency-domain based evaluation method for real-time hybrid simulation
    Xu, Weijie
    Guo, Tong
    Chen, Cheng
    Dongnan Daxue Xuebao (Ziran Kexue Ban)/Journal of Southeast University (Natural Science Edition), 2013, 43 (05): : 923 - 927
  • [40] Gamma/neutron classification with SiPM CLYC detectors using frequency-domain analysis for embedded real-time applications
    Morales, Ivan Rene
    Crespo, Maria Liz
    Bogovac, Mladen
    Cicuttin, Andres
    Kanaki, Kalliopi
    Carrato, Sergio